Feasibility of Swept Synthetic Aperture Ultrasound Imaging

被引:42
作者
Bottenus, Nick [1 ]
Long, Will [1 ]
Zhang, Haichong K. [2 ]
Jakovljevic, Marko [1 ]
Bradway, David P. [1 ]
Boctor, Emad M. [3 ]
Trahey, Gregg E. [1 ,4 ]
机构
[1] Duke Univ, Dept Biomed Engn, Durham, NC 27708 USA
[2] Johns Hopkins Univ, Dept Comp Sci, Baltimore, MD 21218 USA
[3] Johns Hopkins Univ, Dept Radiol, Baltimore, MD 21218 USA
[4] Duke Univ, Med Ctr, Dept Radiol, Durham, NC 27710 USA
基金
美国国家科学基金会;
关键词
Aberration; beamforming; image quality; image resolution; mechanical tracking; synthetic aperture; ultrasound imaging; ARRAY TRANSDUCER; PHASED-ARRAY; MOTION; RESOLUTION; ELEMENT; SYSTEMS; VIVO;
D O I
10.1109/TMI.2016.2524992
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
Ultrasound image quality is often inherently limited by the physical dimensions of the imaging transducer. We hypothesize that, by collecting synthetic aperture data sets over a range of aperture positions while precisely tracking the position and orientation of the transducer, we can synthesize large effective apertures to produce images with improved resolution and target detectability. We analyze the two largest limiting factors for coherent signal summation: aberration and mechanical uncertainty. Using an excised canine abdominal wall as a model phase screen, we experimentally observed an effective arrival time error ranging from 18.3 ns to 58 ns (root-mean-square error) across the swept positions. Through this clutter-generating tissue, we observed a 72.9% improvement in resolution with only a 3.75 dB increase in side lobe amplitude compared to the control case. We present a simulation model to study the effect of calibration and mechanical jitter errors on the synthesized point spread function. The relative effects of these errors in each imaging dimension are explored, showing the importance of orientation relative to the point spread function. We present a prototype device for performing swept synthetic aperture imaging using a conventional 1-D array transducer and ultrasound research scanner. Point target reconstruction error for a 44.2 degree sweep shows a reconstruction precision of 82.8 mu m and 17.8 mu m in the lateral and axial dimensions respectively, within the acceptable performance bounds of the simulation model. Improvements in resolution, contrast and contrast-to-noise ratio are demonstrated in vivo and in a fetal phantom.
引用
收藏
页码:1676 / 1685
页数:10
相关论文
共 41 条
  • [1] Synthetic Aperture Focusing for a Single-Element Transducer Undergoing Helical Motion
    Andresen, Henrik
    Nikolov, Svetoslav Ivanov
    Jensen, Jorgen Arendt
    [J]. IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2011, 58 (05) : 935 - 943
  • [2] Three-Dimensional Synthetic Aperture Focusing Using a Rocking Convex Array Transducer
    Andresen, Henrik
    Nikolov, Svetoslav Ivanov
    Pedersen, Mads Moller
    Buckton, Daniel
    Jensen, Jorgen Arendt
    [J]. IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2010, 57 (05) : 1051 - 1063
  • [3] [Anonymous], 2011, SONAR SYSTEMS
  • [4] Blackall J., 2000, P MICCA OCT
  • [5] Bottenus N., 2013, P SPIE, V8675, P1
  • [6] Bottenus N., 2015, P SPIE MED IMAG ULTR, V9419
  • [7] Bottenus N., 2015, PROC IEEE INT ULTRAS, P1
  • [8] Corl P. D., 1978, 1978 Ultrasonics Symposium Proceedings, P263, DOI 10.1109/ULTSYM.1978.197042
  • [9] DURGIN HW, 1992, ULTRASON, P623, DOI 10.1109/ULTSYM.1992.275925
  • [10] PHASE-ABERRATION CORRECTION USING SIGNALS FROM POINT REFLECTORS AND DIFFUSE SCATTERERS - BASIC PRINCIPLES
    FLAX, SW
    ODONNELL, M
    [J]. IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1988, 35 (06) : 758 - 767