Determination of trace beryllium in geological samples by oxygen doped air-acetylene flame atomic absorption spectrometry

被引:0
作者
Wu, SW [1 ]
Yang, XQ
Chen, WW
Yang, M
Tang, ZY
Jin, ZX
机构
[1] China Univ Geosci, Fac Chem Engn & Mat Sci, Wuhan 430074, Peoples R China
[2] Hubei Inst Nationalitites, Fac Chem & Environm Engn, Enshi 445000, Peoples R China
关键词
oxygen enriched air-acetylene flame; AAS; geological samples; beryllium;
D O I
暂无
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
A new method for the determination of trace beryllium in geological samples by oxygen-air-acetylene flame atomic absorption spectrometry is reported in the present paper. The optimum determination conditions were selected. When suitable amounts of 8-hydroxyquinoline are added, the interference effect from the coexisting elements can be minimized and the sensitivity for the determination of beryllium can be increased. The detection limit(k = 3) is 0.006 mu g center dot mL(-1) for beryllium. The method has been applied to the determination of trace beryllium in geological standard samples, and the results are in good agreement with the certified values with a precision of 4.69% RSD(n = 8).
引用
收藏
页码:785 / 787
页数:3
相关论文
共 4 条
[1]  
CHEN JH, 1998, CHINESE J SPECTRAL E, V18, P16
[2]  
LI LZ, 1991, ROCK MINERAL ANAL, P613
[3]  
Ou H, 2002, SPECTROSC SPECT ANAL, V22, P146
[4]  
WELZ B, 1989, ATOMIC ABSORPTION SP, P314