A quantum dot-based optical immunosensor for human serum albumin detection

被引:64
作者
Tu, Meng-Che [1 ]
Chang, Yun-Tzu [1 ]
Kang, Yu-Ting [1 ]
Chang, Hwan-You [2 ]
Chang, Pin [3 ]
Yew, Tri-Rung [1 ]
机构
[1] Natl Tsing Hua Univ, Dept Mat Sci & Engn, Hsinchu 30013, Taiwan
[2] Natl Tsing Hua Univ, Dept Life Sci, Hsinchu 30013, Taiwan
[3] ITRI S, Microsyst Technol Ctr, Ind Technol Res Inst, Tainan 70955, Taiwan
关键词
Human serum albumin; Quantum dot; Optical immunosensor; Photodiode; 3-Aminopropyltriethoxysilane; Diode laser; C-REACTIVE PROTEIN; IMMUNOASSAY; NANOPARTICLES; ELECTRODES; BIOSENSOR; MYOGLOBIN;
D O I
10.1016/j.bios.2011.11.035
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
In this study, a CdSe/ZnS quantum dot (QD)-based immunosensor using a simple optical system for human serum albumin (HSA) detection is developed. Monoclonal anti-HSA (AHSA) immobilized on 3-aminopropyltriethoxysilane (APTES)-modified glass was used to capture HSA specifically. Bovine serum albumin (BSA) was used to block non-specific sites. The solution, containing AHSA-QD complex prepared by mixing biotinylated polyclonal anti-HSA and streptavidin coated QD, was used to conjugate with the HSA molecules captured on AHSA/BSA/APTES-modified glass for the modification of HSA with QD. A simple optical system, comprising a diode laser (405 nm), an optical lens, a 515-nm-long pass filter, and an Si-photodiode, was used to detect fluorescence and convert it to photocurrent. The current intensity was determined by the amount of QD specifically conjugated with HSA, and was therefore HSA-concentration-dependent and could be used to quantify HSA concentration. The detection limit of the pure QD solution was similar to 3.5 x 10(-12) M, and the detection limit for the CdSe/ZnS QD-based immunosensor developed in this study was approximately 3.2 x 10(-5) mg/ml. This small optical biosensing system shows considerable potential for future applications of on-chip liver-function detection. (c) 2011 Elsevier B.V All rights reserved.
引用
收藏
页码:286 / 290
页数:5
相关论文
共 22 条
[1]   SERRS immunoassay for quantitative human CRP analysis [J].
Campbell, Fiona M. ;
Ingram, Andrew ;
Monaghan, Paul ;
Cooper, Jon ;
Sattar, Naveed ;
Eckersall, P. David ;
Graham, Duncan .
ANALYST, 2008, 133 (10) :1355-1357
[2]   Microfluidic electrochemical immunoarray for ultrasensitive detection of two cancer biomarker proteins in serum [J].
Chikkaveeraiah, Bhaskara V. ;
Mani, Vigneshwaran ;
Patel, Vyomesh ;
Gutkind, J. Silvio ;
Rusling, James F. .
BIOSENSORS & BIOELECTRONICS, 2011, 26 (11) :4477-4483
[3]   An electrochemical tyrosinase-immobilized biosensor for albumin - toward a potential total protein measurement [J].
Chuang, MC ;
Liu, CC ;
Yang, MC .
SENSORS AND ACTUATORS B-CHEMICAL, 2006, 114 (01) :357-363
[4]   On-chip detection of myoglobin based on fluorescence [J].
Darain, Farzana ;
Yager, Paul ;
Gan, Kai Ling ;
Tjin, Swee Chuan .
BIOSENSORS & BIOELECTRONICS, 2009, 24 (06) :1744-1750
[5]  
De M, 2009, NAT CHEM, V1, P461, DOI [10.1038/NCHEM.334, 10.1038/nchem.334]
[6]   GLUCOSE POTENTIOMETRIC ELECTRODES BASED ON MEDIATORLESS BIOELECTROCATALYSIS - A NEW APPROACH [J].
GHINDILIS, AL ;
KUROCHKIN, IN .
BIOSENSORS & BIOELECTRONICS, 1994, 9 (4-5) :353-357
[7]   An electrochemical albumin-sensing system utilizing microfluidic technology [J].
Huang, Chao-June ;
Lu, Chiu-Chun ;
Lin, Thong-Yueh ;
Chou, Tse-Chuan ;
Lee, Gwo-Bin .
JOURNAL OF MICROMECHANICS AND MICROENGINEERING, 2007, 17 (04) :835-842
[8]   A review of molecular recognition technologies for detection of biological threat agents [J].
Iqbal, SS ;
Mayo, MW ;
Bruno, JG ;
Bronk, BV ;
Batt, CA ;
Chambers, JP .
BIOSENSORS & BIOELECTRONICS, 2000, 15 (11-12) :549-578
[9]   Label free novel electrical detection using micromachined PZT monolithic thin film cantilever for the detection of C-reactive protein [J].
Lee, JH ;
Yoon, KH ;
Hwang, KS ;
Park, J ;
Ahn, S ;
Kim, TS .
BIOSENSORS & BIOELECTRONICS, 2004, 20 (02) :269-275
[10]   Determination of albumin concentration by MIP-QCM sensor [J].
Lin, TY ;
Hu, CH ;
Chou, TC .
BIOSENSORS & BIOELECTRONICS, 2004, 20 (01) :75-81