Development of a lactate biosensor based on conducting copolymer bound lactate oxidase

被引:91
|
作者
Suman, S
Singhal, R
Sharma, AL
Malthotra, BD
Pundir, CS [1 ]
机构
[1] Maharshi Dayanand Univ, Dept Biosci, Biochem Res Lab, Rohtak 124001, Haryana, India
[2] Natl Phys Lab, Biomol Elect & Conducting Polymer Res Grp, New Delhi 110012, India
关键词
lactate; lactate oxidase; immobilization; polyaniline-co-fluoroaniline; lactate biosensor; serum;
D O I
10.1016/j.snb.2004.12.016
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An amperometric enzyme electrode was developed for determination of lactate in serum. To prepare this electrode, commercial lactate oxidase from Pediococcus species has been immobilized through glutaraldehyde coupling onto polyaniline-co-fluoroaniline film deposited on an Indium tin oxide (ITO) coated glass plate. This plate acted as working electrode when combined with Pt electrode as counter electrode to the electrometer for the development of a biosensor. The method is based on generation of electrons from H2O2 which is formed from lactic acid by immobilized lactate oxidase. The concentration of lactic acid is directly proportional to the current measured. The enzyme electrode showed optimum response when operated at 42 degrees C in 0.05 M, sodium phosphate buffer pH 6.5 for 50s, The biosensor showed a good performance with a linear response range from 0.1 to 5.5 mM/l. The minimum detection limit of the method is 0.1 mM/l and sensitivity of the sensor is 1.18 mu A/mM/l lactate. This electrode was employed for determination of lactate in serum. The serum values in healthy and diseased persons were in the range 0.51-2.9 and 5.0-15.0 mM. respectively. The analytical recovery of added lactic acid was 71%. Within batch and between batch CV were < 4 and < 5%. respectively. Among the various serum substances tested only 8-hydroxyquinoline, urea, ammonium molybdate and uric acid caused 64, 38, 34 and 31% inhibition, respectively. The electrode Was used 150 times over 26 days without any considerable loss of activity, when stored at 4 degrees C. (c) 2005 Elsevier B.V. All rights reserved.
引用
收藏
页码:768 / 772
页数:5
相关论文
共 50 条
  • [1] Development of Lactate Biosensor based on Electro statically Functionalized Graphene oxide bound Lactate oxidase
    Gera, Meeta
    Jain, V. K.
    Suman
    PHYSICS OF SEMICONDUCTOR DEVICES, 2014, : 413 - 415
  • [2] The development of reagentless lactate biosensor based on a novel conducting polymer
    Haccoun, J.
    Piro, B.
    Noel, V.
    Pham, M. C.
    BIOELECTROCHEMISTRY, 2006, 68 (02) : 218 - 226
  • [3] Lactate biosensor based on the adsorption of polyelectrolyte stabilized lactate oxidase into porous conductive carbon
    Gavalas, VG
    Chaniotakis, NA
    MIKROCHIMICA ACTA, 2001, 136 (3-4) : 211 - 215
  • [4] Design and characterization of a lactate biosensor based on immobilized lactate oxidase onto gold surfaces
    Parra, A
    Casero, E
    Vázquez, L
    Pariente, F
    Lorenzo, E
    ANALYTICA CHIMICA ACTA, 2006, 555 (02) : 308 - 315
  • [5] Lactate Biosensor Based on the Adsorption of Polyelectrolyte Stabilized Lactate Oxidase into Porous Conductive Carbon
    Vasilis G. Gavalas
    Nikolas A. Chaniotakis
    Microchimica Acta, 2001, 136 : 211 - 215
  • [6] Determination of serum lactate with alkylamine glass bound lactate oxidase
    Suman
    Pundir, CS
    INDIAN JOURNAL OF BIOCHEMISTRY & BIOPHYSICS, 2005, 42 (03): : 186 - 189
  • [7] L-lactate selective impedimetric bienzymatic biosensor based on lactate dehydrogenase and pyruvate oxidase
    Chan, Deni
    Barsan, Madalina M.
    Korpan, Yaroslav
    Brett, Christopher M. A.
    ELECTROCHIMICA ACTA, 2017, 231 : 209 - 215
  • [8] Development of an amperometric biosensor for lactate
    Lillis, B
    Grogan, C
    Berney, H
    Lane, B
    1ST ANNUAL INTERNATIONAL IEEE-EMBS SPECIAL TOPIC CONFERENCE ON MICROTECHNOLOGIES IN MEDICINE & BIOLOGY, PROCEEDINGS, 2000, : 394 - 399
  • [9] Lactate biosensor based on a bionanocomposite composed of titanium oxide nanoparticles, photocatalytically reduced graphene, and lactate oxidase
    Elena Casero
    Concepción Alonso
    María Dolores Petit-Domínguez
    Luis Vázquez
    Ana María Parra-Alfambra
    Pablo Merino
    Susana Álvarez-García
    Alicia de Andrés
    Edna Suárez
    Félix Pariente
    Encarnación Lorenzo
    Microchimica Acta, 2014, 181 : 79 - 87
  • [10] Design and optimization of a lactate amperometric biosensor based on lactate oxidase cross-linked with polymeric matrixes
    Romero, Marcelo Ricardo
    Garay, Fernando
    Baruzzi, Ana M.
    SENSORS AND ACTUATORS B-CHEMICAL, 2008, 131 (02) : 590 - 595