The Bi-Surface total knee arthroplasty: Minimum 10-year follow-up study

被引:39
作者
Nakamura, Shinichiro [1 ]
Kobayashi, Masahiko [1 ]
Ito, Hiromu [1 ]
Nakamura, Kenji [2 ]
Ueo, Toyoji [2 ]
Nakamura, Takashi [1 ]
机构
[1] Kyoto Univ, Dept Orthoped Surg, Grad Sch Med, Sakyo Ku, Kyoto 6068507, Japan
[2] Tamatsukuri Koseinenkin Hosp, Dept Orthoped Surg, Tamatsukuri, Ibaraki, Japan
关键词
Total knee arthroplasty; Long-term follow-up; Ceramic femoral component; Deep flexion design; TOTAL CONDYLAR; TIBIAL POST; REPLACEMENT; RANGE; PROSTHESIS; MOTION; DESIGN; FLEXION; SYSTEM; SURVIVORSHIP;
D O I
10.1016/j.knee.2010.02.015
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
The Bi-Surface Knee System (Japan Medical Material, Kyoto, Japan), which has a unique ball-and-socket joint and whose femoral component is made from alumina ceramic, was designed to improve deep knee flexion and long-term durability after total knee arthroplasty. The purpose of this study was to review the clinical results of a minimum 10-year follow-up. Between 1989 and 1997, 507 total knee arthroplasties were carried out in 371 patients. Forty three patients (56 knees) were lost to follow-up. The mean age of the patients at operation was 68.5 years, and the patients were followed up for a mean of 11.7 years. The knees were evaluated on the basis of Knee Society knee score and functional score, radiographs, and Kaplan-Meier survivorship analysis. The knee score was improved from 38.9 +/- 17.4 points preoperatively to 93.3 +/- 7.8 points at the latest follow-up (p<0.001). The functional score was improved from 34.9 +/- 19.3 points to 52.7 +/- 24.1 points (p<0.001). The mean range of flexion was improved from 118.7 +/- 21.7 degrees to 124.2 +/- 20.8 degrees (p<0.001). The critical angle, which means the border to gain more range of flexion postoperatively, was 130.1 degrees. Kaplan-Meier survivorship at 10-year was 95.9% with any operation or radiographic failure as the end point. The corresponding rate was 97.4% with revision of any component as the end point. No ceramic component fracture occurred. The present study demonstrates that good range of flexion was maintained for a long time after total knee arthroplasty with excellent durability. The Bi-Surface Knee System appears to have achieved its design objectives. (C) 2010 Elsevier B.V. All rights reserved.
引用
收藏
页码:274 / 278
页数:5
相关论文
共 46 条
[1]  
ABDEEN AR, 2009, J ARTHROPLA IN PRESS
[2]   Functional analysis of the effect of the posterior stabilising cam in two total knee replacements - A comparison of the Insall/Burstein and bisurface prostheses [J].
Akagi, M ;
Kaneda, E ;
Nakamura, T ;
Ueno, M ;
Yamanashi, W ;
Miyajima, H .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2002, 84B (04) :561-565
[3]   A mechanical comparison of 2 posterior-stabilizing designs - Insall/Burstein 2 knee and bisurface knee [J].
Akagi, M ;
Ueo, T ;
Takagi, H ;
Hamanishi, C ;
Nakamura, T .
JOURNAL OF ARTHROPLASTY, 2002, 17 (05) :627-634
[4]   The bisurface total knee replacement: A unique design for flexion - Four-to-nine-year follow-up study [J].
Akagi, M ;
Nakamura, T ;
Matsusue, Y ;
Ueo, T ;
Nishijyo, K ;
Ohnishi, E .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2000, 82A (11) :1626-1633
[5]   Cemented rotating-platform total knee replacement - A concise follow-up, at a minimum of fifteen years, of a previous report [J].
Callaghan, JJ ;
O'Rourke, MR ;
Iossi, MF ;
Liu, SS ;
Goetz, DD ;
Vittetoe, DA ;
Sullivan, PM ;
Johnston, RC .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2005, 87A (09) :1995-1998
[6]   Tibial post impingement in posterior-stabilized total knee arthroplasty [J].
Callaghan, JJ ;
O'Rourke, MR ;
Goetz, DA ;
Schmalzried, TP ;
Campbell, PA ;
Johnston, RC .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2002, (404) :83-88
[7]   PFC knee replacement - Osteolytic failures from extreme polyethylene degradation [J].
Casey, David ;
Cottrell, Jocelyn ;
DiCarlo, Edward ;
Windsor, Russell ;
Wright, Timothy .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2007, (464) :157-163
[8]   The stability of the cemented tibial component of total knee arthroplasty [J].
Catani, F ;
Leardini, A ;
Ensini, A ;
Cucca, G ;
Bragonzoni, L ;
Toksvig-Larsen, S ;
Giannini, S .
JOURNAL OF ARTHROPLASTY, 2004, 19 (06) :775-782
[9]   Knee Range of Motion During the First Two Years After Use of Posterior Cruciate-Stabilizing or Posterior Cruciate-Retaining Total Knee Prostheses A Randomized Clinical Trial [J].
Chaudhary, R. ;
Beaupre, L. A. ;
Johnston, D. W. C. .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2008, 90A (12) :2579-2586
[10]   Fracture of the polyethylene tibial post in a NexGen posterior-stabilized knee prosthesis [J].
Chiu, YS ;
Chen, WM ;
Huang, CK ;
Chiang, CC ;
Chen, TH .
JOURNAL OF ARTHROPLASTY, 2004, 19 (08) :1045-1049