Method for Testing Motion Analysis Laboratory Measurement Systems

被引:3
作者
Hakkarainen, Marko J. [1 ]
Bragge, Timo [1 ]
Liikavainio, Tuomas [2 ]
Arokoski, Jari [2 ]
Karjalainen, Pasi A. [1 ]
Tarvainen, Mika [1 ]
机构
[1] Univ Eastern Finland, Dept Math & Phys, Kuopio, Finland
[2] Kuopio Univ Hosp, Dept Phys & Rehabil Med, SF-70210 Kuopio, Finland
来源
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME | 2010年 / 132卷 / 11期
关键词
ACCELEROMETERS; GAIT; WALKING;
D O I
10.1115/1.4002368
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
This paper proposes a method for comparing data from accelerometers, optical based 3D motion capture systems, and force platforms (FPs) in the context of spatial and temporal differences. Testing method is based on the motion laboratory accreditation test (MLAT), which can be used to test FP and camera based motion capture components of a motion analysis laboratory. This study extends MLAT to include accelerometer data. Accelerometers were attached to a device similar to the MLAT rod. The elevation of the rod from the plane of the floor is computed and compared with the force platform vector orientation and the rod orientation obtained by optical motion capture system. Orientation of the test device is achieved by forming nonlinear equation group, which describes the components of the measured accelerations. Solution for this equation group is estimated by using the Gauss-Newton method. This expanded MLAT procedure can be used in the laboratory setting were either FP, camera based motion capture, or any other motion capture system is used along with accelerometer measurements. [DOI:10.1115/1.4002368]
引用
收藏
页数:5
相关论文
共 13 条
[1]  
[Anonymous], 2010, DISCRETE TIME SIGNAL
[2]   Human movement analysis using stereophotogrammetry - Part 1: theoretical background [J].
Cappozzo, A ;
Della Croce, U ;
Leardini, A ;
Chiari, L .
GAIT & POSTURE, 2005, 21 (02) :186-196
[3]   A proposed test to support the clinical movement analysis laboratory accreditation process [J].
Holden, JP ;
Selbie, WS ;
Stanhope, SJ .
GAIT & POSTURE, 2003, 17 (03) :205-213
[4]  
Lawson C.L., 1995, SOLVING LEAST SQUARE
[5]   Reproducibility of loading measurements with skin-mounted accelerometers during walking [J].
Liikavainio, Tuomas ;
Bragge, Timo ;
Hakkarainen, Marko ;
Jurvelin, Jukka S. ;
Karjalainen, Pasi A. ;
Arokoski, Jari P. .
ARCHIVES OF PHYSICAL MEDICINE AND REHABILITATION, 2007, 88 (07) :907-915
[6]   Standing balance evaluation using a triaxial accelerometer [J].
Mayagoitia, RE ;
Lötters, JC ;
Veltink, PH ;
Hermens, H .
GAIT & POSTURE, 2002, 16 (01) :55-59
[7]   Trunk accelerometry as a measure of balance control during quiet standing [J].
Moe-Nilssen, R ;
Helbostad, JL .
GAIT & POSTURE, 2002, 16 (01) :60-68
[8]   A new method for evaluating motor control in gait under real-life environmental conditions. Part 2: Gait analysis [J].
Moe-Nilssen, R .
CLINICAL BIOMECHANICS, 1998, 13 (4-5) :328-335
[9]   Automated estimation of initial and terminal contact timing using accelerometers; Development and validation in transtibial amputees and controls [J].
Selles, RW ;
Formanoy, MAG ;
Bussmann, JBJ ;
Janssens, PJ ;
Stam, HJ .
IEEE TRANSACTIONS ON NEURAL SYSTEMS AND REHABILITATION ENGINEERING, 2005, 13 (01) :81-88
[10]   LOWER-EXTREMITY ANGLE MEASUREMENT WITH ACCELEROMETERS - ERROR AND SENSITIVITY ANALYSIS [J].
WILLEMSEN, ATM ;
FRIGO, C ;
BOOM, HBK .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1991, 38 (12) :1186-1193