Design of synchrotron light source and its beamline dedicated to dual-energy x-ray computed tomography

被引:32
作者
Torikoshi, M
Tsunoo, T
Endo, M
Noda, K
Kumada, M
Yamada, S
Soga, F
Hyodo, K
机构
[1] Natl Inst Radiol Sci, Inage Ku, Chiba 2638555, Japan
[2] High Energy Accelerator Res Org, Tsukuba, Ibaraki 3050801, Japan
关键词
x-ray computed tomography; synchrotron radiation; monochromatic x ray; electron density;
D O I
10.1117/1.1383062
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A synchrotron light source dedicated to medical applications has been designed at National Institute of Radiological Sciences. The storage ring, with circumference of 80 m, is designed for acceleration of 2.3 GeV and a stored current of 420 mA. It is equipped with two multipole wigglers to produce sufficient photon flux in a hard x-ray region required for medical applications. The purposes of the synchrotron light source are clinical performance of medical diagnoses clinically and research and development relating with medical applications. One of the most interesting applications for us is dual-energy x-ray computed tomography (CT). It gives the information about electron density of human tissue. The information plays an important role in advancing heavy-ion radiotherapy of cancers. Electron density can be derived from attenuation coefficients measured by different energy x rays. In this paper, a practical method of the dual-energy x-ray CT with synchrotron radiation is proposed with the theoretical consideration. The primitive experiment using monochromatic x rays emitted from radioisotopes proved the procedure of analysis mentioned here effective to derive electron densities from linear attenuation coefficients for two x rays of a different energy. The beamline dedicated to dual-energy x-ray CT is also proposed. It has a multipole wiggler as a light source and it mainly consists of a dual crystal monochromator and a rotating filter for attenuating photon flux of x rays and two-dimensional detector. (C) 2001 Society of Photo-Optical Instrumentation Engineers.
引用
收藏
页码:371 / 377
页数:7
相关论文
共 50 条
  • [41] Design of x-ray diagnostic beam line for a synchrotron radiation source and measurement results
    Garg, Akash Deep
    Karnewar, A. K.
    Ojha, A.
    Shrivastava, B. B.
    Holikatti, A. C.
    Puntambekar, T. A.
    Navathe, C. P.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2014, 754 : 15 - 23
  • [42] Design considerations of source of synchrotron hard X-ray and coherent infrared radiation at JINR
    Alexandrov, VS
    Beloshitsky, PF
    Kazarinov, NY
    Perelstein, EA
    Shaliapin, VN
    Sissakian, AN
    Tyutyunnikov, SI
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2000, 448 (1-2) : 32 - 36
  • [43] Four-dimensional X-ray phase tomography with Talbot interferometer and white synchrotron light
    Momose, Atsushi
    Yashiro, Wataru
    Harasse, Sebastien
    Kuwabara, Hiroaki
    Kawabata, Katsuyuki
    DEVELOPMENTS IN X-RAY TOMOGRAPHY VII, 2010, 7804
  • [44] DAFNE-Light DXR1 Soft X-Ray Synchrotron Radiation Beamline: Characteristics and XAFS Applications
    Balerna, Antonella
    CONDENSED MATTER, 2019, 4 (01): : 1 - 12
  • [45] Observation of internal crack of steel bar by computed tomography and strain mapping near its tip using synchrotron white X-ray
    Shibano, Jun-Ichi
    Kiriyama, Kouji
    Kajiwara, Kentaro
    Shobu, Takahisa
    Suzuki, Kenji
    Arai, Takayuki
    Miura, Setsuo
    Kobayashi, Michiaki
    Zairyo/Journal of the Society of Materials Science, Japan, 2009, 58 (07) : 596 - 602
  • [46] X-ray imaging of a small electron beam in a low-emittance synchrotron light source
    Takano, S
    Masaki, M
    Ohkuma, H
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2006, 556 (01) : 357 - 370
  • [47] Measurement of K series of X-ray fluorescence spectra of Ce and Gd by high energy X-ray from synchrotron radiation source
    Harada, Masaaki
    Shoji, Masahiko
    Kawata, Hiroshi
    Sakurai, Kenji
    BUNSEKI KAGAKU, 2006, 55 (06) : 433 - 439
  • [48] The use of X-ray computed tomography for design and process modeling of aerospace composites: A review
    Naresh, K.
    Khan, K. A.
    Umer, R.
    Cantwell, W. J.
    MATERIALS & DESIGN, 2020, 190
  • [49] Effective atomic number measurement with energy-resolved X-ray computed tomography
    Kanno, Ikuo
    Yamashita, Yoshiki
    Kimura, Masashi
    Inoue, Fukutaro
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2015, 787 : 121 - 124
  • [50] Measurement of coherence length and incoherent source size of hard x-ray undulator beamline at Pohang Light Source-II
    Park, So Yeong
    Hong, Chung Ki
    Lim, Jun
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2014, 85 (04)