Progress in determination of the area function of indenters used for nanoindentation

被引:121
作者
Herrmann, K
Jennett, NM
Wegener, W
Meneve, J
Hasche, K
Seemann, R
机构
[1] Phys Tech Bundesanstalt, D-38116 Braunschweig, Germany
[2] Natl Phys Lab, Teddington TW11 0LW, Middx, England
[3] Vlaamse Instelling Technol Onderzoek, B-2400 Mol, Belgium
关键词
indentation method; indenter; scanning force microscopy; area function;
D O I
10.1016/S0040-6090(00)01367-5
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
It is generally accepted that the most significant source of uncertainty in nanoindentation measurement is the geometry of the indenter tip. In this paper results are presented on the determination of indenter area functions by two methods: (1) direct determination from co-ordinate measurements obtained using a scanning force microscope (SFM); (2) indentation into reference materials with known Young's modulus and Poisson ratio. The direct measurement of geometry of a Berkovich indenter was determined using a traceably calibrated Thermomicroscopes 'Autoprobe CP' SFM that of other mentioned indenters (Berkovich and ball shaped indenters) was determined with a modified SFM Veritekt-3. The modification of the latter is that the movement of the SFM scanner is directly calibrated by three laser interferometers fitted in the co-ordinate axes x, y and z, thus realizing directly traceable measurements. Recently, the original arrangement of the laser-interferometers was improved such that an Abbe error is minimized. The paper reports the results achieved with this arrangement and analyses the uncertainty of the area functions obtained. Good knowledge of the machine compliance is a second prerequisite for the accurate determination of the mechanical properties of thin films, in particular elastic modulus. Using the indenter area calculated from SFM measurements, an iterative procedure of high load indentation into a stiff reference material is assessed. In the alternative method, improvements in the determination of the area function by means of indentation into reference materials, based on the Oliver and Pharr method (J. Mater. Res. 7 (1992) 1564-1583), is reported. Due to the uncertainty of the thus determined area functions essentially depending on the uncertainty of the machine compliance, a combined iterative approach using two reference materials is described. The machine compliance again was determined by high load indentation into a stiff reference material (as in the first method). The indenter area function, however, is derived from another reference material allowing accurate stiffness measurements at shallow depths. This allows measurement of the area function near the indenter tip. Both methods are compared to assess the accuracy of the indentation method for determining the indenter area function and the machine compliance. (C) 2000 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:394 / 400
页数:7
相关论文
共 13 条
  • [1] A metrological scanning force microscope used for coating thickness and other topographical measurements
    Bienias, M
    Gao, S
    Hasche, K
    Seemann, R
    Thiele, K
    [J]. APPLIED PHYSICS A-MATERIALS SCIENCE & PROCESSING, 1998, 66 (Suppl 1): : S837 - S842
  • [2] *CEN, TC1884 CEN
  • [3] A method for interpreting the data from depth-sensing indentation instruments
    Doerner, M. F.
    Nix, W. D.
    [J]. JOURNAL OF MATERIALS RESEARCH, 1986, 1 (04) : 601 - 609
  • [4] *EUR COMM SMT, DETERMINATION HARDNE
  • [5] Field J.E., 1999, YOUNG MODULUS POISSO
  • [6] HASCHE K, 1999, P LAS METR 1999 FLOR
  • [7] HASCHE K, 2000, P QUANT MICR 2000 SE, P219
  • [8] *ISO CD, 145771 ISOCD
  • [9] JAGER G, 1999, P LAS METR 1999 FLOR
  • [10] Depth sensing indentation of thin hard films: A study of modulus measurement sensitivity to indentation parameters
    Jennett, NM
    Meneve, J