Two-Micron Continuous-Wave Laser-Assisted Neuroendoscopy: Clinical Experience of Two Institutions in 524 Procedures

被引:5
|
作者
Schuhmann, Martin U. [1 ]
Kural, Cahit [1 ,2 ]
Lalla, Lisanne [1 ]
Ebner, Florian H. [1 ]
Bock, Christoph [3 ]
Ludwig, Hans-Christoph [3 ]
机构
[1] Eberhard Karls Univ Hosp Tubingen, Dept Neurosurgery, Div Pediat Neurosurg, Tubingen, Germany
[2] Hlth Sci Univ, Gulhane Educ & Res Hosp, Dept Neurosurg, Ankara, Turkey
[3] Georg August Univ Hosp Gottingen, Dept Neurosurg, Div Pediat Neurosurg, Gottingen, Germany
关键词
Cyst fenestration; Endoscopic third ventriculostomy; Laser; Neuroendoscopy; Septostomy; ENDOSCOPIC 3RD VENTRICULOSTOMY; ND-YAG LASER; COLLOID CYSTS; CONTACT LASER; DIODE-LASER; 3RD-VENTRICLE; COMPLICATIONS; SERIES;
D O I
10.1016/j.wneu.2018.09.064
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
OBJECTIVE: To present the clinical experience of 2 neurosurgical centers with the use of a 2-micron continuous-wave laser (2 mu-cwL) system as standard tool in neuroendoscopic procedures and to discuss the safety and efficacy of this system. METHODS: In total, 469 patients underwent neuroendoscopic procedures using 2 mu-cwL between September 2009 and January 2015. All patient data were retrospectively reviewed. In total, 241 (51%) patients were children and 228 (49%) adults. Mean age was 27.5 years (range: 3 days to 83 years). Intraoperative ultrasonography or neuronavigation were used to guide ventricular or cyst puncture and for intraventricular or intracystic orientation if necessary. RESULTS: A total of 524 neuroendoscopic procedures using 2 mu-cwL were performed. Laser-assisted endoscopic third ventriculostomy was the most common procedure in 302 (64%) patients. Cyst fenestration was performed in 124 (26%), septostomy in 45, tumor biopsy in 41, tumor resection in 8, and choroid plexus coagulation in 3 patients. There was no intraoperative complication directly attributable to the use of laser and an overall procedural complication rate of 4.8%. CONCLUSIONS: This large series of 2 mu-cwL as a routine tool in neuroendoscopic procedures demonstrates that 2 mu-cwL is safe for endoscopic third ventriculostomy, septostomy, cyst fenestration, and intraventricular tumor biopsy or resection. As a cutting and coagulation tool, it combines the action of mechanical tools like forceps, balloons, and scissors plus those of electric tools. It therefore renders neuroendoscopic procedures more straightforward with a minimum need to change tools.
引用
收藏
页码:E81 / E88
页数:8
相关论文
共 50 条
  • [1] Hysteroscopic myomectomy using a two-micron continuous wave laser (RevoLix)
    Chen, Ching-Hui
    Lee, Wen-Ling
    Wang, I-Te
    Yen, Yuan-Kuei
    Chiu, Li-Hsuan
    Tzeng, Chii-Ruey
    Liu, Wei-Min
    GYNECOLOGY AND MINIMALLY INVASIVE THERAPY-GMIT, 2013, 2 (03): : 89 - 92
  • [2] Continuous-wave laser-assisted injection of single magnetic nanobeads into living cells
    Zhong, Jing
    Liu, Hengjun
    Maruyama, Hisataka
    Masuda, Taisuke
    Arai, Fumihito
    SENSORS AND ACTUATORS B-CHEMICAL, 2016, 230 : 298 - 305
  • [3] Continuous-wave two-photon terahertz quantum cascade laser
    Khabibullin, R. A.
    Ushakov, D. V.
    Afonenko, A. A.
    Pavlov, A. Yu.
    Galiev, R. R.
    Ponomarev, D. S.
    Vasilyev, A. P.
    Kuzmenkov, A. G.
    Maleev, N. A.
    Zubov, F. I.
    Maksimov, M. V.
    Belov, D. A.
    Ikonnikov, A. V.
    Kuritsyn, D. I.
    Zhukavin, R. Kh.
    Kovalevsky, K. A.
    Anfertev, V. A.
    Vaks, V. L.
    Antonov, A. V.
    Dubinov, A. A.
    Morozov, S. V.
    Gavrilenko, V. I.
    JOURNAL OF APPLIED PHYSICS, 2024, 136 (19)
  • [4] Two micron continuous wave laser vaporesection for the treatment of benign prostatic hyperplasia
    Fu, Wei-Jun
    Hong, Bao-Fa
    Yang, Yong
    Gao, Jiang-Ping
    Zhang, Lei
    Cai, Wei
    Chen, Yao-Fu
    Wang, Xiao-Xiong
    ASIAN JOURNAL OF ANDROLOGY, 2008, 10 (02) : 341 - 342
  • [5] Continuous-wave solid-state two-Stokes Raman laser
    Grabtchikov, A. S.
    Lisinetskii, V. A.
    Orlovich, V. A.
    Schmitt, M.
    Schluecker, S.
    Kuestner, B.
    Kiefer, W.
    QUANTUM ELECTRONICS, 2009, 39 (07) : 624 - 626
  • [6] A Two Micron Continuous Wave Laser System Is Safe and Effective for Peritoneal Access & High Precision Dissection During NOTES® Procedures
    Dray, Xavier
    Donatelli, Gianfranco
    Krishnamurty, Devi Mukkai
    Wroblewski, Ronald J.
    Assumpcao, Lia R.
    Giday, Samuel A.
    Buscaglia, Jonathan M.
    Shin, Eun Ji
    Dubcenco, Elena
    Magno, Priscilla
    Pipitone, Laurie J.
    Marohn, Michael R.
    Kantsevoy, Sergey V.
    Kalloo, Anthony N.
    GASTROINTESTINAL ENDOSCOPY, 2009, 69 (05) : AB162 - AB162
  • [7] Continuous-wave diode laser thermokeratoplasty:: First clinical experience in blind human eyes
    Geerling, G
    Koop, N
    Brinkmann, R
    Tüngler, A
    Wirbelauer, C
    Birngruber, R
    Laqua, H
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1999, 25 (01): : 32 - 40
  • [8] CONTINUOUS-WAVE CO2 LASER-ASSISTED HIGH-TEMPERATURE SUPERCONDUCTING FILM DEPOSITION BY EXCIMER LASER ABLATION
    VONDERBURG, E
    GRILL, W
    DIEGEL, M
    STAFAST, H
    MATERIALS SCIENCE AND ENGINEERING B-SOLID STATE MATERIALS FOR ADVANCED TECHNOLOGY, 1992, 13 (01): : 25 - 28
  • [9] Clinical Outcomes in the First Two Years of Femtosecond Laser-Assisted Cataract Surgery
    Carifi, Gianluca
    Zygoura, Vasiliki
    Kopsachilis, Nikolaos
    AMERICAN JOURNAL OF OPHTHALMOLOGY, 2015, 160 (02) : 396 - 397
  • [10] Clinical Outcomes in the First Two Years of Femtosecond Laser-Assisted Cataract Surgery
    Chee, Soon-Phaik
    Yang, Younian
    Ti, Seng-El
    AMERICAN JOURNAL OF OPHTHALMOLOGY, 2015, 159 (04) : 714 - 719