Hemodynamic performance of small-size bileaflet valves: Pressure drop and laser Doppler anemometry study comparison of three prostheses

被引:10
作者
Grigioni, M [1 ]
Daniele, C [1 ]
D'Avenio, G [1 ]
Barbaro, V [1 ]
机构
[1] Ist Super Sanita, Biomed Engn Lab, I-00161 Rome, Italy
关键词
prosthetic heart valves; laser Doppler anemometry; turbulence; energy dissipation; in vitro testing;
D O I
10.1046/j.1525-1594.2000.06614.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Laser Doppler anemometry (LDA) is a single-point technique which is unparalleled to detect accurately the local properties of the velocity field in a turbulent flow, such as that generated by a prosthetic heart valve (PHV). We propose a correlation between the structure of the flow field in three 19 mm bileaflet PHVs (Sorin Bicarbon, St. Jude Standard, St. Jude HP), investigated at peak systole (6 L/min cardiac output [CO]) with LDA, in kinematic and geometric similarity, and the global parameter of transvalvular pressure drop measured in both steady and pulsatile conditions. The pressure transducers of the same apparatus were used to characterize pressure drops at different flow rates whereas the steady-flow case was studied with a highly accurate tester built in our laboratory. The 2 St. Jude models rank according to their internal orifice diameter (ID) with the standard model (with a smaller IT)) providing higher pressure drops for each flow rate. Sorin Bicarbon, due to its leaflet geometry, generates a more complex flow field with respect to the 2 St. Jude flat-leaflet models and shows improved hemodynamical behavior in pulsatile conditions with respect to the stationary case due to differences in pressure recovery. This study can provide insights into a PHV's local flow structure and global hemodynamical parameters.
引用
收藏
页码:959 / 965
页数:7
相关论文
共 23 条
[1]  
Akutsu T, 1997, ARTIF ORGANS, V21, P1110
[2]   DETERMINATION OF PRINCIPAL REYNOLDS STRESSES IN PULSATILE FLOWS AFTER ELLIPTIC FILTERING OF DISCRETE VELOCITY-MEASUREMENTS [J].
BALDWIN, JT ;
DEUTSCH, S ;
PETRIE, HL ;
TARBELL, JM .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1993, 115 (04) :396-403
[3]  
Barbaro V, 1998, Technol Health Care, V6, P259
[4]   19 mm sized bileaflet valve prostheses' flow field investigated by bidimensional laser Doppler anemometry (part II: maximum turbulent shear stresses) [J].
Barbaro, V ;
Grigioni, M ;
Daniele, C ;
D'Avenio, G ;
Boccanera, G .
INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 1997, 20 (11) :629-636
[5]  
Barbaro V, 1997, J HEART VALVE DIS, V6, P93
[6]  
Black M M, 1995, J Heart Valve Dis, V4 Suppl 1, pS1
[7]  
*CEN EN, 1998, 120061 CEN EN
[8]   Inaccurate and misleading valve sizing: A proposed standard for valve size nomenclature [J].
Christakis, GT ;
Buth, KJ ;
Goldman, BS ;
Fremes, SE ;
Rao, V ;
Cohen, G ;
Borger, MA ;
Weisel, RD .
ANNALS OF THORACIC SURGERY, 1998, 66 (04) :1198-1203
[9]   Comparative study of the hydrodynamic function of six size 19 mm bileaflet heart valves [J].
Fisher, J .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 1995, 9 (12) :692-695
[10]  
*FOOD DRUG ADM, 1985, INT COMP PROSTH HEAR