FLUORESCENCE TEMPERATURE SYSTEM DESIGN BASED ON SYNTHETIC OXIDE CRYSTAL

被引:1
作者
Bajsic, I. [1 ]
Krizan, J. [2 ]
Mozina, J.
机构
[1] Univ Ljubljana, Fac Mech Engn, Lab Measurement Proc Engn, SI-1000 Ljubljana, Slovenia
[2] AMI DOO, Ptuj, Slovenia
关键词
fluorescence; fluorescence intensity ratio; fluorescence lifetime; oxide crystal; temperature sensors; THERMOMETRY;
D O I
10.1080/10739149.2010.508320
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This article provides a description of the system for measuring temperature with Cr-doped crystals and a USB spectrometer. The same principle of measurement may also be used for other thermographic materials characterized by multiple spectral lines of different temperature sensitivities. The article presents some of the results obtained in the development of new generation temperature sensors that apply fluorescent synthetic oxide crystals. Such sensors have potential uses in biomedicine and in highly exacting industrial applications. Research involved some temperature optical characteristics of synthetic oxide crystals with an Al2O3 basis, such as Cr3+MgAl2O4(Cr:Spinel) and Cr3+Y3Al5O12(Cr:YAG), but also other variations doped with rare earth oxides. The method of measuring fluorescence intensity ratio and fluorescence lifetime is examined in detail. The article also provides a description of meter equipment used in measuring the fluorescence intensity ratio and fluorescence lifetime. Computer hardware and software for processing generated measurement signals are catalogued. A short review of obtained results and potential uses of such sensors in the field of temperature metrology is given in the final section of the article.
引用
收藏
页码:283 / 294
页数:12
相关论文
共 18 条
[1]  
ALLISON SW, LED INDUCED FLUORESC
[2]   Improved probe geometry for fluorescence-based fibre-optic temperature sensor [J].
Babnik, A ;
Kobe, A ;
Kuzman, D ;
Bajsic, I ;
Mozina, J .
SENSORS AND ACTUATORS A-PHYSICAL, 1996, 57 (03) :203-207
[3]  
Dewitt D.P.N., 1988, Theory and Practice of Radiation Thermometry
[4]  
FALCENBERG R, 1975, Patent No. 3876382
[5]   Investigations on exponential lifetime measurements for fluorescence thermometry [J].
Fernicola, VC ;
Rosso, L ;
Galleano, R ;
Sun, T ;
Zhang, ZY ;
Grattan, KTV .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2000, 71 (07) :2938-2943
[6]  
Grattan KTV., 1995, FIBER OPTIC FLUORESC
[7]   Two-colour phosphor thermometry for surface temperature measurement [J].
Heyes, AL ;
Seefeldt, S ;
Feist, JP .
OPTICS AND LASER TECHNOLOGY, 2006, 38 (4-6) :257-265
[8]  
KANEDA M, 2006, P M BUS KOR OCT 18 2, P1586
[9]  
KANO Y, 2007, 2006 SICE ICASE INT, P1660
[10]   Thermographic phosphors for high temperature measurements: Principles, current state of the art and recent applications [J].
Khalid, Ashiq Hussain ;
Kontis, Konstantinos .
SENSORS, 2008, 8 (09) :5673-5744