Physical validation of a Monte Carlo-based, phantom-derived approach to computed tomography organ dosimetry under tube current modulation

被引:10
作者
Stepusin, Elliott J. [1 ]
Long, Daniel J. [2 ]
Ficarrotta, Kayla R. [1 ,3 ]
Hintenlang, David E. [1 ,4 ]
Bolch, Wesley E. [1 ]
机构
[1] Univ Florida, J Crayton Pruitt Family Dept Biomed Engn, Gainesville, FL 32611 USA
[2] Mem Sloan Kettering Canc Ctr, Dept Med Phys, 1275 York Ave, New York, NY 10065 USA
[3] Univ S Florida, Dept Chem & Biomed Engn, Tampa, FL 33620 USA
[4] Ohio State Univ, Dept Radiol, Wexner Med Ctr Suite 450, Columbus, OH 43210 USA
关键词
anthropomorphic phantom; computed tomography; Monte Carlo; organ dose; tube current modulation; DOSE CONVERSION COEFFICIENTS; RADIATION-EXPOSURE; CT EXAMINATIONS; SIMULATIONS; SPECTRA; SYSTEM; ADULTS; OSLD;
D O I
10.1002/mp.12461
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To physically validate the accuracy of a Monte Carlo-based, phantom-derived methodology for computed tomography (CT) dosimetry that utilizes organ doses from precomputed axial scans and that accounts for tube current modulation (TCM). Methods: The output of a Toshiba Aquilion ONE CT scanner was modeled, based on physical measurement, in the Monte Carlo radiation transport code MCNPX (v2.70). CT examinations were taken of two anthropomorphic phantoms representing pediatric and adult patients (15-yr-old female and adult male) at various energies, in which physical organ dose measurements were made using optically stimulated luminescence dosimeters (OSLDs). These exams (chest-abdomen-pelvis) were modeled using organ dose data obtained from the computationally equivalent phantom of each anthropomorphic phantom. TCM was accounted for by weighting all organ dose contributions by both the relative attenuation of the phantom and the image-derived mA value (from the DICOM header) at the same z-extent (cranial-caudal direction) of the axial dose data. Results: The root mean squares of percent difference in organ dose when comparing the physical organ dose measurements to the computational estimates were 21.2, 12.1, and 15.1% for the uniform (no attenuation weighting), weighted (computationally derived), and image-based methodologies, respectively. Conclusions: Overall, these data suggest that the Monte Carlo-based dosimetry presented in this work is viable for CT dosimetry. Additionally, for CT exams with TCM, local attenuation weighting of organ dose contributions from precomputed axial dosimetry libraries increases organ dose accuracy. (C) 2017 American Association of Physicists in Medicine
引用
收藏
页码:5423 / 5432
页数:10
相关论文
共 33 条
  • [1] Characteristics of an OSLD in the diagnostic energy range
    Al-Senan, Rani M.
    Hatab, Mustapha R.
    [J]. MEDICAL PHYSICS, 2011, 38 (07) : 4396 - 4405
  • [2] American Association of Physicists in Medicine, 2008, 96 AM ASS PHYS MED
  • [3] [Anonymous], NY TIMES
  • [4] [Anonymous], THESIS
  • [5] Accurate method for computer-generating tungsten anode x-ray spectra from 30 to 140 kV
    Boone, JM
    Seibert, JA
    [J]. MEDICAL PHYSICS, 1997, 24 (11) : 1661 - 1670
  • [6] Accuracy of Monte Carlo simulations compared to in-vivo MDCT dosimetry
    Bostani, Maryam
    Mueller, Jonathon W.
    McMillan, Kyle
    Cody, Dianna D.
    Cagnon, Chris H.
    DeMarco, John J.
    McNitt-Gray, Michael F.
    [J]. MEDICAL PHYSICS, 2015, 42 (02) : 1080 - 1086
  • [7] Attenuation-based size metric for estimating organ dose to patients undergoing tube current modulated CT exams
    Bostani, Maryam
    McMillan, Kyle
    Lu, Peiyun
    Kim, Hyun J.
    Cagnon, Chris H.
    DeMarco, John J.
    McNitt-Gray, Michael F.
    [J]. MEDICAL PHYSICS, 2015, 42 (02) : 958 - 968
  • [8] Validation of a Monte Carlo model used for simulating tube current modulation in computed tomography over a wide range of phantom conditions/challenges
    Bostani, Maryam
    McMillan, Kyle
    DeMarco, John J.
    Cagnon, Chris H.
    McNitt-Gray, Michael F.
    [J]. MEDICAL PHYSICS, 2014, 41 (11)
  • [9] Current concepts - Computed tomography - An increasing source of radiation exposure
    Brenner, David J.
    Hall, Eric J.
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 2007, 357 (22) : 2277 - 2284
  • [10] Stationary table CT dosimetry and anomalous scanner-reported values of CTDIvol
    Dixon, Robert L.
    Boone, John M.
    [J]. MEDICAL PHYSICS, 2014, 41 (01)