4D dose-position verification in radiation therapy using the RADPOS system in a deformable lung phantom

被引:21
|
作者
Cherpak, Amanda [1 ,2 ]
Serban, Monica [3 ]
Seuntjens, Jan [4 ]
Cygler, Joanna E. [1 ,2 ]
机构
[1] Ottawa Hosp Canc Ctr, Ottawa, ON K1H 8L6, Canada
[2] Carleton Univ, Dept Phys, Ottawa, ON K1S 5B6, Canada
[3] Hop Maisonneuve Rosemont Hosp, Serv Radiophys, Montreal, PQ H1T 2M4, Canada
[4] McGill Univ, Med Phys Unit, Montreal, PQ H3G 1A4, Canada
关键词
4D in vivo dosimetry; electromagnetic positioning system; MOSFET; deformable phantom; RADPOS; 4-DIMENSIONAL COMPUTED-TOMOGRAPHY; VOLUME HISTOGRAM ANALYSIS; IN-VIVO DOSIMETRY; RESPIRATORY MOTION; CANCER PATIENTS; TECHNICAL NOTE; PNEUMONITIS; RADIOTHERAPY; PHOTON; REGISTRATION;
D O I
10.1118/1.3515461
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: A novel 4D in vivo dosimetry system (RADPOS), in conjunction with a deformable lung phantom, has been evaluated as a potential quality assurance tool for 4D radiotherapy. Methods: RADPOS detectors, which consist of a MOSFET dosimeter combined with an electromagnetic positioning probe, were placed inside the deformable lung phantom. One detector was positioned directly inside a tumor embedded in the lung phantom and another was positioned inside the lung portion of the phantom, outside the tumor. CT scans were taken with the phantom at three breathing phases, and for each phase, the detector position inside the phantom was read with the RADPOS software and compared to the position as determined from the CT data. These values were also compared to RADPOS measurements taken with the phantom on the couch of a Varian Clinac 6EX linac. The deformable phantom and the RADPOS system were also used in two radiation delivery scenarios: (1) A simulation of a free-breathing delivery and (2) a simulation of an adaptive treatment. Results: Compared to CT imaging, the RADPOS positional accuracy was found to be better than 2.5 mm. The radial displacement measurements taken in the CT and linac rooms agreed to within an average of (0.7 +/- 0.3) mm. Hence, the system can provide relative displacement measurements in the treatment room, consistent with measurements made in the CT room. For the free-breathing delivery, the total dose reported by RADPOS agreed to within 4% and 5% of the treatment planning doses in the tumor and the lung portion of the phantom, respectively. The RADPOS-measured dose values for the adaptive delivery were within 1.5% of the treatment plan values, which was well within the estimated experimental uncertainties. Conclusions: This work has shown that the deformable lung phantom-RADPOS system can be an efficient quality assurance tool for 4D radiation therapy. (C) 2011 American Association of Physicists in Medicine. [DOI: 10.1118/1.3515461]
引用
收藏
页码:179 / 187
页数:9
相关论文
共 39 条
  • [21] Investigation of 4D dose in volumetric modulated arc therapy-based stereotactic body radiation therapy: does fractional dose or number of arcs matter?
    Shintani, Takashi
    Nakamura, Mitsuhiro
    Matsuo, Yukinori
    Miyabe, Yuki
    Mukumoto, Nobutaka
    Mitsuyoshi, Takamasa
    Iizuka, Yusuke
    Mizowaki, Takashi
    JOURNAL OF RADIATION RESEARCH, 2020, 61 (02) : 325 - 334
  • [22] The relative accuracy of 4D dose accumulation for lung radiotherapy using rigid dose projection versus dose recalculation on every breathing phase
    Valdes, Gilmer
    Lee, Chul
    Tenn, Stephen
    Lee, Percy
    Robinson, Clifford
    Iwamoto, Keisuke
    Low, Daniel
    Lamb, James M.
    MEDICAL PHYSICS, 2017, 44 (03) : 1120 - 1127
  • [23] 4D patient dose reconstruction using online measured EPID cine images for lung SBRT treatment validation
    Lin, Mu-Han
    Li, Jinsheng
    Wang, Lu
    Koren, Sion
    Fan, Jiajing
    Forkal, Eugene
    Ma, C. -M.
    MEDICAL PHYSICS, 2012, 39 (10) : 5949 - 5958
  • [24] Ga-68 MAA Perfusion 4D-PET/CT Scanning Allows for Functional Lung Avoidance Using Conformal Radiation Therapy Planning
    Siva, Shankar
    Devereux, Thomas
    Ball, David L.
    MacManus, Michael P.
    Hardcastle, Nicholas
    Kron, Tomas
    Bressel, Mathias
    Foroudi, Farshad
    Plumridge, Nikki
    Steinfort, Daniel
    Shaw, Mark
    Callahan, Jason
    Hicks, Rodney J.
    Hofman, Michael S.
    TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2016, 15 (01) : 114 - 121
  • [25] 4D dose calculation for pencil beam scanning proton therapy of pancreatic cancer using repeated 4DMRI datasets
    Dolde, Kai
    Naumann, Patrick
    David, Christian
    Gnirs, Regula
    Kachelriess, Marc
    Lomax, Antony John
    Saito, Nami
    Weber, Damien Charles
    Pfaffenberger, Asja
    Zhang, Ye
    PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (16)
  • [26] Predictive value of vascular calcification identified in 4D planning CT of lung cancer patients treated with stereotactic body radiation therapy
    Abravan, Azadeh
    van Herk, Marcel
    Brewster, Frank
    Faivre-Finn, Corinne
    McWilliam, Alan
    Osorio, Eliana M. Vasquez
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2020, 78 : 173 - 178
  • [27] Impact of dose on lung ventilation change calculated from 4D-CT using deformable image registration in lung cancer patients treated with SBRT
    Latifi, Kujtim
    Dilling, Thomas J.
    Feygelman, Vladimir
    Moros, Eduardo G.
    Stevens, Craig W.
    Montilla-Soler, Jaime L.
    Zhang, Geoffrey G.
    JOURNAL OF RADIATION ONCOLOGY, 2015, 4 (03) : 265 - 270
  • [28] The feasibility of the pretreatment verification of 2D dose distributions in radiation therapy with small fields using the electronic portal imaging device
    Tuncel, Nina
    Mahdavi, Seied Rabi
    Haghparast, Mohammad
    Karakus, Ismail
    Haghparast, Abbas
    Nazari, Vahideh
    JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2023, 19 : S815 - S820
  • [29] 4D cone-beam computed tomography (CBCT) using a moving blocker for simultaneous radiation dose reduction and scatter correction
    Zhao, Cong
    Zhong, Yuncheng
    Duan, Xinhui
    Zhang, You
    Huang, Xiaokun
    Wang, Jing
    Jin, Mingwu
    PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (11)
  • [30] Comparing phase- and amplitude-gated volumetric modulated arc therapy for stereotactic body radiation therapy using 3D printed lung phantom
    Lee, Minsik
    Yoon, KyoungJun
    Cho, Byungchul
    Kim, Su Ssan
    Song, Si Yeol
    Choi, Eun Kyung
    Ahn, SeungDo
    Lee, Sang-Wook
    Kwak, JungWon
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2019, 20 (02): : 107 - 113