Intra-operative navigation of knee kinematics and the influence of osteoarthritis

被引:5
作者
Massin, Philippe [1 ]
Boyer, Patrick [1 ]
Hajage, David [2 ]
Kilian, Pascal [1 ]
Tubach, Florence [2 ]
机构
[1] Paris Diderot Univ, Dept Orthoped Surg, Bichat Hosp, F-75877 Paris 18, France
[2] Hop Xavier Bichat, Dept Clin Res, Paris, France
关键词
Knee kinematics; Computer-assisted surgery; Total knee replacement; Knee osteoarthritis; Cadaver knee; POSTERIOR CONDYLAR OFFSET; TIBIOFEMORAL MOVEMENT; LIVING KNEE; FLUOROSCOPIC ANALYSIS; MULTICENTER ANALYSIS; ARTHROPLASTY; MRI; FLEXION; ROTATION; SHAPES;
D O I
10.1016/j.knee.2010.06.007
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Intra-operative assessment of knee kinematics should optimise implantation of total knee arthroplasties. The purpose of this work was to validate the data delivered by an adapted navigation system in 10 healthy cadaver knees and to investigate the kinematics of 10 osteoarthritic (OA) knees in patients undergoing total knee replacement. The system displayed the magnitude of axial rotation, the position of the instantaneous centre of axial rotation and the displacements of the condyles. Successive cycles from full extension to 140 degrees of flexion in the same knee produced a mean external rotation of 20 degrees +/- 10 degrees, which was correlated to knee flexion (r=0.6 +/- 0.2 in healthy knees, r=0.8 +/- 0.2 in OA knees). The centre of axial rotation migrated posteriorly an average of 8.2 mm in both groups. The posterior displacements were 4 mm +/- 5 mm in healthy and 5 mm +/- 6 mm in OA knees for the medial condyle, and 21 mm +/- 9 mm in healthy and 21 mm 10 mm in OA knees for the lateral condyle. The medial condyle lifted off beyond 110 degrees of flexion. Results in healthy knees were consistent with those reported in the current literature. The kinematics of healthy and of OA knees with an intact anterior cruciate ligament did not differ significantly. (C) 2010 Elsevier B.V. All rights reserved.
引用
收藏
页码:259 / 264
页数:6
相关论文
共 22 条
[1]   The international knee documentation committee subjective knee evaluation form - Normative data [J].
Anderson, AF ;
Irrgang, JJ ;
Kocher, MS ;
Mann, BJ ;
Harrast, JJ .
AMERICAN JOURNAL OF SPORTS MEDICINE, 2006, 34 (01) :128-135
[2]   Altered knee kinematics in ACL-deficient non-copers: A comparison using dynamic MRI [J].
Barrance, PJ ;
Williams, GN ;
Snyder-Mackler, L ;
Buchanan, TS .
JOURNAL OF ORTHOPAEDIC RESEARCH, 2006, 24 (02) :132-140
[3]   Fluoroscopic analysis of the kinematics of deep flexion in total knee arthroplasty - Influence of posterior condylar offset [J].
Bellemans, J ;
Banks, S ;
Victor, J ;
Vandenneucker, H ;
Moemans, A .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2002, 84B :50-53
[4]   THE ENVELOPE OF PASSIVE KNEE-JOINT MOTION [J].
BLANKEVOORT, L ;
HUISKES, R ;
DELANGE, A .
JOURNAL OF BIOMECHANICS, 1988, 21 (09) :705-+
[5]  
Dennis DA, 1998, CLIN ORTHOP RELAT R, P47
[6]   In vivo determination of normal and anterior cruciate ligament-deficient knee kinematics [J].
Dennis, DA ;
Mahfouz, MR ;
Komistek, RD ;
Hoff, W .
JOURNAL OF BIOMECHANICS, 2005, 38 (02) :241-253
[7]   A multicenter analysis of axial femorotibial rotation after total knee arthroplasty [J].
Dennis, DA ;
Komistek, RD ;
Mahfouz, MR ;
Walker, SA ;
Tucker, A .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2004, (428) :180-189
[8]  
Dennis DA, 1996, CLIN ORTHOP RELAT R, P107
[9]   The movement of the normal tibio-femoral joint [J].
Freeman, MAR ;
Pinskerova, V .
JOURNAL OF BIOMECHANICS, 2005, 38 (02) :197-208
[10]   The movement of the knee studied by magnetic resonance imaging [J].
Freeman, MAR ;
Pinskerova, V .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2003, (410) :35-43