Effects of bismuth breast shielding on iodine quantification in dual-energy computed tomography: an experimental phantom study

被引:2
作者
Jang, Hyunsik [1 ]
Hur, Jin [1 ]
Choi, Byoung Wook [1 ]
Im, Dong Jin [1 ]
Hong, Yoo Jin [1 ]
Kim, Young Jin [1 ]
Lee, Hye-Jeong [1 ]
机构
[1] Yonsei Univ, Severance Hosp, Coll Med, Dept Radiol,Res Inst Radiol Sci, Seoul, South Korea
基金
新加坡国家研究基金会;
关键词
Bismuth; dual-energy computed tomography; iodine; radiation protection; radiation dosage; THORACIC CT; IONIZING-RADIATION; IMAGE QUALITY; TUBE-CURRENT; CANCER; REDUCTION;
D O I
10.1177/0284185118768098
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: Although the bismuth breast shield can reduce radiation exposure to the breast during dual-energy computed tomography (DECT), it can potentially affect material quantification on DECT due to artifacts. Purpose: To evaluate the effects of bismuth breast shielding on iodine quantification and radiation exposure in DECT. Material and Methods: Small balloons were made with 0.2%, 0.6% and 1.0% blended iodinated contrast (370 mg/mL of iodine) with water. The balloons were located at both anterior and posterior lungs in an adult anthropomorphic chest phantom. DECT was performed with and without breast shielding. Afterwards, iodine concentration values were measured for each balloon on the iodine maps. Absorbed radiation doses in the breast were measured with the optically stimulated luminescence dosimeter. Results: After shielding, we obtained significantly decreased iodine quantification for all three concentrations with 0.78 +/- 0.13 to 0.46 +/- 0.13 mg/mL, 2.31 +/- 0.17 to 1.68 +/- 0.19 mg/mL, and 3.82 +/- 0.10 to 2.84 +/- 0.20 mg/mL at the anterior location, and 0.72 +/- 0.11 to 0.48 +/- 0.09 mg/mL, 2.24 +/- 0.13 to 1.87 +/- 0.21 mg/mL, and 3.75 +/- 0.16 to 3.15 +/- 0.14 mg/mL at the posterior location for the 0.2%, 0.6%, and 1.0% balloons, respectively (P = 0.001 for all). After shielding, absorbed radiation doses to the breast significantly decreased by 14.8% (4.32 +/- 0.33 to 3.68 +/- 0.30 mGy; P= 0.005). Conclusion: Although using the bismuth breast shield may decrease radiation exposure to the breast on DECT, it may also significantly affect iodine quantification.
引用
收藏
页码:1475 / 1481
页数:7
相关论文
共 23 条
  • [1] Characteristics of an OSLD in the diagnostic energy range
    Al-Senan, Rani M.
    Hatab, Mustapha R.
    [J]. MEDICAL PHYSICS, 2011, 38 (07) : 4396 - 4405
  • [2] TISSUE SIGNATURES WITH DUAL-ENERGY COMPUTED-TOMOGRAPHY
    DICHIRO, G
    BROOKS, RA
    KESSLER, RM
    JOHNSTON, GS
    JONES, AE
    HERDT, JR
    SHERIDAN, WT
    [J]. RADIOLOGY, 1979, 131 (02) : 521 - 523
  • [3] In-plane bismuth breast shields for pediatric CT: Effects on radiation dose and image quality using experimental and clinical data
    Fricke, BL
    Donnelly, LF
    Frush, DP
    Yoshizumi, T
    Varchena, V
    Poe, SA
    Lucaya, J
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2003, 180 (02) : 407 - 411
  • [4] Quantitative assessment of selective in-plane shielding of tissues in computed tomography through evaluation of absorbed dose and image quality
    Geleijns, J.
    Salvado Artells, M.
    Veldkamp, W. J. H.
    Lopez Tortosa, M.
    Calzado Cantera, A.
    [J]. EUROPEAN RADIOLOGY, 2006, 16 (10) : 2334 - 2340
  • [5] QUANTITATIVE BONE-MINERAL ANALYSIS USING DUAL ENERGY COMPUTED TOMOGRAPHY
    GENANT, HK
    BOYD, D
    [J]. INVESTIGATIVE RADIOLOGY, 1977, 12 (06) : 545 - 551
  • [6] Ionising radiation and cancer risks: What have we learned from epidemiology?
    Gilbert, Ethel S.
    [J]. INTERNATIONAL JOURNAL OF RADIATION BIOLOGY, 2009, 85 (06) : 467 - 482
  • [7] Shielding Artificially Increases the Attenuation of Water: Study of CT Gradient Attenuation Induced by Shielding (CT GAINS)
    Goldin, Dmitry
    Emerson, Scott
    Farah, Michael C.
    Sokhandon, Farnoosh
    Nandalur, Kiran R.
    Kumar, Shawn
    Mezwa, Duane G.
    [J]. ACADEMIC RADIOLOGY, 2013, 20 (02) : 231 - 237
  • [8] Orbital, thyroid, and breast superficial radiation shielding for patients undergoing diagnostic CT
    Hopper, KD
    [J]. SEMINARS IN ULTRASOUND CT AND MRI, 2002, 23 (05) : 423 - 427
  • [9] The breast: In-plane x-ray protection during diagnostic thoracic CT - Shielding with bismuth radioprotective garments
    Hopper, KD
    King, SH
    Lobell, ME
    TenHave, TR
    Weaver, JS
    [J]. RADIOLOGY, 1997, 205 (03) : 853 - 858
  • [10] Dual-Energy CT: General Principles
    Johnson, Thorsten R. C.
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2012, 199 (05) : S3 - S8