High perimeter impedance defibrillation electrodes reduce skin burns in transthoracic cardioversion

被引:14
作者
Garcia, LA
Pagan-Carlo, LA
Stone, MS
Kerber, RE
机构
[1] Univ Iowa Hosp & Clin, Dept Internal Med, Div Cardiol, Iowa City, IA 52242 USA
[2] Univ Iowa Hosp & Clin, Dept Dermatol, Iowa City, IA 52242 USA
[3] Univ Iowa Hosp & Clin, Dept Pathol, Iowa City, IA 52242 USA
关键词
D O I
10.1016/S0002-9149(98)00571-2
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
During cardioversion, skin burns result from current preferentially flowing at the electrode edge. We tested new electrodes with high perimeter impedance to yield more uniform current distribution; these electrodes reduced histopathologic skin injury.
引用
收藏
页码:1125 / +
页数:4
相关论文
共 8 条
[1]   THE EFFECT OF GELLED-PAD DESIGN ON THE PERFORMANCE OF ELECTROSURGICAL DISPERSIVE ELECTRODES [J].
CARUSO, PM ;
PEARCE, JA ;
DEWITT, DP .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1982, 104 (04) :324-329
[2]  
FERRARI RK, 1996, Patent No. 5571165
[3]  
GEUZE RH, 1983, MED BIOL ENG COMPUT, V21, P518, DOI 10.1007/BF02442644
[4]   OPTIMAL ELECTRODE DESIGNS FOR ELECTROSURGERY, DEFIBRILLATION, AND EXTERNAL CARDIAC PACING [J].
KIM, Y ;
BENFAHY, J ;
TUPPER, BJ .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1986, 33 (09) :845-853
[5]   SIMULATED AND EXPERIMENTAL STUDIES OF TEMPERATURE ELEVATION AROUND ELECTROSURGICAL DISPERSIVE ELECTRODES [J].
KIM, YM ;
WEBSTER, JG ;
TOMPKINS, WJ .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1984, 31 (11) :681-692
[6]  
OLSON WH, 1978, MED INSTRUM, P667
[7]   Nature and determinants of skin ''burns'' after transthoracic cardioversion [J].
PaganCarlo, LA ;
Stone, MS ;
Kerber, RE .
AMERICAN JOURNAL OF CARDIOLOGY, 1997, 79 (05) :689-691
[8]   ANALYSIS AND CONTROL OF THE CURRENT DISTRIBUTION UNDER CIRCULAR DISPERSIVE ELECTRODES [J].