Features of High-Temperature Calibration of W/Re Thermocouples

被引:6
作者
Ulanovskiy, A. [1 ]
Edler, F. [2 ]
Fischer, J. [2 ]
Oleynikov, P. [3 ]
Zaytsev, P. [3 ]
Pokhodun, A. [4 ]
机构
[1] Obninsk Thermoelect Co OTC Ltd, Obninsk 249033, Russia
[2] Phys Tech Bundesanstalt, D-10587 Berlin, Germany
[3] Sci Ind Assoc SIA LUCH, Podolsk 142100, Russia
[4] D I Mendeleyev Inst Metrol VNIIM, St Petersburg 190005, Russia
关键词
High temperatures; IEC; 60584; Thermoelectric properties; W/Re thermocouple;
D O I
10.1007/s10765-014-1780-4
中图分类号
O414.1 [热力学];
学科分类号
摘要
Investigations of Type A (W-5 %Re/W-20 %Re) thermocouples were performed at several laboratories to validate their reference function before its standardization in the new issue of the international standard IEC 60584. The Type A thermocouples investigated were equipped with sealed protection tubes made of sapphire which were filled with an inert gas (argon). The investigations at Russian laboratories were performed mainly in carbon-free high-temperature furnaces. The calibration results obtained in the temperature range (600 to 1850) in the carbon-free environment were within % tolerance limits and confirmed the suitability of Type A thermocouples for industrial applications. In contrast, the Type A thermocouple 89/95-P investigated at PTB (Germany) was exposed to a carbon environment when annealed at and when it was calibrated at metal-carbon eutectic (Me-C) fixed points. Measurements made at Me-C fixed points did not deviate from the reference function by more than about 6 K at the first stage when temperatures were below . However, the inhomogeneity of the thermoelements increased continuously after the calibration at the Me-C eutectic fixed points. The additional measurements at the peritectic fixed point () resulted in a continuous emf drift to deviations from the reference function of about (100 to 150) which corresponds to a temperature equivalent of about (9 to 14) K. The thermoelectric stability and homogeneity of the thermocouple 89/95-P during these investigations was checked by repeated measurements at the freezing point of copper ().
引用
收藏
页码:433 / 443
页数:11
相关论文
共 13 条
[1]  
[Anonymous], 2002, 85852001 GOSTR
[2]  
[Anonymous], 2003, ASTME230E230M12
[3]   Construction and investigation of Pt/Pd thermocouples in the framework of the EUROMET Project 857 [J].
Edler, F. ;
Morice, R. ;
Pearce, J. .
INTERNATIONAL JOURNAL OF THERMOPHYSICS, 2008, 29 (01) :199-209
[4]   A cobalt-carbon eutectic fixed point for the calibration of contact thermometers at temperatures above 1100°C [J].
Edler, F ;
Baratto, AC .
METROLOGIA, 2005, 42 (04) :201-207
[5]  
Hall B.F., 1962, J INSTRUM CONTROL SY, V35
[6]  
Intemational Electrotechnical Comission, 2013, IEC6058412013
[7]   THERMOCOUPLES FOR USE IN CARBON ATMOSPHERES [J].
NADLER, MR ;
KEMPTER, CP .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1961, 32 (01) :43-&
[8]   C-W (Carbon-Tungsten) [J].
Okamoto, H. .
JOURNAL OF PHASE EQUILIBRIA AND DIFFUSION, 2008, 29 (06) :543-544
[9]   Self-Calibration of a W/Re Thermocouple Using a Miniature Ru-C (1954 °C) Eutectic Cell [J].
Ongrai, O. ;
Pearce, J. V. ;
Machin, G. ;
Sweeney, S. J. .
TEMPERATURE: ITS MEASUREMENT AND CONTROL IN SCIENCE AND INDUSTRY, VOL 8, 2013, 1552 :504-509
[10]   Investigation of Pd-C cells to improve thermocouple calibration [J].
Pearce, J. V. ;
Edler, F. ;
Elliott, C. J. ;
Failleau, G. ;
Morice, R. ;
Ogura, H. .
METROLOGIA, 2011, 48 (05) :375-381