Effect of scan pattern on complete-arch scans with 4 digital scanners

被引:123
作者
Latham, Jason [1 ]
Ludlow, Mark [2 ]
Mennito, Anthony [2 ]
Kelly, Abigail [3 ]
Evans, Zachary [4 ]
Renne, Walter [2 ]
机构
[1] Med Univ South Carolina, Coll Dent Med, Dept Orthodont, Charleston, SC 29425 USA
[2] Med Univ South Carolina, Coll Dent Med, Dept Oral Rehabil & Restorat Dent, Charleston, SC 29425 USA
[3] Med Univ South Carolina, Dept Med, Div Biostat & Epidemiol, Charleston, SC 29425 USA
[4] Med Univ South Carolina, Coll Dent Med, Dept Stomatol, Charleston, SC 29425 USA
关键词
IMPLANT IMPRESSIONS; INTRAORAL SCANNERS; ACCURACY; RELIABILITY;
D O I
10.1016/j.prosdent.2019.02.008
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Statement of problem. Complete-arch digital scans are becoming popular as digital dentistry is adopted for expanded clinical situations such as complete-arch prostheses, removable prostheses, extensive implant-supported treatment, and orthodontic aligners. Whether the scan pattern technique affects the trueness and precision of complete-arch scans and whether differences in accuracy exist among different scanners remain unclear. Furthermore, each manufacturer recommends a different scan pattern, but evidence of the superiority of the manufacturer's recommended pattern is lacking. Purpose. The purpose of this in vitro study was to determine whether the scan pattern affects the trueness, precision, and speed of complete-arch digital scans performed by using 4 different digital scanning systems. Material and methods. A custom model used as the reference standard was fabricated with teeth having the same refractive index as dentin and enamel to simulate the natural dentition. The scan of the custom typodont was obtained by using an ATOS III Triple Scan 3D optical scanner. This study evaluated the CEREC Omnicam, Planmeca Emerald, Align iTero Element, and 3Shape TRIOS 3. Experimental scans were obtained from each of the 4 different digital scanning systems by using 4 unique scan patterns by experienced clinicians. Four experimental scans were acquired from each of the scanners by using 4 distinct scan patterns for a total of 16 scans for each scanner. Scan patterns 1 to 4 were based on the operator manuals for each different scanner. The scan time was recorded for each scan. All experimental scans were converted to standard tessellation language (STL) format, and a comprehensive metrology program, Geomagic Control X, was used to compare the reference standard scan with the experimental scans. Results. For trueness, the scanner (P<.001), scan pattern (P=.001), and their interaction (P<.001) were found to be significant. Overall, scan pattern 2 showed the highest average trueness and precision. Likewise, for overall scan pattern precision, the scanner, scan pattern, and their interaction were found to be significant (P<.001). Conclusions. Scan pattern affected trueness and precision for some scanners, but not for others. Differences exist in the complete-arch scan speed, trueness, and precision of individual scanners. Scan pattern can play an important role in the success of digital scanning.
引用
收藏
页码:85 / 95
页数:11
相关论文
共 28 条
[21]   Evaluation of the effect scan pattern has on the trueness and precision of six intraoral digital impression systems [J].
Mennito, Anthony S. ;
Evans, Zachary P. ;
Lauer, Abigail W. ;
Patel, Ravi B. ;
Ludlow, Mark E. ;
Renne, Walter G. .
JOURNAL OF ESTHETIC AND RESTORATIVE DENTISTRY, 2018, 30 (02) :113-118
[22]   Impact of digital intraoral scan strategies on the impression accuracy using the TRIOS Pod scanner [J].
Mueller, Philipp ;
Ender, Andreas ;
Joda, Tim ;
Katsoulis, Joannis .
QUINTESSENCE INTERNATIONAL, 2016, 47 (04) :343-349
[23]   Precision and Accuracy of a Digital Impression Scanner in Full-Arch Implant Rehabilitation [J].
Pesce, Paolo ;
Pera, Francesco ;
Setti, Paolo ;
Menini, Maria .
INTERNATIONAL JOURNAL OF PROSTHODONTICS, 2018, 31 (02) :171-175
[24]  
RENNE W, 2017, DENT, V118, P36
[25]  
Rutkunas V, 2017, EUR J ORAL IMPLANTOL, V10, P101
[26]   Comparing the accuracy (trueness and precision) of models of fixed dental prostheses fabricated by digital and conventional workflows [J].
Sim, Ji-Young ;
Jang, Yeon ;
Kim, Woong-Chul ;
Kim, Hae-Young ;
Lee, Dong-Hwan ;
Kim, Ji-Hwan .
JOURNAL OF PROSTHODONTIC RESEARCH, 2019, 63 (01) :25-30
[27]   Use of digital impression systems with intraoral scanners for fabricating restorations and fixed dental prostheses [J].
Takeuchi, Yoshimasa ;
Koizumi, Hiroyasu ;
Furuchi, Mika ;
Sato, Yohei ;
Ohkubo, Chikahiro ;
Matsumura, Hideo .
JOURNAL OF ORAL SCIENCE, 2018, 60 (01) :1-7
[28]  
Wimmer T, 2017, J PROSTHET DENT, V5, P36