Standard Versus Custom-Made Acetabular Implants in Revision Total Hip Arthroplasty

被引:18
作者
Tikhilov, Rashid M. [1 ]
Dzhavadov, Alisagib A. [1 ]
Kovalenko, Anton N. [1 ]
Bilyk, Stanislav S. [1 ]
Denisov, Alexey O. [1 ]
Shubnyakov, Igor I. [1 ]
机构
[1] Minist Hlth Russian Federat, Vreden Natl Med Res Ctr Traumatol & Orthoped, St Petersburg, Russia
关键词
hip joint; revision total hip arthroplasty; custom-made acetabular implants; standard acetabular implants; three-dimensional imaging; FOLLOW-UP; BONE DEFECT; RECONSTRUCTION; COMPONENTS; AUGMENTS; CLASSIFICATION; SURVIVAL; GRAFT; CUPS;
D O I
10.1016/j.arth.2021.09.003
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Today, various options are used for the reconstruction of acetabular bone loss in revision total hip arthroplasty (RTHA). The aim of the study was to compare the outcomes of using standard acetabular implants (SAIs) and custom-made acetabular implants (CMAIs) in RTHA in cases with extensive acetabular bone loss. Methods: This was a comparative analysis of the results of 106 operations of RTHA performed during the period from January 2013 to December 2019. In 61 cases (57.5%), CMAIs were used. In 45 cases (42.5%), SAIs were implanted. Results: The incidence of aseptic loosening of the acetabular component after RTHA in uncontained loss of bone stock of the acetabulum (type III-IV as per the Gross and Saleh classification) using the CMAI was less than that using the SAI (2.4% and 10.0%, respectively). The most significant differences in aseptic loosening rates were noted after implantation of the CMAI and SAI in pelvic discontinuity with uncontained bone defect (0% and 60.0%, respectively; P < .001). Conclusion: The ideal indications for the use of the CMAI are uncontained defects and pelvic discontinuity with uncontained loss of bone stock (types III-V Gross and Saleh classification). Treatment of these defects with the SAI leads to a higher incidence of aseptic loosening requiring re-revisions. Further observation is required to assess the effectiveness of using the CMAI and SAI in the long-term follow-up period. (c) 2021 Elsevier Inc. All rights reserved.
引用
收藏
页码:119 / 125
页数:7
相关论文
共 35 条
[1]   Management of Paprosky type three B acetabular defects by custom-made components: early results [J].
Aprato, Alessandro ;
Giachino, Matteo ;
Bedino, Paolo ;
Mellano, Danilo ;
Piana, Raimondo ;
Masse, Alessandro .
INTERNATIONAL ORTHOPAEDICS, 2019, 43 (01) :117-122
[2]   Are Custom Triflange Acetabular Components Effective for Reconstruction of Catastrophic Bone Loss? [J].
Berasi, Carl C. ;
Berend, Keith R. ;
Adams, Joanne B. ;
Ruh, Erin L. ;
Lombardi, Adolph V., Jr. .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2015, 473 (02) :528-535
[3]   Comparative Epidemiology of Revision Arthroplasty: Failed THA Poses Greater Clinical and Economic Burdens Than Failed TKA [J].
Bozic, Kevin J. ;
Kamath, Atul F. ;
Ong, Kevin ;
Lau, Edmund ;
Kurtz, Steve ;
Chan, Vanessa ;
Vail, Thomas P. ;
Rubash, Harry ;
Berry, Daniel J. .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2015, 473 (06) :2131-2138
[4]   Acetabular custom-made implants for severe acetabular bone defect in revision total hip arthroplasty: a systematic review of the literature [J].
Chiarlone, Francesco ;
Zanirato, Andrea ;
Cavagnaro, Luca ;
Alessio-Mazzola, Mattia ;
Felli, Lamberto ;
Burastero, Giorgio .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2020, 140 (03) :415-424
[5]  
Christie MJ, 2001, CLIN ORTHOP RELAT R, P216
[6]  
DANTONIO JA, 1989, CLIN ORTHOP RELAT R, P126
[7]   Survivorship and Clinical Outcomes of Custom Triflange Acetabular Components in Revision Total Hip Arthroplasty: A Systematic Review [J].
De Martino, Ivan ;
Strigelli, Vanni ;
Cacciola, Giorgio ;
Gu, Alex ;
Bostrom, Mathias P. ;
Sculco, Peter K. .
JOURNAL OF ARTHROPLASTY, 2019, 34 (10) :2511-2518
[8]   Revision total hip arthroplasty for pelvic discontinuity [J].
DeBoer, David K. ;
Christie, Michael J. ;
Brinson, Martha F. ;
Morrison, J. Craig .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2007, 89A (04) :835-840
[9]  
Fryhofer George W, 2020, J Clin Orthop Trauma, V11, P22, DOI 10.1016/j.jcot.2019.11.004
[10]   Trabecular Titanium™ cups and augments in revision total hip arthroplasty: clinical results, radiology and survival outcomes [J].
Gallart, Xavier ;
Fernandez-Valencia, Jenaro A. ;
Riba, Josep ;
Bori, Guillem ;
Garcia, Sebastian ;
Tornero, Eduard ;
Combalia, Andreu .
HIP INTERNATIONAL, 2016, 26 (05) :486-491