Effects of titanium contamination caused by iron-free high-performance liquid chromatography systems on peak shape and retention of drugs with chelating properties

被引:26
作者
De Pra, Mauro [1 ]
Greco, Giorgia [1 ]
Krajewski, Matthias P. [1 ]
Martin, Markus M. [1 ]
George, Ed [2 ]
Bartsch, Nora [3 ]
Steiner, Frank [1 ]
机构
[1] Thermo Fisher Sci, Dornierstr 4, D-82210 Germering, Germany
[2] Thermo Fisher Sci, 355 River Oaks Pkwy, San Jose, CA 95134 USA
[3] Thermo Fisher Sci, Hanna Kunath Str 11, D-28199 Bremen, Germany
关键词
Metal contamination; Biocompatible HPLC; Fluoroquinolones; Peak tailing; System corrosion; MASS-TRANSFER KINETICS; STRESS-CORROSION; PASSIVE FILM; TI; ANTIBIOTICS; COMPLEXES; BEHAVIOR; MODEL;
D O I
10.1016/j.chroma.2019.460619
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Iron-free HPLC systems, better known as biocompatible systems, are generally regarded to be chemically more inert compared to conventional HPLC systems. In this work, we studied the chromatographic behavior of some classes of compounds of pharmaceutical interest, analyzed with iron-free systems. Issues typically associated with metal contamination, i.e. strong peak tailing, were observed when using an amide polar-embedded column. Effects of the contamination were visible when anhydrous methanolacetonitrile was used, indicating that this solvent, albeit generally considered safe for conventional HPLC systems, induce corrosion of iron-free systems. The confirmation of titanium as main acting contaminant came from systematically studying the contribution of each wetted component of the HPLC system on peak shape of affected molecules. Quantification of titanium by ICP-MS analysis of effluents provided further evidence on the source of contamination. A mechanistic description of the complex interaction between titanium ions, organic molecules, and column stationary phase is proposed. In the perspective of developing methods that are fully portable between stainless steel and titanium systems, recommendations are given in terms of potentially sensitive molecules, suitable mobile phase conditions, and type of column to be used. (C) 2019 Elsevier B.V. All rights reserved.
引用
收藏
页数:10
相关论文
共 32 条
[1]  
[Anonymous], 1967, D3868 NASA TN
[2]  
Bosch M.R. Elisabeth, 2002, ANAL CHEM, V69, P1970, DOI [10.1021/ac961126a., DOI 10.1021/AC961126A]
[3]   Understanding Mixed-Mode Retention Mechanisms in Liquid Chromatography with Hydrophobic Stationary Phases [J].
Cavazzini, Alberto ;
Marchetti, Nicola ;
Guzzinati, Roberta ;
Pasti, Luisa ;
Ciogli, Alessia ;
Gasparrini, Francesco ;
Lagana, Aldo .
ANALYTICAL CHEMISTRY, 2014, 86 (10) :4919-4926
[4]   Spectrofluorimetric study of the acid-base equilibria and complexation behavior of the fluoroquinolone antibiotics ofloxacin, norfloxacin, ciprofloxacin and pefloxacin in aqueous solution [J].
Drakopoulos, AI ;
Ioannou, PC .
ANALYTICA CHIMICA ACTA, 1997, 354 (1-3) :197-204
[5]   SIMPLIFIED DESCRIPTION OF HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC SEPARATION UNDER OVERLOAD CONDITIONS, BASED ON THE CRAIG DISTRIBUTION MODEL .2. EFFECT OF ISOTHERM TYPE, AND EXPERIMENTAL-VERIFICATION OF COMPUTER-SIMULATIONS FOR A SINGLE BAND [J].
EBLE, JE ;
GROB, RL ;
ANTLE, PE ;
SNYDER, LR .
JOURNAL OF CHROMATOGRAPHY, 1987, 384 :45-79
[6]   THE STRESS-CORROSION AND PRE-EXPOSURE EMBRITTLEMENT OF TITANIUM IN METHANOLIC SOLUTIONS OF HYDROCHLORIC-ACID [J].
EBTEHAJ, K ;
HARDIE, D ;
PARKINS, RN .
CORROSION SCIENCE, 1985, 25 (06) :415-429
[7]   Chromatographic determination of metallic impurities in reversed-phase HPLC columns [J].
Engelhardt, H ;
Lobert, T .
ANALYTICAL CHEMISTRY, 1999, 71 (09) :1885-1892
[8]   COMPARISON AND CHARACTERIZATION OF REVERSED PHASES [J].
ENGELHARDT, H ;
JUNGHEIM, M .
CHROMATOGRAPHIA, 1990, 29 (1-2) :59-68
[9]  
Fehrenbach T., 2017, HPLC EXPERT, P223, DOI [10.1002/9783527694945.ch7, DOI 10.1002/9783527694945.CH7]
[10]   Peak tailing and mass transfer kinetics in linear chromatography [J].
Fornstedt, T ;
Zhong, GM ;
Guiochon, G .
JOURNAL OF CHROMATOGRAPHY A, 1996, 741 (01) :1-12