A predictive tool for the self-management of diabetes (Librae): evaluation using a continuous glucose monitoring system

被引:3
作者
Franklin, VL
Wilson, AW
Butler, RA
Greene, SA
机构
[1] Univ Dundee, Ninewells Hosp & Med Sch, Dundee DD1 9SY, Scotland
[2] Robert Gordon Univ, Sch Engn, Aberdeen AB9 1FR, Scotland
[3] Robert Gordon Univ, Sch Comp, Aberdeen AB9 1FR, Scotland
[4] Librae Ltd, Aberdeen AB10 1FR, Scotland
关键词
diabetes; decision support tools; compartmental models; glucose sensors;
D O I
10.1111/j.1464-5491.2005.01770.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Aims Diabetes self-management involves a difficult balancing act between insulin, food and exercise. The challenge is to develop innovative, validated algorithms to aid patient decision-making and optimize glycaemic control. 'Librae' is a computerized diabetes simulator in diary format, developed as an educational predictive tool for patients, reducing 'trial and error' by allowing patients to simulate and experiment with dietary or insulin adjustments on a 'body double'. We have evaluated the predictive ability of Librae using continuous blood glucose monitoring (CGMS). Methods Patients with Type 1 Diabetes attending the Paediatric Clinic were invited to use 'Librae' for 1 week and were then fitted with a CGMS for 72 h. The predictive ability of 'Librae' was compared with concurrent data obtained from the CGMS. Results Seven thousand nine hundred and sixty paired blood glucose values were obtained from the 11 patients who completed the study. 'Librae' underestimated the measured CGMS values, the error having a positive mean of 0.35 mmol/l (95% confidence interval 0.22-0.48 mmol/l). However, Librae tended to overestimate at low levels of blood glucose readings, and underestimate at high levels of blood glucose readings. Conclusion The modelled values of 'Librae' correlated well with the CGMS data, but clinically unacceptable errors occurred at extremes of blood glucose levels. Concurrent CGMS recordings have provided a large data set to modify and improve the existing Librae model and patient feedback has led to improvements in its usability. Librae may provide a useful tool to improve diabetes self-management education and optimize glycaemic control.
引用
收藏
页码:21 / 25
页数:5
相关论文
共 50 条
  • [31] Continuous glucose monitoring versus self-monitoring of blood glucose in the management of cystic fibrosis related diabetes: A systematic review and meta-analysis
    Kumar, Shanal
    Soldatos, Georgia
    Ranasinha, Sanjeeva
    Teede, Helena
    Pallin, Michael
    JOURNAL OF CYSTIC FIBROSIS, 2023, 22 (01) : 39 - 49
  • [32] A Modified User-Oriented Heuristic Evaluation of a Mobile Health System for Diabetes Self-management Support
    Georgsson, Mattias
    Staggers, Nancy
    Weir, Charlene
    CIN-COMPUTERS INFORMATICS NURSING, 2016, 34 (02) : 77 - 84
  • [33] Empowering Diabetes Self-management by Gamifying Intelligent System: myTrybeCare Concept
    Bouras, Adam
    Usop, Kat
    Popescu, Mihail
    2018 IEEE 4TH MIDDLE EAST CONFERENCE ON BIOMEDICAL ENGINEERING (MECBME), 2018, : 137 - 140
  • [34] The LMC Skills, Confidence & Preparedness Index (SCPI): development and evaluation of a novel tool for assessing self-management in patients with diabetes
    Mbuagbaw, Lawrence
    Aronson, Ronnie
    Walker, Ashleigh
    Brown, Ruth E.
    Orzech, Naomi
    HEALTH AND QUALITY OF LIFE OUTCOMES, 2017, 15
  • [35] Continuous Glucose Monitoring Integration for Remote Diabetes Management: Virtual Diabetes Care with Case Studies
    Carlson, Anders L.
    Martens, Thomas W.
    Johnson, Libby
    Criego, Amy B.
    DIABETES TECHNOLOGY & THERAPEUTICS, 2021, 23 : S56 - S65
  • [36] A Personalized Recommendation System to Support Diabetes Self-Management for American Indians
    Alian, Shadi
    Li, Juan
    Pandey, Vikram
    IEEE ACCESS, 2018, 6 : 73041 - 73051
  • [37] A mHealth Architecture for Diabetes Self-Management System Research-in-Progress
    El-Gayar, Omar
    Timsina, Prem
    Nawar, Nevine
    AMCIS 2013 PROCEEDINGS, 2013,
  • [38] Does continuous glucose monitoring have clinical utility in contemporary management of diabetes?
    Cameron, FJ
    Ambler, GR
    JOURNAL OF PAEDIATRICS AND CHILD HEALTH, 2004, 40 (03) : 79 - 84
  • [39] Continuous glucose monitoring in persons with type 2 diabetes not using insulin
    Cowart, Kevin
    Updike, Wendy H.
    Franks, Rachel
    EXPERT REVIEW OF MEDICAL DEVICES, 2021, 18 (11) : 1049 - 1055
  • [40] The LMC Skills, Confidence & Preparedness Index (SCPI): development and evaluation of a novel tool for assessing self-management in patients with diabetes
    Lawrence Mbuagbaw
    Ronnie Aronson
    Ashleigh Walker
    Ruth E. Brown
    Naomi Orzech
    Health and Quality of Life Outcomes, 15