A Label-Free Interdigitated Microelectrodes Immunosensor for Pesticide Detection

被引:4
作者
Marrakchi, M. [1 ]
Campos Sanchez, I. [2 ]
Helali, S. [3 ]
Mejri, N. [1 ]
Soto Camino, J. [2 ]
Gonzalez-Martinez, M. A. [2 ]
Hamdi, M. [1 ]
Abdelghani, A. [3 ]
机构
[1] Carthage Univ, Lab Ecol & Technol Microbienne, INSAT, Charguia 1080, Tunisia
[2] Univ Politecn Valencia, Dept Quim, Valencia 46022, Spain
[3] Carthage Univ, Nanotechnol Lab, INSAT, Charguia 1080, Tunisia
关键词
Immunosensor; Atrazine; Physisorption; Electrochemical Impedance Spectroscopy; SOLID-PHASE EXTRACTION; LIQUID-CHROMATOGRAPHY; ATRAZINE RESIDUES; SAMPLES; IMMUNOASSAY; PERFORMANCE; HERBICIDES; DESIGN; WATER; DISKS;
D O I
10.1166/sl.2011.1800
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A sensitive, simple and label-free immunosensor for atrazine detection was developed. The immunosensor was based on antibody physisorption as immobilization technique. This technique was proposed as a cost-effective gold immuno-functionalization strategy alternative to SAM-based strategy for pesticide detection. Each step of immunosensor construction was characterized by electrochemical impedance spectroscopy, and the antibody-atrazine interaction was measured in a dynamic range from 10 to 150 ng/mL. This method was highly selective for atrazine and propazine, showing weak or no cross-reactivity to other similar compounds like terbutylazine. A limit detection of 10 ng/mL of atrazine was detected in PBS buffer with a good reproductibility. Lower limit detection in tap water was obtained.
引用
收藏
页码:2203 / 2206
页数:4
相关论文
共 17 条
[1]   USE OF EXTRACTION DISKS FOR TRACE ENRICHMENT OF VARIOUS PESTICIDES FROM RIVER WATER AND SIMULATED SEAWATER SAMPLES FOLLOWED BY LIQUID-CHROMATOGRAPHY RAPID-SCANNING UV VISIBLE AND THERMOSPRAY MASS-SPECTROMETRY DETECTION [J].
BARCELO, D ;
DURAND, G ;
BOUVOT, V ;
NIELEN, M .
ENVIRONMENTAL SCIENCE & TECHNOLOGY, 1993, 27 (02) :271-277
[2]   A Physisorbed Interface Design of Biomolecules for Selective and Sensitive Protein Detection [J].
Choi, Seokheun ;
Chae, Junseok .
JALA, 2010, 15 (03) :172-178
[3]   In situ electrodepo sited nanoparticles for facilitating electron transfer across self-assembled monolayers in biosensor design [J].
Du, Dan ;
Ding, Jiawang ;
Cai, Jie ;
Zhang, Jianming ;
Liu, Li .
TALANTA, 2008, 74 (05) :1337-1343
[4]  
EEC Directive, 1980, OFFICIAL J EUROPEAN
[5]   Rapid immunoanalytical method for the determination of atrazine residues in olive oil [J].
Garcés-García, M ;
Morais, S ;
Gonzaález-Martínez, MA ;
Puchades, R ;
Maquieira, A .
ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2004, 378 (02) :484-489
[6]   Development of a highly sensitive enzyme-linked immunosorbent assay for atrazine - Performance evaluation by flow injection immunoassay [J].
Gascon, J ;
Oubina, A ;
Ballesteros, B ;
Barcelo, D ;
Camps, F ;
Marco, MP ;
GonzalezMartinez, MA ;
Morais, S ;
Puchades, R ;
Maquieira, A .
ANALYTICA CHIMICA ACTA, 1997, 347 (1-2) :149-162
[7]  
Gascon J., 1997, ACS SYM SER, V657, P254
[8]  
González-Martínez MA, 1999, ANAL CHIM ACTA, V386, P201, DOI 10.1016/S0003-2670(99)00032-X
[9]   Atrazine analysis using an impedimetric immunosensor based on mixed biotinylated self-assembled monolayer [J].
Hleli, S ;
Martelet, C ;
Abdelghani, A ;
Burais, N ;
Jaffrezic-Renault, N .
SENSORS AND ACTUATORS B-CHEMICAL, 2006, 113 (02) :711-717
[10]   Breakthrough volumes increased by the addition of salt in the on-line solid-phase extraction and liquid chromatography of pesticides in environmental water [J].
MartinEsteban, A ;
Fernandez, P ;
Camara, C .
INTERNATIONAL JOURNAL OF ENVIRONMENTAL ANALYTICAL CHEMISTRY, 1996, 63 (02) :127-135