Radiation dose reduction with frame rate conversion in X-ray fluoroscopic imaging systems with flat panel detector: basic study and clinical retrospective analysis

被引:8
作者
Sakai, Noriyuki [1 ]
Tabei, Katsuyuki [1 ]
Sato, Jiro [1 ]
Imae, Toshikazu [1 ]
Suzuki, Yuichi [1 ]
Takenaka, Shigeharu [1 ]
Yano, Keiichi [1 ]
Abe, Osamu [1 ]
机构
[1] Univ Tokyo Hosp, Dept Radiol, Bunkyo Ku, 7-3-1 Hongo, Tokyo 1138655, Japan
基金
日本学术振兴会;
关键词
Fluoroscopy; Retrospective studies; Comparative study; X-rays; Radiation dosage; AUTOMATED PATIENT RECOGNITION; INDUCED SKIN INJURIES; CHEST RADIOGRAPHS; INTENSIFIER; CHILDREN;
D O I
10.1007/s00330-018-5620-y
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
ObjectivesTo (a) evaluate the interpolation frames of frame rate conversion (FRC) compared with fluoroscopic frames of conventional method, and (b) compare radiation dose and fluoroscopy time between various clinical examinations without and with FRC retrospectively.MethodsThis study consisted of a basic study and a clinical retrospective analysis. The radiation dosimetry, visual assessment and measurements of contrast to noise ratio were examined. Similarity between interpolation frames and fluoroscopic frames was evaluated using normalised cross-correlation values. In the clinical retrospective analysis approved by the institutional review board, we extracted 270 examinations performed without FRC (conventional group, 12.5 pulses/s) and with FRC (FRC group, 6.25 pulses/s) from 23 May to 31 December 2016. The fluoroscopy parameters and demographics of the two groups of the clinical examinations were compared. Statistical analyses were performed with Wilcoxon signed-rank test, Brunner-Munzel test and (2) test.ResultsIn the basic study, the only significant difference was that the radiation dose of FRC was approximately half that of the conventional method in the same fluoroscopy time (p = .031). The interpolation frames of FRC were similar to the fluoroscopic frames of the conventional method. In the clinical retrospective analysis, the only significant difference was that FRC reduced the fluoroscopy dose by 48% and the total dose by 31% compared with the conventional method (p < .001). There was no significant difference in the others.ConclusionFRC significantly reduced the radiation dose without extending the fluoroscopy time and maintaining the image quality compared to the conventional method.Key Points center dot Although X-ray fluoroscopic techniques are widely used for various clinical purposes, X-ray fluoroscopic examinations have radiation risks.center dot Frame rate conversion is an image processing technique for radiation dose reduction.center dot Clinical retrospective analysis showed that FRC reduces radiation doses of patients.
引用
收藏
页码:985 / 992
页数:8
相关论文
共 16 条
  • [1] How to set up and apply reference levels in fluoroscopy at a national level
    Aroua, A.
    Rickli, H.
    Stauffer, J.-C.
    Schnyder, P.
    Trueb, P. R.
    Valley, J.-F.
    Vock, P.
    Verdun, F. R.
    [J]. EUROPEAN RADIOLOGY, 2007, 17 (06) : 1621 - 1633
  • [2] Radiation-induced skin injuries and fluoroscopy
    Berlin, L
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2001, 177 (01) : 21 - 25
  • [3] Risk of cancer from diagnostic X-rays:: estimates for the UK and 14 other countries
    Berrington de González, A
    Darby, S
    [J]. LANCET, 2004, 363 (9406) : 345 - 351
  • [4] The ALARA concept in pediatric cardiac catheterization: techniques and tactics for managing radiation dose
    Justino, Henri
    [J]. PEDIATRIC RADIOLOGY, 2006, 36 (Suppl 2) : 146 - 153
  • [5] Measurement of observer agreement
    Kundel, HL
    Polansky, M
    [J]. RADIOLOGY, 2003, 228 (02) : 303 - 308
  • [6] Ultra low-dose VCUG in children using a modern flat detectorunit
    Linke, Sara Y. S.
    Tsiflikas, Ilias
    Herz, Klaus
    Szavay, Phillipp
    Gatidis, Sergios
    Schaefer, Juergen F.
    [J]. EUROPEAN RADIOLOGY, 2016, 26 (06) : 1678 - 1685
  • [7] Comparison between radiation exposure levels using an image intensifier and a flat-panel detector-based system in image-guided central venous catheter placement in children weighing less than 10 kg
    Miraglia, Roberto
    Maruzzelli, Luigi
    Cortis, Kelvin
    Piazza, Marcello
    Gerasia, Roberta
    Maggio, Simona
    Tuzzolino, Fabio
    Luca, Angelo
    [J]. PEDIATRIC RADIOLOGY, 2015, 45 (02) : 235 - 240
  • [8] Potential usefulness of biological fingerprints in chest radiographs for automated patient recognition and identification
    Morishita, J
    Katsuragawa, S
    Sasaki, Y
    Doi, K
    [J]. ACADEMIC RADIOLOGY, 2004, 11 (03) : 309 - 315
  • [9] An automated patient recognition method based on an image-matching technique using previous chest radiographs in the picture archiving and communication system environment
    Morishita, J
    Katsuragawa, S
    Kondo, K
    Doi, K
    [J]. MEDICAL PHYSICS, 2001, 28 (06) : 1093 - 1097
  • [10] Fluoroscopy-controlled voiding cystourethrography in infants and children: Are the radiation risks trivial?
    Perisinakis, K
    Raissaki, M
    Damilakis, J
    Stratakis, J
    Neratzoulakis, J
    Gourtsoyiannis, N
    [J]. EUROPEAN RADIOLOGY, 2006, 16 (04) : 846 - 851