Diagnostic accuracy of R-wave detection by insertable cardiac monitors

被引:19
作者
De Coster, Margot [1 ]
Demolder, Anthony [1 ]
De Meyer, Veerle [1 ]
Vandenbulcke, Filiep [1 ]
Van Heuverswyn, Frederic [1 ]
De Pooter, Jan [1 ]
机构
[1] Ghent Univ Hosp, Dept Cardiol, C Heymanslaan 10, B-9000 Ghent, Belgium
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 2020年 / 43卷 / 05期
关键词
arrhythmia detection; implantable loop recorder; insertable cardiac monitor; oversensing; R-wave; undersensing; IMPLANTABLE LOOP RECORDER;
D O I
10.1111/pace.13912
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background Insertable cardiac monitors (ICM) allow prolonged rhythm monitoring, but the diagnostic performance can be hampered by false positive arrhythmia alerts related to inadequate R-wave sensing. This study assesses the prevalence and predictors of inadequate R-wave sensing (both over- and undersensing) among different ICM types. Methods Patients implanted with an ICM at Ghent University Hospital between January 2017 and August 2018 were included. ICM tracings recorded at interrogation or transmitted by remote monitoring were reviewed for inadequate R-wave sensing leading to false arrhythmia alerts. Patient and implant characteristics were retrieved from the medical records and implant reports. Results The study screened 135 patients (age 59 +/- 19 years, 44% female) implanted with different ICM types: Reveal LINQ (TM) and XT (Medtronic): n = 92 (68%), Confirm and Confirm Rx (Abbott): n = 35 (26%), and BioMonitor 2 (Biotronik): n = 8 (6%). ICM tracings were analyzed in 112 patients (83%). False arrhythmia alerts occurred in 22 (20%) patients, most frequently related to undersensing (77%). False diagnosis of bradycardia or pause was documented in 64%, false high ventricular rates in 14%, and false atrial fibrillation alerts in 22%. Occurrence of R-wave changes was not related to patient characteristics or implant R-wave sensing. A trend toward higher number of inadequate R-wave sensing seems to occur with nonparasternal implant sites (P = .074). Conclusions False arrhythmia alerts due to inadequate R-wave sensing occurred in 20% of ICM patients independent of implant features and patient characteristics.
引用
收藏
页码:511 / 517
页数:7
相关论文
共 10 条
[1]   Implantable loop recorder in clinical practice [J].
Bisignani, Antonio ;
De Bonis, Silvana ;
Mancuso, Luigi ;
Ceravolo, Gianluca ;
Bisignani, Giovanni .
JOURNAL OF ARRHYTHMIA, 2019, 35 (01) :25-32
[2]  
Brignole M, 2018, EUR HEART J, V39, P1883, DOI 10.1093/eurheartj/ehy037
[3]   Indications for the use of diagnostic implantable and external ECG loop recorders [J].
Brignole, Michele ;
Vardas, Panos ;
Hoffman, Ellen ;
Huikuri, Heikki ;
Moya, Angel ;
Ricci, Renato ;
Sulke, Neil ;
Wieling, Wouter ;
Auricchio, Angelo ;
Lip, Gregory Y. H. ;
Almendral, Jesus ;
Kirchhof, Paulus ;
Aliot, Etienne ;
Gasparini, Maurizio ;
Braunschweig, Frieder .
EUROPACE, 2009, 11 (05) :671-687
[4]   Feasibility of Remote Monitoring Using a Novel Long-Dipole Insertable Cardiac Monitor [J].
Ciconte, Giuseppe ;
Vicedomini, Gabriele ;
Giacopelli, Daniele ;
Calovic, Zarko ;
Conti, Manuel ;
Spinelli, Bernardo ;
Saviano, Massimo ;
Santinelli, Vincenzo ;
Pappone, Carlo .
JACC-CLINICAL ELECTROPHYSIOLOGY, 2018, 4 (04) :559-561
[5]   An Anatomic-Based Approach for the Placement of Implantable Loop Recorders [J].
Grubb, Blair P. ;
Welch, Marlene ;
Kanjwal, Khalil ;
Karabin, Beverly ;
Kanjwal, Yousuf .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2010, 33 (09) :1149-1152
[6]  
Kirchhof P, 2016, EUROPACE, V18, DOI [10.5603/KP.2016.0172, 10.1016/j.rec.2016.11.033, 10.1093/europace/euw295]
[7]   R-Wave Sensing in an Implantable Cardiac Monitor without ECG-Based Preimplant Mapping: Results from a Multicenter Clinical Trial [J].
Krahn, Andrew D. ;
Pickett, Robert A. ;
Sakaguchi, Scott ;
Shaik, Naushad ;
Cao, Jian ;
Norman, Holly S. ;
Guerrero, Patricia .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2014, 37 (04) :505-511
[8]   Clinical impact, safety, and accuracy of the remotely monitored implantable loop recorder Medtronic Reveal LINQ™ [J].
Maines, Massimiliano ;
Zorzi, Alessandro ;
Tomasi, Giancarlo ;
Angheben, Carlo ;
Catanzariti, Domenico ;
Piffer, Lucio ;
Del Greco, Maurizio .
EUROPACE, 2018, 20 (06) :1050-1057
[9]   Diagnostic and therapeutic value of implantable loop recorder: A tertiary care center experience [J].
Padmanabhan, Deepak ;
Kancharla, Krishna ;
El-Harasis, Majd A. ;
Isath, Ameesh ;
Makkar, Nayani ;
Noseworthy, Peter A. ;
Friedman, Paul A. ;
Cha, Yong-Mei ;
Kapa, Suraj .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2019, 42 (01) :38-45
[10]   Improving sensing and detection performance in subcutaneous monitors [J].
van Dam, Peter ;
van Groeningen, Chris ;
Houben, Richard P. M. ;
Hampton, David R. .
JOURNAL OF ELECTROCARDIOLOGY, 2009, 42 (06) :580-583