Neutron diffraction residual strain measurements in nanostructured hydroxyapatite coatings for orthopaedic implants

被引:26
作者
Ahmed, R. [1 ,2 ]
Faisal, N. H. [1 ,2 ]
Paradowska, A. M. [3 ]
Fitzpatrick, M. E. [4 ]
Khor, K. A. [5 ]
机构
[1] Heriot Watt Univ, Sch Engn & Phys Sci, Edinburgh EH14 4AS, Midlothian, Scotland
[2] Alfaisal Univ, Coll Engn, Riyadh 11533, Saudi Arabia
[3] Rutherford Appleton Lab, ISIS, Didcot OX11 0QX, Oxon, England
[4] Open Univ, Milton Keynes MK7 6AA, Bucks, England
[5] Nanyang Technol Univ, Sch Mech & Aerosp Engn, Singapore 639798, Singapore
关键词
Hydroxyapatite coatings; Residual strain; Neutron diffraction; Plasma spray; Orthopaedic implants; SPRAY CERMET COATINGS; MECHANICAL-PROPERTIES; PLASMA; STRESS; ORIENTATION; SUBSTRATE; STRENGTH; TITANIUM; MODULUS;
D O I
10.1016/j.jmbbm.2011.07.003
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The failure of an orthopaedic implant can be initiated by residual strain inherent to the hydroxyapatite coating (HAC). Knowledge of the through-thickness residual strain profile in the thermally sprayed hydroxyapatite coating/substrate system is therefore important in the development of a new generation of orthopaedic implants. As the coating microstructure is complex, non-destructive characterization of residual strain, e.g. using neutron diffraction, provides a useful measure of through thickness strain profile without altering the stress field. This first detailed study using a neutron diffraction technique, non-destructively evaluates the through thickness strain measurement in nanostructured hydroxyapatite plasma sprayed coatings on a titanium alloy substrate (as-sprayed, heat treated, and heat treated then soaked in simulated body fluid (SBF)). The influence of crystallographic plane orientation on the residual strain measurement is shown to indicate texturing in the coating. This texturing is expected to influence both the biological and fracture response of HA coatings. Results are discussed in terms of the influence of heat-treatment and SBF on the residual stress profile for these biomedical coatings. The results show that the through thickness residual strain in all three coatings was different for different crystallographic planes but was on average tensile. It is also concluded that the heat-treatment and simulated body fluid exposure had a significant effect on the residual strain profile in the top layers of HAC. (C) 2011 Elsevier Ltd. All rights reserved.
引用
收藏
页码:2043 / 2054
页数:12
相关论文
共 45 条
[1]   Residual strain measurements in thermal spray cermet coatings via neutron diffraction [J].
Ahmed, R. ;
Yu, H. ;
Stewart, S. .
JOURNAL OF TRIBOLOGY-TRANSACTIONS OF THE ASME, 2007, 129 (02) :411-418
[2]   Neutron Diffraction Residual Strain Measurements in Plasma Sprayed Nanostructured Hydroxyapatite Coatings for Orthopaedic Implants [J].
Ahmed, R. ;
Faisal, N. H. ;
Knupfer, S. M. ;
Paradowska, A. M. ;
Fitzpatrick, M. E. ;
Khor, K. A. ;
Cizek, J. .
MECHANICAL STRESS EVALUATION BY NEUTRONS AND SYNCHROTRON RADIATION, 2010, 652 :309-+
[3]   Neutron diffraction residual strain measurements in post-treated thermal spray cermet coatings [J].
Ahmed, R. ;
Yu, H. ;
Stoica, V. ;
Edwards, L. ;
Santisteban, J. R. .
MATERIALS SCIENCE AND ENGINEERING A-STRUCTURAL MATERIALS PROPERTIES MICROSTRUCTURE AND PROCESSING, 2008, 498 (1-2) :191-202
[4]  
Boivin Georges., 2007, MEDICOGRAPHIA, V29, P126
[5]  
CHE L, 2007, RESIDUAL STRESS STAT, P150
[6]  
CHRISTOPHER MR, 1995, BIOMATERIALS, V16, P691
[7]  
CHUNG YC, 2005, SURFACE COATINGS TEC, V190, P122
[8]   BIOMINERALIZATION A crystal-clear view [J].
Coelfen, Helmut .
NATURE MATERIALS, 2010, 9 (12) :960-961
[9]   Use of Rietveld refinement for elastic macrostrain determination and for evaluation of plastic strain history from diffraction spectra [J].
Daymond, MR ;
Bourke, MAM ;
VonDreele, RB ;
Clausen, B ;
Lorentzen, T .
JOURNAL OF APPLIED PHYSICS, 1997, 82 (04) :1554-1562
[10]  
EDWARDS L, 2003, ANAL RESIDUAL STRESS, V233