Performance of the cesium iodide calorimeter at the KTeV experiment at Fermilab

被引:4
作者
Prasad, V [1 ]
机构
[1] Univ Chicago, Enrico Fermi Inst, Chicago, IL 60637 USA
关键词
D O I
10.1016/S0168-9002(00)01239-0
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We describe the KTeV purr cesium iodide calorimeter and explain the calibration of the detector. The calorimeter has a global energy resolution better than 0.75% and an energy non-linearity less than 0.5% between 3 and 75 GeV. We discuss the performance of the calorimeter during the data-taking runs in 1996 and 1997 and briefly comment on the performance in 1999. (C) 2001 Published by Elsevier Science B.V.
引用
收藏
页码:341 / 343
页数:3