Effect of secondary particles on image quality of dynamic flat panels in carbon ion scanning beam treatment

被引:4
作者
Mori, S. [1 ]
Amano, S. [2 ]
Furukawa, T. [1 ]
Shirai, T. [1 ]
Noda, K. [1 ]
机构
[1] Natl Inst Radiol Sci, Res Ctr Charged Particle Therapy, Med Phys Res Program, Chiba 260, Japan
[2] Chiba Univ, Res Ctr Frontier Med Engn, Chiba, Japan
关键词
MOTION; RADIOTHERAPY; LUNG; TRACKING; THERAPY; PROTON; PATTERN;
D O I
10.1259/bjr.20140567
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: Real-time markerless tumour tracking using radiographic fluoroscopic imaging is one of the better solutions to improving respiratory-gated radiotherapy. However, particle beams cause secondary particles from patients, which could affect radiographs. Here, we evaluated the quality of radiographs during carbon ion pencil beam scanning (CPBS) irradiation for respiratory gating. Methods: A water phantom and chest phantom were used. The phantoms were irradiated with CPBS at 290 MeV n(-1) from orthogonal directions. Dose rates were 3.4x10(8), 1.14x10(8) and 3.79x10(7) particles per second. A dynamic flat panel detector (DFPD) was installed on the upstream (DFPD1) or downstream (DFPD2) side of the vertical irradiation port. DFPD images were acquired during CPBS at 15.00, 7.50 and 3.75 frames per second (fps). Charge on the DFPD was cleaned using fast readout technique every 30 fps. DFPD images were acquired during CPBS with radiographic exposure, and results with and without fast readout technique were compared. Results: Secondary particles were visualized as spots or streak-like shapes. Capture of secondary particles from the horizontal beam direction was lower with fast readout technique than without it. With regard to beam irradiation direction dependency, CPBS from the horizontal direction resulted in a greater magnitude of secondary particles reaching DFPD2 than reaching DFPD1. When CPBS was delivered from the vertical direction, however, the magnitude of secondary particles on both DFPDs was very similar. Conclusion: Fast readout technique minimized the effect of secondary particles on DFPD images during CPBS. Advances in knowledge: This technique may be useful for markerless tumour tracking for respiratory gating.
引用
收藏
页数:8
相关论文
共 23 条
[1]   Breathing pattern in humans: diversity and individuality [J].
Benchetrit, G .
RESPIRATION PHYSIOLOGY, 2000, 122 (2-3) :123-129
[2]   Towards fluoroscopic respiratory gating for lung tumours without radiopaque markers [J].
Berbeco, RI ;
Mostafavi, H ;
Sharp, GC ;
Jiang, SB .
PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (19) :4481-4490
[3]   Motion in radiotherapy: particle therapy [J].
Bert, C. ;
Durante, M. .
PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (16) :R113-R114
[4]   Comparison of carbon ion radiotherapy to photon radiation alone or in combination with temozolomide in patients with high-grade gliomas: Explorative hypothesis-generating retrospective analysis [J].
Combs, Stephanie E. ;
Bruckner, Thomas ;
Mizoe, Jun-Etso ;
Kamada, Tadashi ;
Tsujii, Hirohiko ;
Kieser, Meinhard ;
Debus, Juergen .
RADIOTHERAPY AND ONCOLOGY, 2013, 108 (01) :132-135
[5]   Multiple template-based fluoroscopic tracking of lung tumor mass without implanted fiducial markers [J].
Cui, Ying ;
Dy, Jennifer G. ;
Sharp, Gregory C. ;
Alexander, Brian ;
Jiang, Steve B. .
PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (20) :6229-6242
[6]   Performance of the NIRS fast scanning system for heavy-ion radiotherapy [J].
Furukawa, Takuji ;
Inaniwa, Taku ;
Sato, Shinji ;
Shirai, Toshiyuki ;
Takei, Yuka ;
Takeshita, Eri ;
Mizushima, Kota ;
Iwata, Yoshiyuki ;
Himukai, Takeshi ;
Mori, Shinichiro ;
Fukuda, Shigekazu ;
Minohara, Shinichi ;
Takada, Eiichi ;
Murakami, Takeshi ;
Noda, Koji .
MEDICAL PHYSICS, 2010, 37 (11) :5672-5682
[7]   Motion mitigation in intensity modulated particle therapy by internal target volumes covering range changes [J].
Graeff, Christian ;
Durante, Marco ;
Bert, Christoph .
MEDICAL PHYSICS, 2012, 39 (10) :6004-6013
[8]   Hypofractionated carbon ion therapy delivered with scanned ion beams for patients with hepatocellular carcinoma - feasibility and clinical response [J].
Habermehl, Daniel ;
Debus, Juergen ;
Ganten, Tom ;
Ganten, Maria-Katharina ;
Bauer, Julia ;
Brecht, Ingo C. ;
Brons, Stephan ;
Haberer, Thomas ;
Haertig, Martin ;
Jaekel, Oliver ;
Parodi, Katia ;
Welzel, Thomas ;
Combs, Stephanie E. .
RADIATION ONCOLOGY, 2013, 8
[9]  
HUGELIN A, 1989, RESPIRATORY CONTROL : A MODELING PERSPECTIVE, P353
[10]  
International Electrotechnical Commission, 2008, RADIOTHERAPY EQUIPME, V61217