Very low wear of non-remelted highly cross-linked polyethylene cups -: An RSA study lasting up to 6 years

被引:52
作者
Rohri, Stephan M. [1 ]
Li, Ming Guo [2 ]
Nilsson, Kjell-Gunnar [3 ]
Nivbrant, Bo [2 ]
机构
[1] Ullevaal Univ Hosp, Dept Orthopaed, Div Neurosci Orthopaed & Rehabil, Oslo, Norway
[2] Univ Western Australia, Perth Orthopaed Inst, Perth, WA, Australia
[3] Umea Univ Hosp, Umea, Sweden
关键词
D O I
10.1080/17453670710014509
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background and purpose Highly cross-linked polyethylenes (PEs) all appear to reduce wear dramatically in laboratory studies, although there is substantial variation in this respect between manufacturers. Non-remelted cross-linked PE is almost as tough as unirradiated PE, but is not completely stable and can oxidize in vivo, as has been shown in recent retrievals studies. We had reported low wear and good clinical performance after 2 years for 10 non-remelted highly crosslinked PE cups compared to 16 conventional PE cups sterilized by gamma-in-air. Method Because of possible degradation by free radicals, we followed up both cohorts for 5 years (conventional PE) and 6 years (highly cross-linked PE). Result Mean (CI) proximal head penetration over the observation time was linear and measured 0.08 (0.02-0.13) mm for cross-linked PE and 0.42 (0.23-0.62) mm for conventional PE, and total penetration was 0.23 (0.1-0.35) mm and 0.75 (0.05-1.4) mm respectively. After subtracting creep, the annual wear for non-remelted highly cross-linked PE was below 6 put. The cups had equally low migration and few radiolucencies. Interpretation The theoretical possibility of oxidation in non-remelted highly cross-linked PE may not show clinically. However, it may be that cemented cups with their thicker PE are more forgiving than metal-backed cups with thin PE moving in the locking mechanism. So far, we can conclude that the Crossfire highly crosslinked polyethylene cups performed very well clinically, with extremely low wear even after almost 6 years. This is reassuring, but care should be taken in extrapolating these results to other cross-linked PEs or uncemented cups where toughness of PE is more of an issue.
引用
收藏
页码:739 / 745
页数:7
相关论文
共 36 条
  • [1] Wear and surface cracking in early retrieved highly cross-linked polyethylene acetabular liners
    Bradford, L
    Baker, DA
    Graham, J
    Chawan, A
    Ries, MD
    Pruitt, LA
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2004, 86A (06) : 1271 - 1282
  • [2] Deng M, 1998, J BIOMED MATER RES, V40, P214, DOI 10.1002/(SICI)1097-4636(199805)40:2<214::AID-JBM6>3.3.CO
  • [3] 2-M
  • [4] Highly cross-linked polyethylene in cemented THA -: Randomized study of 61 hips
    Digas, G
    Kärrholm, J
    Thanner, J
    Malchau, H
    Herberts, P
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2003, (417) : 126 - 138
  • [5] The Otto Aufranc Award -: Highly cross-linked polyethylene in Total Hip Arthroplasty -: Randomized evaluation of penetration rate in cemented and uncemented sockets using radiostereometric analysis
    Digas, G
    Kärrholm, J
    Thanner, J
    Malchau, H
    Herberts, P
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2004, (429) : 6 - 16
  • [6] DIGAS G, 2005, ACTA ORTHOP SCAND S, V315, P3
  • [7] Clinical performance of a durasul highly cross-linked polyethylene acetabular liner for total hip arthroplasty at five years
    Dorr, LD
    Wan, ZN
    Shahrdar, C
    Sirianni, L
    Boutary, M
    Yun, A
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2005, 87A (08) : 1816 - 1821
  • [8] The basis for a second-generation highly cross-linked UHMWPE
    Dumbleton, John H.
    D'Antonio, James A.
    Manley, Michael T.
    Capello, William N.
    Wang, Aiguo
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2006, (453) : 265 - 271
  • [9] A randomized prospective evaluation of outcomes after total hip arthroplasty using cross-linked marathon and non-cross-linked Enduron polyethylene liners
    Engh, C. Anderson, Jr.
    Stepniewski, Adam S.
    Ginn, Stuart D.
    Beykirch, Sarah E.
    Sychterz-Terefenko, Christi J.
    Hopper, Robert H., Jr.
    Engh, Charles A.
    [J]. JOURNAL OF ARTHROPLASTY, 2006, 21 (06) : 17 - 25
  • [10] Wear, debris, and biologic activity of cross-linked polyethylene in the knee - Benefits and potential concerns
    Fisher, J
    McEwen, HMJ
    Tipper, JL
    Galvin, AL
    Ingram, J
    Kamali, A
    Stone, MH
    Ingham, E
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2004, (428) : 114 - 119