Optimization of experimental parameters for packed column supercritical fluid chromatography

被引:9
作者
Roman, John M. [1 ]
Abbott, Eric [1 ]
Xu, Xia [1 ]
Fox, Stephen D. [1 ]
Veenstra, Timothy D. [1 ]
Issaq, Haleem J. [1 ]
机构
[1] NCI, SAIC Frederick Inc, Adv Technol program, Lab Proteom & Analyt Technol, Frederick, MD 21702 USA
基金
美国国家卫生研究院;
关键词
optimization; experimental parameters; packed column; supercritical fluid chromatography;
D O I
10.1080/10826070701435046
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The present research is a study of experimental parameters that affect the resolution of a test mixture of estrogens by packed column supercritical fluid chromatography. The parameters that were evaluated included fluid flow rate, effect on resolution and retention times of the type and concentration of four organic modifiers, namely methanol, ethanol, isopropanol, and acetonitrile. Also, the effect of column type on the resolution of a mixture of estrogen metabolites was studied. Two packing materials cyanopropyl silica and Betasil diol silica were selected for this study. The following percentages of each of the packing materials were used in columns connected in series for the separation of the test mixture:100.0% CPS, 37.5% CPS/62.5% Diol, 50% CPS/50% Diol, 62.5 CPS/32.5% Diol, and 100.0% Diol. The results indicated that connecting two columns having the same dimensions (a ratio of one-to-one CPS/Diol) in series gave the best resolution. There was no effect on retention times or resolution when the two columns were reversed, i.e., column order has no effect. Also, retention times were a function of the organic modifier; methanol gave the shortest retention times, with approximately the same separation factor as the other three modifiers.
引用
收藏
页码:2037 / 2044
页数:8
相关论文
共 8 条
[1]   Derivatization of ethinylestradiol with dansyl chloride to enhance electrospray ionization: Application in trace analysis of ethinylestradiol in rhesus monkey plasma [J].
Anari, MR ;
Bakhtiar, R ;
Zhu, B ;
Huskey, S ;
Franklin, RB ;
Evans, DC .
ANALYTICAL CHEMISTRY, 2002, 74 (16) :4136-4144
[2]  
BERGER TA, 1997, SUPERCRITICAL FLUID, P19
[3]   NILE RED AS A SOLVATOCHROMIC DYE FOR MEASURING SOLVENT STRENGTH IN NORMAL LIQUIDS AND MIXTURES OF NORMAL LIQUIDS WITH SUPERCRITICAL AND NEAR CRITICAL FLUIDS [J].
DEYE, JF ;
BERGER, TA ;
ANDERSON, AG .
ANALYTICAL CHEMISTRY, 1990, 62 (06) :615-622
[4]  
ISSAQ HJ, 1983, J LIQ CHROMATOGR, V6, P59
[5]  
KLESPER E, 1962, J ORG CHEM, V27, P700
[6]  
SHHOENMAKERS PJ, 1986, OPITIMIZATION CHROMA
[7]   Analysis of fifteen estrogen metabolites using packed column supercritical fluid chromatography-mass spectrometry [J].
Xu, X ;
Roman, JM ;
Veenstra, TD ;
Van Anda, J ;
Ziegler, RG ;
Issaq, HJ .
ANALYTICAL CHEMISTRY, 2006, 78 (05) :1553-1558
[8]   Measuring fifteen endogenous estrogens simultaneously in human urine by high-performance liquid chromatography-mass spectrometry [J].
Xu, X ;
Veenstra, TD ;
Fox, SD ;
Roman, JM ;
Issaq, HJ ;
Falk, R ;
Saavedra, JE ;
Keefer, LK ;
Ziegler, RG .
ANALYTICAL CHEMISTRY, 2005, 77 (20) :6646-6654