Diurnal Intraocular Pressure Fluctuations with Self-tonometry in Glaucoma Patients and Suspects: A Clinical Trial

被引:32
作者
Huang, Jessie [1 ,2 ]
Katalinic, Paula [1 ,2 ]
Kalloniatis, Michael [1 ,2 ]
Hennessy, Michael P. [1 ,3 ]
Zangerl, Barbara [1 ,2 ]
机构
[1] Univ New South Wales Sydney, Ctr Eye Hlth, Sydney, NSW, Australia
[2] Univ New South Wales Sydney, Sch Optometry & Vis Sci, Sydney, NSW, Australia
[3] Prince Wales Hosp, Dept Ophthalmol, Sydney, NSW, Australia
基金
英国医学研究理事会;
关键词
OCULAR HYPERTENSION; REBOUND TONOMETER; RISK-FACTOR; ICARE-ONE; PROGRESSION; 24-HOUR; ACCURACY; TERM; PATTERNS;
D O I
10.1097/OPX.0000000000001172
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
SIGNIFICANCE: This article shows that self-tonometry can provide robust measures of diurnal intraocular pressure (IOP) and also detect changes to IOP in response to treatment within a short period of monitoring. These advances in IOP monitoring may contribute to improved management of glaucoma patients and suspects. PURPOSE: The aim of this study was to prospectively investigate the utility of rebound self-tonometry performed over several weeks for detecting diurnal IOP fluctuations in glaucoma patients and suspects and also initial response to topical treatment in glaucoma patients. METHODS: Forty patients were recruited following glaucoma-specific examination. Subsequent to successful training with the iCare HOME tonometer, patients were instructed to measure IOP, in a sitting position, four times a day over 4 to 6 weeks. Date, time, laterality, and IOP downloaded from the tonometer and clinical examination data, including applanation IOP and corneal thickness, were analyzed. A user satisfaction survey was also administered at study completion. t Test and analysis of variance were used to compare groups and IOP across days. Pearson correlation was used to compare measurements to Goldmann applanation tonometry and IOP measurements from the first day/s to the overall mean IOP. RESULTS: Twenty-seven patients (18 suspects and 9 glaucoma patients) completed data collection. Patients self-measured IOP on 118 (29) occasions for 40 (+/- 7.4) days. Two dominant patterns of fluctuation were revealed: peak IOP upon awakening (n = 11) and at midday (n = 13). Diurnal IOP measured in the first 7 days showed strong correlation to diurnal IOP across the entire study period (r(2) = 0.82, P < .0001). Within 24 hours of treatment commencement (latanoprost 0.005% ophthalmic solution), IOP reduced from 23.9 (+/- 5.2) to 16.1 (+/- 2.6) mmHg. Overall, patients rated the instrument as easy to use, although difficulties with correct alignment were expressed. CONCLUSIONS: Rebound self-tonometry demonstrated utility for measuring diurnal IOP fluctuations in most patients, hence enhancing management of patient with or at risk of developing glaucoma.
引用
收藏
页码:88 / 95
页数:8
相关论文
共 45 条
[1]   Statistical guidelines for the analysis of data obtained from one or both eyes [J].
Armstrong, Richard A. .
OPHTHALMIC AND PHYSIOLOGICAL OPTICS, 2013, 33 (01) :7-14
[2]  
Asrani S, 2000, J GLAUCOMA, V9, P134
[3]  
Australian Government National Health and Medical Research Council (NHMRC), NHMRC GUID SCREEN PR
[4]   Clinical utility of Intraocular pressure monitoring outside of normal office hours in patients with glaucoma [J].
Barkana, Yaniv ;
Anis, Sarah ;
Liebmann, Jeffrey ;
Tello, Celso ;
Ritch, Robert .
ARCHIVES OF OPHTHALMOLOGY, 2006, 124 (06) :793-797
[5]   Fluctuation of intraocular pressure and glaucoma progression in the early manifest glaucoma trial [J].
Bengtsson, Boel ;
Leske, M. Cristina ;
Hyman, Leslie ;
Heijl, Anders .
OPHTHALMOLOGY, 2007, 114 (02) :205-209
[6]   Impact of intraocular pressure regulation on visual fields in open-angle glaucoma [J].
Bergeå, B ;
Bodin, L ;
Svedbergh, B .
OPHTHALMOLOGY, 1999, 106 (05) :997-1004
[7]   Intraocular pressure fluctuation - A risk factor for visual field progression at low intraocular pressures in the advanced glaucoma intervention study [J].
Caprioli, Joseph ;
Coleman, Anne L. .
OPHTHALMOLOGY, 2008, 115 (07) :1123-1129
[8]   Self-tonometry as a complement in the investigation of glaucoma patients [J].
Chen, Enping ;
Querat, Laurence ;
Akerstedt, Christina .
ACTA OPHTHALMOLOGICA, 2016, 94 (08) :788-792
[9]   Twenty-four-hour efficacy of preservative-free tafluprost for open-angle glaucoma patients, assessed by home intraocular pressure (Icare-ONE) and blood-pressure monitoring [J].
Cho, Soon Young ;
Kim, Yong Yeon ;
Yoo, Chungkwon ;
Lee, Tae-Eun .
JAPANESE JOURNAL OF OPHTHALMOLOGY, 2016, 60 (01) :27-34
[10]   Evaluation of a new rebound tonometer for self-measurement of intraocular pressure [J].
Dabasia, Priya L. ;
Lawrenson, John G. ;
Murdoch, Ian E. .
BRITISH JOURNAL OF OPHTHALMOLOGY, 2016, 100 (08) :1139-1143