Evaluation of MatriXX for IMRT and VMAT dose verifications in peripheral dose regions

被引:41
作者
Han, Zhaohui [1 ,2 ]
Ng, Sook Kien [1 ,2 ]
Bhagwat, Mandar S. [1 ,2 ]
Lyatskaya, Yulia [1 ,2 ]
Zygmanski, Piotr [1 ,2 ]
机构
[1] Brigham & Womens Hosp, Dept Radiat Oncol, Boston, MA 02115 USA
[2] Harvard Univ, Sch Med, Boston, MA 02115 USA
关键词
MatriXX 2D ion chamber array; peripheral dose; IMRT QA; VMAT dose verification; MODULATED RADIATION-THERAPY; PIXEL IONIZATION-CHAMBER; QUALITY-ASSURANCE; ION-CHAMBER; DIODE-ARRAY; RADIOTHERAPY; PHOTON; CALIBRATION; DETECTOR; BEAMS;
D O I
10.1118/1.3455707
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: MatriXX is a two-dimensional ion chamber array designed for IMRT/VMAT (RapidArc, IMAT, etc.) dose verifications. Its dosimetric properties have been characterized for megavoltage beams in a number of studies; however, to the best of the authors' knowledge, there is still a lack of an investigation into its performance in the peripheral or low dose regions. In this work, the authors have carried out a systematic study on this issue. Methods: The authors compare the performance of MatriXX with a cylindrical ion chamber in solid water phantoms in the peripheral dose regions. The comparisons are performed for a number of typical irradiation conditions that involve different gantry and/or MLC motions, field sizes, and distances to the target including static gantry/open fields, static gantry/sweeping MLC gap (mimicking an IMRT delivery), dynamic gantry/oscillating sweeping MLC gap (mimicking a VMAT delivery), as well as clinical IMRT and VMAT plans. Results: MatriXX, when used according to the manufacturer's recommendations, is found to disagree with an ion chamber in peripheral dose regions. This disagreement has been attributed to four types of MatriXX errors, namely, positive bias, over-response to scattered doses, round-off error, and angular dependence, all of which contribute to dose inaccuracies in the peripheral regions. The positive bias, which is independent of the dose level, is cumulative when MatriXX operates in the movie mode. The accumulation is proportional to the number of movie frames (snaps) when the sampling time is greater than 500 ms and is proportional to the overall movie time for a sampling time shorter than 500 ms. This behavior suggests multiple sources of the bias. MatriXX is also found to over-respond to peripheral doses by about 2.0% for the regions investigated in this work (3-15 cm from the field edge), where phantom scatter and collimator scatter dominate. Round-off error is determined to be due to insufficient precision in conversion of the raw signals to MatriXX software data for low doses. Angular dependence is defined as the dose response of MatriXX at different gantry angles. Up to 8% difference in detector response has been observed between 0 and 180. Possible sources of these errors are discussed and a correction method is suggested. With corrections, MatriXX shows good agreement with the ion chamber in all cases involving different gantry and/or MLC dynamics, as well as the clinical plans. For both primary and peripheral doses, MatriXX shows dose linearity down to 2 cGy with an accuracy of within 1% of the local dose. Conclusions: The performance of MatriXX has been systematically evaluated in the peripheral dose regions. Major sources of error associated with MatriXX are identified and a correction method is suggested. This method has been successfully tested using both experimental and clinical plans. In all cases, good agreements between MatriXX and an ion chamber are achieved after corrections. The authors conclude that with proper corrections, MatriXX can be reliably used for peripheral dose measurements within the ranges studied. (c) 2010 American Association of Physicists in Medicine. [DOI: 10.1118/1.3455707]
引用
收藏
页码:3704 / 3714
页数:11
相关论文
共 24 条
[1]   Dosimetric characterization of a large area pixel-segmented ionization chamber [J].
Amerio, S ;
Boriano, A ;
Bourhaleb, F ;
Cirio, R ;
Donetti, M ;
Fidanzio, A ;
Garelli, E ;
Giordanengo, S ;
Madon, E ;
Marchetto, F ;
Nastasi, U ;
Peroni, C ;
Piermattei, A ;
Freire, CJS ;
Sardo, A ;
Trevisiol, E .
MEDICAL PHYSICS, 2004, 31 (02) :414-420
[2]   Use of a two-dimensional ionization chamber array for proton therapy beam quality assurance [J].
Arjomandy, Bijan ;
Sahoo, Narayan ;
Ding, Xiaoning ;
Gillin, Michael .
MEDICAL PHYSICS, 2008, 35 (09) :3889-3894
[3]   A feasibility study to calculate unshielded fetal doses to pregnant patients in 6-MV photon treatments using Monte Carlo methods and anatomically realistic phantoms [J].
Bednarz, Bryan ;
Xu, X. George .
MEDICAL PHYSICS, 2008, 35 (07) :3054-3061
[4]   Peripheral dose measurement with a MOSFET detector [J].
Butson, MJ ;
Cheung, T ;
Yu, PKN .
APPLIED RADIATION AND ISOTOPES, 2005, 62 (04) :631-634
[5]   Radiation damage studies of a recycling integrator VLSI chip for dosimetry and control of therapeutical beams [J].
Cirio, R ;
Bourhaleb, F ;
Degiorgis, PG ;
Donetti, M ;
Marchetto, F ;
Marletti, M ;
Mazza, G ;
Peroni, C ;
Rizzi, E ;
SanzFreire, C .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2002, 482 (03) :752-760
[6]   A method for the inter-calibration of a matrix of sensors [J].
Donetti, M ;
Garelli, E ;
Marchetto, F ;
Boriano, A ;
Bourhaleb, F ;
Cirio, R ;
Cornelius, I ;
Giordanengo, S ;
La Rosa, A ;
Nastasi, U ;
Peroni, C .
PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (03) :485-495
[7]   Intensity-modulated radiation therapy, protons, and the risk of second cancers [J].
Hall, Eric J. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2006, 65 (01) :1-7
[8]   Dosimetric evaluation of a 2D pixel ionization chamber for implementation in clinical routine [J].
Herzen, J. ;
Todorovic, M. ;
Cremers, F. ;
Platz, V. ;
Albers, D. ;
Bartels, A. ;
Schmidt, R. .
PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (04) :1197-1208
[9]  
*IBA DOS GMBH, OMNIPRO IMRT US GUID
[10]   A 2-D diode array and analysis software for verification of intensity modulated radiation therapy delivery [J].
Jursinic, PA ;
Nelms, BE .
MEDICAL PHYSICS, 2003, 30 (05) :870-879