Respiratory snorkel and valve system for breath-by-breath gas analysis in swimming

被引:56
作者
Keskinen, KL
Rodríguez, FA
Keskinen, OP
机构
[1] Univ Jyvaskyla, Dept Biol Phys Activ, Neuromuscular Res Ctr, FIN-40014 Jyvaskyla, Finland
[2] Univ Barcelona, Dept Biomed Sci, Inst Nacl Educ Fis Catalunya, Barcelona, Spain
关键词
oxygen uptake; ventilation; gas analysis; respiratory valve;
D O I
10.1034/j.1600-0838.2003.00319.x
中图分类号
G8 [体育];
学科分类号
04 ; 0403 ;
摘要
The present study aimed to compare a standard facemask (CM) and a newly modified swimming snorkel and valve system (SV) for breath-by-breath (B x B) gas analysis (K4 b(2), Cosmed, Rome, Italy), and to validate the system under controlled laboratory conditions before being used in swimming. Nine healthy males performed two bouts of a stepwise exercise on an electrically braked stationary bicycle on separate days. Ventilatory and gas exchange parameters were analyzed using the same B x B portable system, with subjects breathing alternatively through the two different valves. Agreement between both methods was evaluated by Passing-Bablok regression analysis. The gas exchange values measured using the SV were highly correlated with those obtained using the CM (R-2 values 40.9). However, differences existed between the two series of measurements so that most ventilatory and gas exchange parameters were lower (3-7%) with the SV. The error when using the SV device was mainly systematic along the whole range of measurement. Accordingly, linear regression equations were developed to further improve the accuracy of the measures when using the SV. Therefore, the modified respiratory SV system can be considered as a valid device for collecting expired gas for B x B analysis, comparable to the standard facemask, with the advantage of being suitable for measurements during swimming.
引用
收藏
页码:322 / 329
页数:8
相关论文
共 20 条
[1]  
*AM COLL SPORTS ME, 1995, ACSMS GUID EX TEST P, P69
[2]   Measuring agreement in method comparison studies [J].
Bland, JM ;
Altman, DG .
STATISTICAL METHODS IN MEDICAL RESEARCH, 1999, 8 (02) :135-160
[3]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[4]   A NEW RESPIRATORY VALVE SYSTEM FOR MEASURING OXYGEN-UPTAKE DURING SWIMMING [J].
DALMONTE, A ;
SARDELLA, F ;
ALIPPI, B ;
FAINA, M ;
MANETTA, A .
EUROPEAN JOURNAL OF APPLIED PHYSIOLOGY AND OCCUPATIONAL PHYSIOLOGY, 1994, 69 (02) :159-162
[5]   The VO2 slow component in swimming [J].
Demarie, S ;
Sardella, F ;
Billat, V ;
Magini, W ;
Faina, M .
EUROPEAN JOURNAL OF APPLIED PHYSIOLOGY, 2001, 84 (1-2) :95-99
[6]  
DIPRAMPE.PE, 1974, J APPL PHYSIOL, V37, P1
[7]   Field testing of VO2peak in cross-country skiers with portable breath-by-breath system [J].
Doyon, KH ;
Perrey, S ;
Abe, D ;
Hughson, RL .
CANADIAN JOURNAL OF APPLIED PHYSIOLOGY-REVUE CANADIENNE DE PHYSIOLOGIE APPLIQUEE, 2001, 26 (01) :1-11
[8]  
Hausswirth C, 1997, INT J SPORTS MED, V18, P449
[9]   OXYGEN-UPTAKE DURING SWIMMING IN MAN [J].
HOLMER, I .
JOURNAL OF APPLIED PHYSIOLOGY, 1972, 33 (04) :502-&
[10]   RELIABILITY OF MEASUREMENT OF OXYGEN-UPTAKE BY A PORTABLE TELEMETRIC SYSTEM [J].
KAWAKAMI, Y ;
NOZAKI, D ;
MATSUO, A ;
FUKUNAGA, T .
EUROPEAN JOURNAL OF APPLIED PHYSIOLOGY AND OCCUPATIONAL PHYSIOLOGY, 1992, 65 (05) :409-414