Tsukuba remote monitoring system for continuous-flow artificial heart

被引:17
作者
Kosaka, R
Sankai, Y
Takiya, R
Jikuya, T
Yamane, T
Tsutsui, T
机构
[1] Univ Tsukuba, Inst Engn Mech & Syst, Sankai Lab Syst & Engn Mech, Tsukuba, Ibaraki 3058573, Japan
[2] Univ Tsukuba, Inst Clin Med, Tsukuba, Ibaraki 3058573, Japan
[3] Natl Inst Adv Ind Sci & Technol, Tsukuba, Ibaraki, Japan
关键词
smart artificial heart; remote monitoring; physiological model; cellular phone;
D O I
10.1046/j.1525-1594.2003.00037.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In order to make long-term medical treatment with the use of an artificial heart effective, we developed the Tsukuba remote monitoring system, which enables medical staff to manage the physiological condition of patients and the driving condition of the artificial heart at anytime from a remote place. This remote monitoring system has three functions: first, a remote monitoring function, which enables medical staff to monitor measured data from a remote place anytime by using not only a personal computer but also a cellular phone; second, an analyzing function, which estimates the unmeasured physiological behavior in the body based on a mathematical physiological model; and third, a warning function, which detects physiological problems and malfunction of the artificial heart by applying the if-then rule and sends a warning message to the medical staff. As a result of applying this system to animal experiments, we have confirmed the effectiveness of the proposed system.
引用
收藏
页码:897 / 906
页数:10
相关论文
共 17 条
[1]  
Abe Y., 1997, Japanese Journal of Artificial Organs, V26, P21
[2]  
HASHIMOTO H, 1993, JPN J ARTIF ORGANS, V22, P724
[3]  
KITAMORI S, 1983, J SICE, V22, P599
[4]  
Kosaka R, 2002, P ANN INT IEEE EMBS, P1571, DOI 10.1109/IEMBS.2002.1106543
[5]   Online parameter identification of second-order systemic circulation model using the delta operator [J].
Kosaka, R ;
Sankai, Y ;
Jikuya, T ;
Yamane, T ;
Tsutsui, T .
ARTIFICIAL ORGANS, 2002, 26 (11) :967-970
[6]   On-line parameter identification of systemic circulation using the delta operator [J].
Kosaka, R ;
Sankai, Y ;
Jikuya, T ;
Yamane, T ;
Tsutsui, T .
ARTIFICIAL ORGANS, 2002, 26 (08) :723-726
[7]   HeartMate left ventricular assist devices: A multigeneration of implanted blood pumps [J].
Maher, TR ;
Butler, KC ;
Poirier, VL ;
Gernes, DB .
ARTIFICIAL ORGANS, 2001, 25 (05) :422-426
[8]  
Mita T., 1988, Journal of the Society of Instrument and Control Engineers, V27, P1058
[9]  
Mussivand T, 1997, ASAIO J, V43, P230
[10]   Terumo implantable left ventricular assist system: Results of long-term animal study [J].
Nojiir, C ;
Kijima, T ;
Maekawa, J ;
Horiuchi, K ;
Kido, T ;
Sugiyama, T ;
Mori, T ;
Sugiura, N ;
Asada, T ;
Ozaki, T ;
Suzuki, M ;
Akamatsu, T ;
Akutsu, T .
ASAIO JOURNAL, 2000, 46 (01) :117-122