Fluorescence Performance Standards for Confocal Microscopy

被引:0
作者
Ruettinger, Steffen [1 ]
Kapusta, Peter [2 ]
Voellkopf, Volker [2 ]
Koberling, Felix [2 ]
Erdmann, Rainer [2 ]
Macdonald, Rainer [1 ]
机构
[1] Phys Tech Bundesanstalt, Abbe Str 2-12, D-10587 Berlin, Germany
[2] Picoquant GmbH, D-12489 Berlin, Germany
来源
MULTIPHOTON MICROSCOPY IN THE BIOMEDICAL SCIENCES X | 2010年 / 7569卷
关键词
confocal microscopy; reference sample; performance standard; single molecule detection; FCS; FLIM;
D O I
10.1117/12.840501
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
State of the art confocal microscopes offer diffraction limited (or even better) spatial resolution, highest (single molecule) sensitivity and ps-fluorescence lifetime measurement accuracy. For developers, manufacturers, as well as users of confocal microscopes it is mandatory to assign values to these qualities. In particular for users, it is often not easy to ascertain that the instrument is properly aligned as a large number of factors influence resolution or sensitivity. Therefore, we aspire to design a set of performance standards to be deployed on a day-to-day fashion in order to check the instruments characteristics. The main quantities such performance standard must address are: Spatial resolution Sensitivity Fluorescence lifetime To facilitate the deployment and thus promote wide range adoption in day-to-day performance testing the corresponding standards have to be ready made, easy to handle and to store. The measurement procedures necessary should be available on as many different setups as possible and the procedures involved in their deployment should be as easy as possible. To this end, we developed two performance standards to accomplish the mentioned goals: Resolution reference Combined molecular brightness and fluorescence lifetime reference The first one is based on sub-resolution sized Tetra-Speck (TM) fluorescent beads or alternatively on single molecules on a glass surface to image and to determine quantitatively the confocal volume, while the latter is a liquid sample containing fluorescent dyes of different concentrations and spectral properties. Both samples are sealed in order to ease their use and prolong their storage life. Currently long-term tests are performed to ascertain durability and road capabilities.
引用
收藏
页数:10
相关论文
共 5 条
[1]   Two-focus fluorescence correlation spectroscopy: A new tool for accurate and absolute diffusion measurements [J].
Dertinger, Thomas ;
Pacheco, Victor ;
von der Hocht, Iris ;
Hartmann, Rudolf ;
Gregor, Ingo ;
Enderlein, Joerg .
CHEMPHYSCHEM, 2007, 8 (03) :433-443
[2]   Recent advances in time-correlated single-photon counting [J].
Koberling, Felix ;
Kraemer, Benedikt ;
Tannert, Sebastian ;
Ruettinger, Steffen ;
Ortmann, Uwe ;
Patting, Matthias ;
Wahl, Michael ;
Ewers, Benjamin ;
Kapusta, Peter ;
Erdmann, Rainer .
SINGLE MOLECULE SPECTROSCOPY AND IMAGING, 2008, 6862
[3]   Time-resolved confocal fluorescence imaging and spectrocopy system with single molecule sensitivity and sub-micrometer resolution [J].
Wahl, M ;
Koberling, F ;
Patting, M ;
Rahn, H ;
Erdmann, R .
CURRENT PHARMACEUTICAL BIOTECHNOLOGY, 2004, 5 (03) :299-308
[4]   Dead-time optimized time-correlated photon counting instrument with synchronized, independent timing channels [J].
Wahl, Michael ;
Rahn, Hans-Juergen ;
Gregor, Ingo ;
Erdmann, Rainer ;
Enderlein, Joerg .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2007, 78 (03)
[5]   Scalable time-correlated photon counting system with multiple independent input channels [J].
Wahl, Michael ;
Rahn, Hans-Juergen ;
Roehlicke, Tino ;
Kell, Gerald ;
Nettels, Daniel ;
Hillger, Frank ;
Schuler, Ben ;
Erdmann, Rainer .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2008, 79 (12)