Electrochemiluminescence (ECL)-Based Electrochemical Imaging Using a Massive Array of Bipolar Ultramicroelectrodes

被引:64
作者
Anderson, Todd J. [1 ]
Defnet, Peter A. [1 ]
Zhang, Bo [1 ]
机构
[1] Univ Washington, Dept Chem, Seattle, WA 98195 USA
基金
美国国家卫生研究院; 美国国家科学基金会;
关键词
ELECTROGENERATED CHEMILUMINESCENCE; ELECTRODE ARRAY; MICROELECTRODE; MICROSCOPY;
D O I
10.1021/acs.analchem.0c00921
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
In this report, we describe the fabrication, characterization, and use of a massive array of closed bipolar ultra-microelectrodes (UMEs) in electrochemical imaging applications. The bipolar UME array is 1 cm(2) in size and contains >146 000 carbon electrodes embedded in a 15 mu m thick insulating and freestanding membrane of Parylene C. Structural characterization with optical and electron microscopies shows that the carbon UMEs are highly uniform in size, shape, and interelectrode spacing. The bipolar UME array was used in electrochemical imaging to probe highly dynamic redox processes in which the reduction of redox molecules on one side the array is electrically coupled to an oxidative electrochemilumescence (ECL) process on the opposite side. This allows one to simultaneously monitor electrochemical reactions on hundreds of thousands of individual electrodes with millisecond temporal resolution. Our results suggest that microfabricated closed bipolar UME arrays can be useful for imaging fast and transient electrochemical processes in which scanning probe methods are inapplicable due to their limited temporal resolution.
引用
收藏
页码:6748 / 6755
页数:8
相关论文
共 48 条
[1]   Scanning Electrochemical Microscopy [J].
Amemiya, Shigeru ;
Bard, Allen J. ;
Fan, Fu-Ren F. ;
Mirkin, Michael V. ;
Unwin, Patrick R. .
ANNUAL REVIEW OF ANALYTICAL CHEMISTRY, 2008, 1 (01) :95-131
[2]   A wireless electrochemiluminescence detector applied to direct and indirect detection for electrophoresis on a microfabricated glass device [J].
Arora, A ;
Eijkel, JCT ;
Morf, WE ;
Manz, A .
ANALYTICAL CHEMISTRY, 2001, 73 (14) :3282-3288
[3]   Visualization and Modeling of the hydrodynamics of an impinging microjet [J].
Bitziou, E ;
Rudd, NC ;
Edwards, MA ;
Unwin, PR .
ANALYTICAL CHEMISTRY, 2006, 78 (05) :1435-1443
[4]   Scanning Ion Conductance Microscopy [J].
Chen, Chiao-Chen ;
Zhou, Yi ;
Baker, Lane A. .
ANNUAL REVIEW OF ANALYTICAL CHEMISTRY, VOL 5, 2012, 5 :207-228
[5]   In Situ Imaging Facet-Induced Spatial Heterogeneity of Electrocatalytic Reaction Activity at the Subparticle Level via Electrochemiluminescence Microscopy [J].
Chen, Ying ;
Zhao, Dongbo ;
Fu, Jiaju ;
Gou, Xiaodan ;
Jiang, Dechen ;
Dong, Hao ;
Zhu, Jun-Jie .
ANALYTICAL CHEMISTRY, 2019, 91 (10) :6829-6835
[6]   Wireless electrochemical DNA microarray sensor [J].
Chow, Kwok-Fan ;
Mavre, Francois ;
Crooks, Richard M. .
JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 2008, 130 (24) :7544-+
[7]   A Large-Scale, Wireless Electrochemical Bipolar Electrode Microarray [J].
Chow, Kwok-Fan ;
Mavre, Francois ;
Crooks, John A. ;
Chang, Byoung-Yong ;
Crooks, Richard M. .
JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 2009, 131 (24) :8364-+
[8]   Steady-State Voltammetry of a Microelectrode in a Closed Bipolar Cell [J].
Cox, Jonathan T. ;
Guerrette, Joshua P. ;
Zhang, Bo .
ANALYTICAL CHEMISTRY, 2012, 84 (20) :8797-8804
[9]   Electrogenerated Chemiluminescence in Submicrometer Wells for Very High-Density Biosensing [J].
Cui, Chen ;
Jin, Rong ;
Jiang, Dechen ;
Zhang, Jianrong ;
Zhu, Jun-Jie .
ANALYTICAL CHEMISTRY, 2020, 92 (01) :578-582
[10]   Detection of Transient Nanoparticle Collision Events Using Electrochemiluminescence on a Closed Bipolar Microelectrode [J].
Defnet, Peter A. ;
Zhang, Bo .
CHEMELECTROCHEM, 2020, 7 (01) :252-259