Custom-fit total knee arthroplasty (OtisKnee) results in malalignment

被引:124
作者
Klatt, Brian A. [2 ,3 ]
Goyal, Nitin [2 ,3 ]
Austin, Matthew S. [2 ,3 ]
Hozack, William J. [1 ,2 ,3 ]
机构
[1] Thomas Jefferson Univ Hosp, Rothman Inst Orthopaed, Philadelphia, PA 19107 USA
[2] Thomas Jefferson Univ, Dept Orthopaed, Philadelphia, PA 19107 USA
[3] Rothman Inst, Dept Orthopaed, Philadelphia, PA USA
关键词
knee; total knee arthroplasty; custom guides; computer navigation;
D O I
10.1016/j.arth.2007.10.001
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
We report on 4 patients who underwent total knee arthroplasty with OtisKnee system (OtisMed, Hayward, Calif). An image-free computer navigation system was used to evaluate the deformities and the recommended ruts. The recommended custom cuts were as follows: valgus/varus cuts on the femur (5.5 degrees valgus to 0.5 degrees varus) in reference to the mechanical axis, flexion cuts on the femur (4 degrees-9 degrees of flexion); femoral rotation was within 1 degrees of the epicondylar axis; valgus/varus cut on the tibia (3 degrees of valgus to 7.5 degrees of varus); tibial slope (5.5 degrees of anterior slope to 0.5 degrees of posterior slope). The custom OtisKnee system guides recommended alignment of the components that was more than 3 degrees off of mechanical axis. The potential for malalignment with this system places implants at high risk of early failure.
引用
收藏
页码:26 / 29
页数:4
相关论文
共 20 条
[1]  
ANDERSON KC, 2005, J ARTHROPLASTY S3, V20, P7
[2]   Alignment in total knee arthroplasty -: A comparison of computer-assisted surgery with the conventional technique [J].
Bäthis, H ;
Perlick, L ;
Tingart, M ;
Lüring, C ;
Zurakowski, D ;
Grifka, J .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2004, 86B (05) :682-687
[3]  
Berend M, 2004, CLIN ORTHOP RELAT R, P428
[4]   Leg axis after computer-navigated total knee arthroplasty - A prospective randomized trial comparing computer-navigated and manual implantation [J].
Decking, R ;
Markmann, Y ;
Fuchs, J ;
Puhl, W ;
Scharf, HP .
JOURNAL OF ARTHROPLASTY, 2005, 20 (03) :282-288
[5]   The effects of varus tibial alignment on proximal tibial surface strain in total knee arthroplasty - The posteromedial hot spot [J].
Green, GV ;
Berend, KR ;
Berend, ME ;
Glisson, RR ;
Vail, TP .
JOURNAL OF ARTHROPLASTY, 2002, 17 (08) :1033-1039
[6]   Computer-assisted navigation increases precision of component placement in total knee arthroplasty [J].
Haaker, RG ;
Stockheim, M ;
Kamp, M ;
Proff, G ;
Breitenfelder, J ;
Ottersbach, A .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2005, (433) :152-159
[7]   CORONAL ALIGNMENT AFTER TOTAL KNEE REPLACEMENT [J].
JEFFERY, RS ;
MORRIS, RW ;
DENHAM, RA .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1991, 73 (05) :709-714
[8]  
JENNY JY, 2005, J ARTHROPLASTY, V20, P7
[9]   Hybrid control of smart structures using a novel wavelet-based algorithm [J].
Kim, HJ ;
Adeli, H .
COMPUTER-AIDED CIVIL AND INFRASTRUCTURE ENGINEERING, 2005, 20 (01) :7-22
[10]  
KLEIN GR, 2004, J ARTHROPLASTY, V19, P8