A wide-acceptance Compton spectrometer for spectral characterization of a medical x-ray source

被引:9
作者
Espy, Michelle A. [1 ]
Gehring, A. [1 ]
Belian, A. [1 ]
Haines, T. [1 ]
Hunter, J. [1 ]
James, M. [1 ]
Klasky, M. [1 ]
Mendez, J. [1 ]
Moir, D. [1 ]
Sedillo, R. [1 ]
Shurter, R. [1 ]
Stearns, J. [1 ]
Van Syoc, K. [1 ]
Volegov, P. [1 ]
机构
[1] Los Alamos Natl Lab, POB 1663, Los Alamos, NM 87545 USA
来源
MEDICAL IMAGING 2016: PHYSICS OF MEDICAL IMAGING | 2016年 / 9783卷
关键词
Spectral characterization; Bremsstrahlung spectra; Compton spectrometer;
D O I
10.1117/12.2216269
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
Accurate knowledge of the x-ray spectra used in medical treatment and radiography is important for dose calculations and material decomposition analysis. Indirect measurements via transmission through materials are possible. However, such spectra are challenging to measure directly due to the high photon fluxes. One method of direct measurement is via a Compton spectrometer (CS) method. In this approach, the x-rays are converted to a much lower flux of electrons via Compton scattering on a converter foil (typically beryllium or aluminum). The electrons are then momentum selected by bending in a magnetic field. With tight angular acceptance of electrons into the magnet of similar to 1 deg, there is a linear correlation between incident photon energy and electron position recorded on an image plate. Here we present measurements of Bremsstrahlung spectrum from a medical therapy machine, a Scanditronix M22 Microtron. Spectra with energy endpoints from 6 to 20 MeV are directly measured, using a CS with a wide energy range from 0.5 to 20 MeV. We discuss the sensitivity of the device and the effects of converter material and collimation on the accuracy of the reconstructed spectra. Approaches toward improving the sensitivity, including the use of coded apertures, and potential future applications to characterization of spectra are also discussed.
引用
收藏
页数:11
相关论文
共 15 条
[1]  
[Anonymous], IEEE NUCL SCI S
[2]  
Berry Phil, COMMUNICATION
[3]   Absolute calibration of image plates for electrons at energy between 100 keV and 4 MeV [J].
Chen, Hui ;
Back, Norman L. ;
Bartal, Teresa ;
Beg, F. N. ;
Eder, David C. ;
Link, Anthony J. ;
MacPhee, Andrew G. ;
Ping, Yuan ;
Song, Peter M. ;
Throop, Alan ;
Van Woerkom, Linn .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2008, 79 (03)
[4]   CT scanner x-ray spectrum estimation from transmission measurements [J].
Duan, Xinhui ;
Wang, Jia ;
Yu, Lifeng ;
Leng, Shuai ;
McCollough, Cynthia H. .
MEDICAL PHYSICS, 2011, 38 (02) :993-997
[5]   ACHROMATIC MAGNETIC MIRROR FOR ION BEAMS [J].
ENGE, HA .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1963, 34 (04) :385-&
[6]  
Gehring A. E., 2014, P SPIE RAD DETECTORS, V9215
[7]  
Goorley J.T, 2013, LA-UR- 13-22934
[8]  
Hawkins A.R., LAUR933544
[9]  
Katsuyuki T., 2013, MED PHYS, V40
[10]   A COMPTON-SCATTERING SPECTROMETER FOR DETERMINING X-RAY PHOTON ENERGY-SPECTRA [J].
MATSCHEKO, G ;
RIBBERFORS, R .
PHYSICS IN MEDICINE AND BIOLOGY, 1987, 32 (05) :577-594