Validity of automated measurement of left ventricular ejection fraction and volume using the Philips EPIQ system

被引:11
作者
Hovnanians, Ninel [1 ]
Win, Theresa [2 ]
Makkiya, Mohammed [3 ]
Zheng, Qi [3 ]
Taub, Cynthia [3 ]
机构
[1] Jacobi Med Ctr, Albert Einstein Coll Med, Dept Med, Bronx, NY USA
[2] Montefiore Med Ctr, Albert Einstein Coll Med, Dept Med, Bronx, NY 10467 USA
[3] Montefiore Med Ctr, Albert Einstein Coll Med, Dept Cardiol, Bronx, NY 10467 USA
来源
ECHOCARDIOGRAPHY-A JOURNAL OF CARDIOVASCULAR ULTRASOUND AND ALLIED TECHNIQUES | 2017年 / 34卷 / 11期
关键词
automated LVEF; cardiac volume; echocardiography; ECHOCARDIOGRAPHY; STANDARD;
D O I
10.1111/echo.13705
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
PurposeTo assess the efficiency and reproducibility of automated measurements of left ventricular (LV) volumes and LV ejection fraction (LVEF) in comparison to manually traced biplane Simpson's method. MethodThis is a single-center prospective study. Apical four- and two-chamber views were acquired in patients in sinus rhythm. Two operators independently measured LV volumes and LVEF using biplane Simpson's method. In addition, the image analysis software a2DQ on the Philips EPIQ system was applied to automatically assess the LV volumes and LVEF. Time spent on each analysis, using both methods, was documented. Concordance of echocardiographic measures was evaluated using intraclass correlation (ICC) and Bland-Altman analysis. ResultsManual tracing and automated measurement of LV volumes and LVEF were performed in 184 patients with a mean age of 67.317.3years and BMI 28.0 +/- 6.8kg/m(2). ICC and Bland-Altman analysis showed good agreements between manual and automated methods measuring LVEF, end-systolic, and end-diastolic volumes. The average analysis time was significantly less using the automated method than manual tracing (116 vs 217seconds/patient, P<.0001). ConclusionAutomated measurement using the novel image analysis software a2DQ on the Philips EPIQ system produced accurate, efficient, and reproducible assessment of LV volumes and LVEF compared with manual measurement.
引用
收藏
页码:1575 / 1583
页数:9
相关论文
共 8 条
  • [1] Assessment of Left Ventricular Volumes with Echocardiography and Cardiac Magnetic Resonance Imaging: Real-Life Evaluation of Standard versus New Semiautomatic Methods
    Aurich, Matthias
    Andre, Florian
    Keller, Marius
    Greiner, Sebastian
    Hess, Alexander
    Buss, Sebastian J.
    Katus, Hugo A.
    Mereles, Derliz
    [J]. JOURNAL OF THE AMERICAN SOCIETY OF ECHOCARDIOGRAPHY, 2014, 27 (10) : 1017 - 1024
  • [2] Analysis of Left Ventricular Volumes and Function: A Multicenter Comparison of Cardiac Magnetic Resonance Imaging, Cine Ventriculography, and Unenhanced and Contrast-Enhanced Two-Dimensional and Three-Dimensional Echocardiography
    Hoffmann, Rainer
    Barletta, Giuseppe
    von Bardeleben, Stephan
    Vanoverschelde, Jean Louis
    Kasprzak, Jaroslaw
    Greis, Christian
    Becher, Harald
    [J]. JOURNAL OF THE AMERICAN SOCIETY OF ECHOCARDIOGRAPHY, 2014, 27 (03) : 292 - 301
  • [3] Fully Automated Versus Standard Tracking of Left Ventricular Ejection Fraction and Longitudinal Strain The FAST-EFs Multicenter Study
    Knackstedt, Christian
    Bekkers, Sebastiaan C. A. M.
    Schummers, Georg
    Schreckenberg, Marcus
    Muraru, Denisa
    Badano, Luigi P.
    Franke, Andreas
    Bavishi, Chirag
    Omar, Alaa Mabrouk Salem
    Sengupta, Partho P.
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2015, 66 (13) : 1456 - 1466
  • [4] History of echocardiography and its future applications in medicine
    Krishnamoorthy, Vijay K.
    Sengupta, Partho P.
    Gentile, Federico
    Khandheria, Bijoy K.
    [J]. CRITICAL CARE MEDICINE, 2007, 35 (08) : S309 - S313
  • [6] Computer-assisted determination of left ventricular endocardial borders reduces variability in the echocardiographic assessment of ejection fraction
    Maret, Eva
    Brudin, Lars
    Lindstrom, Lena
    Nylander, Eva
    Ohlsson, Jan L.
    Engvall, Jan E.
    [J]. CARDIOVASCULAR ULTRASOUND, 2008, 6 (1)
  • [7] Reliability of reporting left ventricular systolic function by echocardiography: A systematic review of 3 methods
    McGowan, JH
    Cleland, JGF
    [J]. AMERICAN HEART JOURNAL, 2003, 146 (03) : 388 - 397
  • [8] Machine-Learning Algorithms to Automate Morphological and Functional Assessments in 2D Echocardiography
    Narula, Sukrit
    Shameer, Khader
    Omar, Alaa Mabrouk Salem
    Dudley, Joel T.
    Sengupta, Partho P.
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2016, 68 (21) : 2287 - 2295