Label-free whole blood cell differentiation based on multiple frequency AC impedance and light scattering analysis in a micro flow cytometer

被引:39
作者
Simon, Peter [1 ]
Frankowski, Marcin [1 ]
Bock, Nicole [1 ]
Neukammer, Joerg [1 ]
机构
[1] Phys Tech Bundesanstalt, Abbestr 2-12, D-10587 Berlin, Germany
关键词
SEPARATION; CLASSIFICATION; SPECTROSCOPY; COUNTS;
D O I
10.1039/c6lc00128a
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
We developed a microfluidic sensor for label-free flow cytometric cell differentiation by combined multiple AC electrical impedance and light scattering analysis. The measured signals are correlated to cell volume, membrane capacity and optical properties of single cells. For an improved signal to noise ratio, the microfluidic sensor incorporates two electrode pairs for differential impedance detection. One-dimensional sheath flow focusing was implemented, which allows single particle analysis at kHz count rates. Various monodisperse particles and differentiation of leukocytes in haemolysed samples served to benchmark the microdevice applying combined AC impedance and side scatter analyses. In what follows, we demonstrate that AC impedance measurements at selected frequencies allow label-free discrimination of platelets, erythrocytes, monocytes, granulocytes and lymphocytes in whole blood samples involving dilution only. Immunofluorescence staining was applied to validate the results of the label-free cell analysis. Reliable differentiation and enumeration of cells in whole blood by AC impedance detection have the potential to support medical diagnosis for patients with haemolysis resistant erythrocytes or abnormally sensitive leucocytes, i.e. for patients suffering from anaemia or leukaemia.
引用
收藏
页码:2326 / 2338
页数:13
相关论文
共 60 条
[1]  
[Anonymous], J CLIN DIAGN RES
[2]  
[Anonymous], LAB HEMATOL
[3]  
[Anonymous], SYSMEX J INT
[4]   The good, the bad, and the tiny: a review of microflow cytometry [J].
Ateya, Daniel A. ;
Erickson, Jeffrey S. ;
Howell, Peter B., Jr. ;
Hilliard, Lisa R. ;
Golden, Joel P. ;
Ligler, Frances S. .
ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2008, 391 (05) :1485-1498
[5]   Quality counts: new parameters in blood cell counting [J].
Briggs, C. ;
Mellors, I. ;
Roderick, A. ;
Ward, A. ;
O'Malley, C. ;
Barker, J. ;
De La Salle, B. ;
McTaggart, P. ;
Hyde, K. ;
Machin, S. J. .
INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2009, 31 (03) :277-297
[6]   Microfluidic Impedance Flow Cytometry Enabling High-Throughput Single-Cell Electrical Property Characterization [J].
Chen, Jian ;
Xue, Chengcheng ;
Zhao, Yang ;
Chen, Deyong ;
Wu, Min-Hsien ;
Wang, Junbo .
INTERNATIONAL JOURNAL OF MOLECULAR SCIENCES, 2015, 16 (05) :9804-9830
[7]   An integrated dielectrophoretic chip for continuous bioparticle filtering, focusing, sorting, trapping, and detecting [J].
Cheng, I-Fang ;
Chang, Hsien-Chang ;
Hou, Diana ;
Chang, Hsueh-Chia .
BIOMICROFLUIDICS, 2007, 1 (02)
[8]   Microfluidic Impedance-Based Flow Cytometry [J].
Cheung, Karen C. ;
Di Berardino, Marco ;
Schade-Kampmann, Grit ;
Hebeisen, Monika ;
Pierzchalski, Arkadiusz ;
Bocsi, Jozsef ;
Mittag, Anja ;
Tarnok, Attila .
CYTOMETRY PART A, 2010, 77A (07) :648-666
[9]   Continuous blood cell separation by hydrophoretic filtration [J].
Choi, Sungyoung ;
Song, Seungjeong ;
Choi, Chulhee ;
Park, Je-Kyun .
LAB ON A CHIP, 2007, 7 (11) :1532-1538
[10]  
Coulter W. H., 1956, Proceedings of the National Electronics Conference, V12, P1034