4D dose-position verification in radiation therapy using the RADPOS system in a deformable lung phantom

被引:21
作者
Cherpak, Amanda [1 ,2 ]
Serban, Monica [3 ]
Seuntjens, Jan [4 ]
Cygler, Joanna E. [1 ,2 ]
机构
[1] Ottawa Hosp Canc Ctr, Ottawa, ON K1H 8L6, Canada
[2] Carleton Univ, Dept Phys, Ottawa, ON K1S 5B6, Canada
[3] Hop Maisonneuve Rosemont Hosp, Serv Radiophys, Montreal, PQ H1T 2M4, Canada
[4] McGill Univ, Med Phys Unit, Montreal, PQ H3G 1A4, Canada
关键词
4D in vivo dosimetry; electromagnetic positioning system; MOSFET; deformable phantom; RADPOS; 4-DIMENSIONAL COMPUTED-TOMOGRAPHY; VOLUME HISTOGRAM ANALYSIS; IN-VIVO DOSIMETRY; RESPIRATORY MOTION; CANCER PATIENTS; TECHNICAL NOTE; PNEUMONITIS; RADIOTHERAPY; PHOTON; REGISTRATION;
D O I
10.1118/1.3515461
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: A novel 4D in vivo dosimetry system (RADPOS), in conjunction with a deformable lung phantom, has been evaluated as a potential quality assurance tool for 4D radiotherapy. Methods: RADPOS detectors, which consist of a MOSFET dosimeter combined with an electromagnetic positioning probe, were placed inside the deformable lung phantom. One detector was positioned directly inside a tumor embedded in the lung phantom and another was positioned inside the lung portion of the phantom, outside the tumor. CT scans were taken with the phantom at three breathing phases, and for each phase, the detector position inside the phantom was read with the RADPOS software and compared to the position as determined from the CT data. These values were also compared to RADPOS measurements taken with the phantom on the couch of a Varian Clinac 6EX linac. The deformable phantom and the RADPOS system were also used in two radiation delivery scenarios: (1) A simulation of a free-breathing delivery and (2) a simulation of an adaptive treatment. Results: Compared to CT imaging, the RADPOS positional accuracy was found to be better than 2.5 mm. The radial displacement measurements taken in the CT and linac rooms agreed to within an average of (0.7 +/- 0.3) mm. Hence, the system can provide relative displacement measurements in the treatment room, consistent with measurements made in the CT room. For the free-breathing delivery, the total dose reported by RADPOS agreed to within 4% and 5% of the treatment planning doses in the tumor and the lung portion of the phantom, respectively. The RADPOS-measured dose values for the adaptive delivery were within 1.5% of the treatment plan values, which was well within the estimated experimental uncertainties. Conclusions: This work has shown that the deformable lung phantom-RADPOS system can be an efficient quality assurance tool for 4D radiation therapy. (C) 2011 American Association of Physicists in Medicine. [DOI: 10.1118/1.3515461]
引用
收藏
页码:179 / 187
页数:9
相关论文
共 39 条
  • [31] Evaluation of the Effects of Dose on Lung Ventilation Calculated from 4D-CT using Deformable Image Registration An Application of a Ventilation Calculation Algorithm based on 4D-CT
    Latifi, Kujtim
    Dilling, Thomas J.
    Stevens, Craig W.
    Feygelman, Vladimir
    Moros, Eduardo G.
    Zhang, Geoffrey G.
    BIOIMAGING: PROCEEDINGS OF THE INTERNATIONAL CONFERENCE ON BIOIMAGING, 2014, : 5 - 11
  • [32] Quantitative assessment of intertarget position variations based on 4D-CT and 4D-CBCT simulations in single-isocenter multitarget lung stereotactic body radiation therapy
    Zhang, Siyu
    Guo, Chang
    Xu, Jun
    Qian, Pudong
    Guo, Jiali
    Liu, Tingting
    Wu, Yifan
    Hong, Jun
    Wang, Qi
    He, Xia
    Sun, Li
    JOURNAL OF CANCER RESEARCH AND CLINICAL ONCOLOGY, 2024, 150 (07)
  • [33] Real-Time Tumor Motion Tracking in 3D Using Planning 4D CT Images during Image-Guided Radiation Therapy
    Choi, Jang-Hwan
    Lee, Sooyeul
    ALGORITHMS, 2018, 11 (10):
  • [34] Quantifying the Effect of 4-Dimensional Computed Tomography-Based Deformable Dose Accumulation on Representing Radiation Damage for Patients with Locally Advanced Non-Small Cell Lung Cancer Treated with Standard-Fractionated Intensity-Modulated Radiation Therapy
    He, Yulun
    Cazoulat, Guillaume
    Wu, Carol
    Svensson, Stina
    Almodovar-Abreu, Lusmeralis
    Rigaud, Bastien
    McCollum, Emma
    Peterson, Christine
    Wooten, Zachary
    Rhee, Dong Joo
    Balter, Peter
    Pollard-Larkin, Julianne
    Cardenas, Carlos
    Court, Laurence
    Liao, Zhongxing
    Mohan, Radhe
    Brock, Kristy
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2024, 118 (01): : 231 - 241
  • [35] Detailed dosimetric evaluation of inter-fraction and respiratory motion in lung stereotactic body radiation therapy based on daily 4D cone beam CT images
    Huesa-Berral, Carlos
    Juan-Cruz, Celia
    van Kranen, Simon
    Rossi, Maddalena
    Belderbos, Jose
    Azcona, Juan Diego
    Burguete, Javier
    Sonke, Jan-Jakob
    PHYSICS IN MEDICINE AND BIOLOGY, 2023, 68 (01)
  • [36] Impact of tumor position displacement during end-exhalation breath-hold condition on tumor dose in lung stereotactic body radiation therapy using volumetric modulated arc therapy
    Kamima, Tatsuya
    Moriya, Shunsuke
    Sakae, Takeji
    Miyauchi, Hikaru
    Ito, Yasushi
    Tokumasu, Kenji
    Yoshioka, Yasuo
    CLINICAL AND TRANSLATIONAL RADIATION ONCOLOGY, 2025, 51
  • [37] Impact of a large breast separation on radiation dose delivery to the ipsilateral lung as result of respiratory motion quantified using free breathing and 4D CT-based planning in patients with locally advanced breast cancers: A potential for adverse clinical implications
    Heineman, Thomas E.
    Sabbas, Albert
    Delamerced, Marilynn Santos
    Chiu, Ya-lin
    Smith, Michael
    Parashar, Bhupesh
    Wernicke, A. Gabriella
    JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2013, 9 (01) : 154 - 160
  • [38] Movement assessment of breast and organ-at-risks using free-breathing, self-gating 4D magnetic resonance imaging workflow for breast cancer radiation therapy
    Habatsch, Melanie
    Schneider, Manuel
    Requardt, Martin
    Doussin, Sylvain
    PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2022, 22 : 111 - 114
  • [39] Investigation of interfractional variation in lung tumor position under expiratory-phase breath hold using cone-b eam compute d tomography in stereotactic body radiation therapy
    Miura, Hideharu
    Ozawa, Shuichi
    Nakao, Minoru
    Doi, Yoshiko
    Adachi, Yoshinori
    Kenjo, Masahiko
    Nagata, Yasushi
    MEDICAL DOSIMETRY, 2021, 46 (04) : 370 - 373