摘要
The effect of time-of-flight (TOF) and point spread function (PSF) modeling in image reconstruction has not been well studied for cardiac PET. This study assesses their separate and combined influence on Rb-82 myocardial perfusion imaging in obese patients. Thirty-six obese patients underwent rest-stress Rb-82 cardiac PET. Images were reconstructed with and without TOF and PSF modeling. Perfusion was quantitatively compared using the AHA 17-segment model for patients grouped by BMI, cross-sectional body area in the scanner field of view, gender, and left ventricular myocardial volume. Summed rest scores (SRS), summed stress scores (SSS), and summed difference scores (SDS) were compared. TOF improved polar map visual uniformity and increased septal wall perfusion by up to 10%. This increase was greater for larger patients, more evident for patients grouped by cross-sectional area than by BMI, and more prominent for females. PSF modeling increased perfusion by about 1.5% in all cardiac segments. TOF modeling generally decreased SRS and SSS with significant decreases between 2.4 and 3.0 (P < .05), which could affect risk stratification; SDS remained about the same. With PSF modeling, SRS, SSS, and SDS were largely unchanged. TOF and PSF modeling affect regional and global perfusion, SRS, and SSS. Clinicians should consider these effects and gender-dependent differences when interpreting Rb-82 perfusion studies.
引用
收藏
页码:1521 / 1545
页数:25
相关论文
共 49 条
  • [1] The effect of time-of-flight and point spread function modeling on 82Rb myocardial perfusion imaging of obese patients
    Paul K. R. Dasari
    Judson P. Jones
    Michael E. Casey
    Yuanyuan Liang
    Vasken Dilsizian
    Mark F. Smith
    Journal of Nuclear Cardiology, 2018, 25 : 1521 - 1545
  • [2] Quantification of myocardial blood flow with 82Rb: Validation with 15O-water using time-of-flight and point-spread-function modeling
    Germino, Mary
    Ropchan, Jim
    Mulnix, Tim
    Fontaine, Kathryn
    Nabulsi, Nabeel
    Ackah, Eric
    Feringa, Herman
    Sinusas, Albert J.
    Liu, Chi
    Carson, Richard E.
    EJNMMI RESEARCH, 2016, 6
  • [3] Effect of time-of-flight and point spread function modeling on detectability of myocardial defects in PET
    Schaefferkoetter, Joshua
    Ouyang, Jinsong
    Rakvongthai, Yothin
    Nappi, Carmela
    El Fakhri, Georges
    MEDICAL PHYSICS, 2014, 41 (06)
  • [4] 82Rb PET/CT: entering a new area of myocardial perfusion imaging?
    Valenta, Ines
    Schindler, Thomas Hellmut
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2012, 39 (08) : 1231 - 1232
  • [5] Effect of iterations and time of flight on normal distributions of 82Rb PET relative perfusion and myocardial blood flow
    Poitrasson-Riviere, Alexis
    Moody, Jonathan B.
    Renaud, Jennifer M.
    Hagio, Tomoe
    Arida-Moody, Liliana
    Murthy, Venkatesh L.
    Ficaro, Edward P.
    JOURNAL OF NUCLEAR CARDIOLOGY, 2022, 29 (05) : 2612 - 2623
  • [6] Quantitative 82Rb dynamic pet perfusion analysis with kinetic modeling for myocardial viability: Can we get away with just 82Rb perfusion kinetics?
    Ananthasubramaniam, Karthik
    Arumugam, Parthiban
    JOURNAL OF NUCLEAR CARDIOLOGY, 2019, 26 (02) : 387 - 390
  • [7] Impact of point spread function modeling and time-of-flight on myocardial blood flow and myocardial flow reserve measurements for rubidium-82 cardiac PET
    Armstrong, Ian S.
    Tonge, Christine M.
    Arumugam, Parthiban
    JOURNAL OF NUCLEAR CARDIOLOGY, 2014, 21 (03) : 467 - 474
  • [8] Detection and severity classification of extracardiac interference in 82Rb PET myocardial perfusion imaging
    Orton, Elizabeth J.
    Al Harbi, Ibraheem
    Klein, Ran
    Beanlands, Rob S. B.
    deKemp, Robert A.
    Wells, R. Glenn
    MEDICAL PHYSICS, 2014, 41 (10)
  • [9] Simulation of Low-Dose Protocols for Myocardial Perfusion 82Rb Imaging
    Lassen, Martin Lyngby
    Otaki, Yuka
    Kavanagh, Paul
    Miller, Robert J. H.
    Berman, Daniel S.
    Slomka, Piotr J.
    JOURNAL OF NUCLEAR MEDICINE, 2021, 62 (08) : 1112 - 1117
  • [10] The Importance of Time-of-Flight Reconstruction and Point Spread Modeling in the Measurement of Myocardial Blood Flow Parameters
    Case, James A.
    CURRENT CARDIOLOGY REPORTS, 2021, 23 (07)