摘要
Developing mechanical devices to restore natural locomotion for transfemoral amputees still raises many challenges. One of them is the development of an efficient control strategy for the prosthesis active joints, with the objective of making it flexible and intuitive to use. This paper reports the modelling of the knee joint of the transfemoral CYBERLEGs Beta-Prosthesis, an actuated elastic ankle-knee prosthesis designed with the objective to minimize its total power consumption. More precisely, we report the static model of the prosthetic knee that was used to build its low-level controller, converting the desired torque into a reference position for the joint actuator. Bench testing of the prosthetic knee was performed to show that this static model is able to accurately predict the actual behaviour of the device.
引用
收藏
页码:776 / 782
页数:7
相关论文
共 30 条
  • [11] Energetic analysis and optimization of a MACCEPA actuator in an ankle prosthesis Energetic evaluation of the CYBERLEGs alpha-prosthesis variable stiffness actuator during a realistic load cycle
    Geeroms, Joost
    Flynn, Louis
    Jimenez-Fabian, Rene
    Vanderborght, Bram
    Lefeber, Dirk
    [J]. AUTONOMOUS ROBOTS, 2018, 42 (01) : 147 - 158
  • [12] The human ankle during walking: implications for design of biomimetic ankle prostheses
    Hansen, AH
    Childress, DS
    Miff, SC
    Gard, SA
    Mesplay, KP
    [J]. JOURNAL OF BIOMECHANICS, 2004, 37 (10) : 1467 - 1474
  • [13] Torque control of an active elastic transfemoral prosthesis via quasi-static modelling
    Heins, Sophie
    Flynn, Louis
    Geeroms, Joost
    Lefeber, Dirk
    Ronsse, Renaud
    [J]. ROBOTICS AND AUTONOMOUS SYSTEMS, 2018, 107 : 100 - 115
  • [14] Sliding-Bar MACCEPA for a Powered Ankle Prosthesis
    Jimenez-Fabian, R.
    Flynn, L.
    Geeroms, J.
    Vitiello, N.
    Vanderborght, B.
    Lefeber, D.
    [J]. JOURNAL OF MECHANISMS AND ROBOTICS-TRANSACTIONS OF THE ASME, 2015, 7 (04):
  • [15] A Robotic Leg Prosthesis Design, Control, and Implementation
    Lawson, Brian E.
    Mitchell, Jason E.
    Truex, Don
    Shultz, Amanda
    Ledoux, Elissa
    Goldfarb, Michael
    [J]. IEEE ROBOTICS & AUTOMATION MAGAZINE, 2014, 21 (04) : 70 - 81
  • [16] Limb Loss Task Force/Amputee Coalition (Knoxville Tennessee), 2012, ROADM PREV LIMB LOSS
  • [17] Ossur, RHEO KNEE
  • [18] Ossur, Power Knee
  • [19] Ossur, SYMB LEG 3
  • [20] Ottobock, C LEG KNEE PROSTH LE