Accuracy and reproducibility of low dose insulin administration using pen-injectors and syringes

被引:60
作者
Gnanalingham, MG
Newland, P
Smith, CP
机构
[1] Univ Manchester, Booth Hall Childrens Hosp, Dept Paediat, Manchester M9 2AA, Lancs, England
[2] Univ Manchester, Booth Hall Childrens Hosp, Dept Clin Biochem, Manchester M9 2AA, Lancs, England
关键词
pen-injector device; insulin syringes; insulin administration; low dose insulin;
D O I
10.1136/adc.79.1.59
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
Many children with diabetes require small doses of insulin administered with syringes or pen-injector devices (at the Booth Hall Paediatric Diabetic Clinic, 20% of children aged 0-5 years receive 1-2 U insulin doses). To determine how accurately and reproducibly small doses are delivered, 1, 2, 5, and 10 U doses of soluble insulin (100 U/ml) were dispensed in random order 15 times from five new NovoPens (1.5 mi), five ED-Pens (1.5 mi), and by five nurses using 30 U syringes. Each dose was weighed, and intended and actual doses compared. The two pen-injectors delivered less insulin than syringes, differences being inversely proportional to dose. For 1 U (mean (SD)): 0.89 (0.04) U (NovoPen), 0.92 (0.03) U (BD-Pen), 1.23 (0.09)U (syringe); and for 10 U: 9.8 (0.1) U (NovoPen), 9.9 (0.1)U (ED-Pen), 10.1 (0.1)U (syringe). The accuracy (percentage errors) of the pen-injectors was similar and more accurate than syringes delivering 1, 2, and 5 U of insulin. Errors for 1 U: 11(4)% (NovoPen), 8(3)% (BD-Pen), 23(9)% (syringe). The reproducibility (coefficient of variation) of actual doses was similar (< 7%) for all three devices, which were equally consistent at underdosing (pen-injectors) or overdosing (syringes) insulin. All three devices, especially syringes, are unacceptably inaccurate when delivering 1 U doses of insulin. Patients on low doses need to be educated that their dose may alter when they transfer from one device to another.
引用
收藏
页码:59 / 62
页数:4
相关论文
共 13 条
[1]   DOSAGE ACCURACY OF SELF-MIXED VS PREMIXED INSULIN [J].
BELL, DSH ;
CLEMENTS, RS ;
PERENTESIS, G ;
RODDAM, R ;
WAGENKNECHT, L .
ARCHIVES OF INTERNAL MEDICINE, 1991, 151 (11) :2265-2269
[2]  
CASELLA SJ, 1993, PEDIATRICS, V91, P1155
[3]   EFFECTS OF ACCIDENTAL INTRAMUSCULAR INJECTION ON INSULIN ABSORPTION IN IDDM [J].
FRID, A ;
GUNNARSSON, R ;
GUNTNER, P ;
LINDE, B .
DIABETES CARE, 1988, 11 (01) :41-45
[4]   FACTORS INFLUENCING THE ABSORPTION, SERUM-INSULIN CONCENTRATION, AND BLOOD-GLUCOSE RESPONSES AFTER INJECTIONS OF REGULAR INSULIN AND VARIOUS INSULIN MIXTURES [J].
GALLOWAY, JA ;
SPRADLIN, CT ;
NELSON, RL ;
WENTWORTH, SM ;
DAVIDSON, JA ;
SWARNER, JL .
DIABETES CARE, 1981, 4 (03) :366-376
[5]   Rising incidence of insulin dependent diabetes in children aged under 5 years in the Oxford region: time trend analysis [J].
Gardner, SG ;
Bingley, PJ ;
Sawtell, PA ;
Weeks, S ;
Gale, EAM ;
Bell, RAF ;
Dunger, DB ;
Mukhtar, A ;
OMalley, BP ;
Silk, BR ;
Smith, EH ;
Scott, RDM ;
Latham, PJ ;
Lakhani, PK ;
Paton, RC ;
Ackland, FM ;
Fox, CJ ;
Griffin, NK ;
Matthews, DR ;
Neil, HAW ;
Mann, NP ;
Simpson, HCR ;
Brown, RS ;
Knight, AH ;
Cowen, JM ;
Pearce, JC ;
Cheetham, CH ;
Gallen, IW ;
Sandler, L ;
Westcott, T .
BRITISH MEDICAL JOURNAL, 1997, 315 (7110) :713-717
[6]   THE KINETICS OF INSULIN ADMINISTRATION BY INSULIN PENS [J].
GINSBERG, BH ;
PARKES, JL ;
SPARACINO, C .
HORMONE AND METABOLIC RESEARCH, 1994, 26 (12) :584-587
[7]  
Gold A. E., 1995, P351
[8]   AN ASSESSMENT OF THE ACCURACY OF NOVOPEN-1 DELIVERY AFTER PROLONGED USE [J].
GORDON, D ;
WILSON, M ;
PATERSON, KR ;
SEMPLE, CG .
DIABETIC MEDICINE, 1990, 7 (04) :364-366
[9]  
KOLENDORF K, 1978, EUR J CLIN PHARMACOL, V14, P117, DOI 10.1007/BF00607442
[10]   Subcutaneous or intramuscular injections of insulin in children [J].
Polak, M ;
Beregzaszi, M ;
Belarbi, N ;
Benali, K ;
Hassan, M ;
Czernichow, P ;
TubianaRufi, N .
DIABETES CARE, 1996, 19 (12) :1434-1436