Open-Field Mouse Brain PET: Design Considerations and Detector Development

被引:0
|
作者
Kyme, Andre Z. [1 ,2 ,3 ]
Gong, Kuang [1 ]
Judenhofer, Martin S. [1 ]
Bec, Julien [1 ]
Du, Junwei [1 ]
Qi, Jinyi [1 ]
Cherry, Simon R. [1 ]
Meikle, Steven R. [2 ,3 ]
机构
[1] Univ Calif Davis, Dept Biomed Engn, Davis, CA 95616 USA
[2] Univ Sydney, Fac Hlth Sci, Sydney, NSW, Australia
[3] Univ Sydney, Brain & Mind Ctr, Sydney, NSW, Australia
来源
2015 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC) | 2015年
基金
澳大利亚研究理事会;
关键词
Positron emission tomography; awake animal imaging; scanner simulation; depth-of-interaction detectors;
D O I
暂无
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
'Open-field' PET imaging offers the key capability of correlating functional changes in the brain of an awake animal with its behavioral response to environmental or pharmacologic challenges. Previously we have demonstrated the feasibility of this concept for rats using motion compensation techniques. That system, however, is not suitable for imaging the mouse brain due to limitations imposed by the use of a commercial PET scanner. Therefore, we are designing a purpose-built PET scanner which optimizes the geometry, motion tracking and imaging performance for open-field imaging of the mouse brain. We simulated the sensitivity and spatial resolution performance of four candidate scanner designs: ring, parallel plate, and two box designs. The block detector was a 23x23 array comprising 0.785 x 0.785 x 20 mm(3) LSO crystals. A ML-EM reconstruction with DoI capability was used to determine the DoI resolution necessary to achieve approximately uniform and isotropic sub-millimeter spatial resolution throughout the FoV. The results showed that 3 mm DoI resolution was sufficient to achieve the required spatial resolution performance for all scanners except the parallel-plate design. However, the sensitivity advantage of the overlapping box design (peak absolute sensitivity of 16% and 36% improvement over the ring design) suggested this unconventional design is favored for imaging the mouse brain. We also designed a dual-ended readout DoI-encoding detector module based on a 6x6 array of through-silicon-via SiPM arrays to meet the DoI requirement. The next stage of work is to characterize the performance of this detector module and use it to build a prototype bench-top scanner for initial testing of the scanner concept. For the final scanner design we propose to slide the scanner axially on rails according to the animal's motion, rather than move the animal (as in our previous design). This will allow faster motion without disturbing the animal's behavior, and also achieve a large axial FoV for animal movement at minimal cost.
引用
收藏
页数:3
相关论文
共 50 条
  • [1] Open-field mouse brain PET: design optimisation and detector characterisation
    Kyme, Andre Z.
    Judenhofer, Martin S.
    Gong, Kuang
    Bec, Julien
    Selfridge, Aaron
    Du, Junwei
    Qi, Jinyi
    Cherry, Simon R.
    Meikle, Steven R.
    PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (15): : 6207 - 6225
  • [2] DoI Detector Design and Characterization for Open-Field Mouse Brain PET
    Kyme, Andre Z.
    Judenhofer, Martin S.
    Meikle, Steven R.
    Cherry, Simon R.
    2016 IEEE NUCLEAR SCIENCE SYMPOSIUM, MEDICAL IMAGING CONFERENCE AND ROOM-TEMPERATURE SEMICONDUCTOR DETECTOR WORKSHOP (NSS/MIC/RTSD), 2016,
  • [3] Motion-Adaptive Gantry Development for Open-Field Mouse PET
    Zhang, Xinwen
    Bec, Julien
    Cherry, Simon R.
    Meikle, Steven R.
    Kyme, Andre Z.
    2019 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2019,
  • [4] Design and validity of a miniaturized open-field aberrometer
    Bhatt, Uday K.
    Sheppard, Amy L.
    Shah, Sunil
    Dua, Harmindar Singh
    Mihashi, Toshifumi
    Yamaguchi, Tatsuo
    Wolffsohn, James S.
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2013, 39 (01): : 36 - 40
  • [5] Design of a compact open-field wavefront sensor
    Borjeson, Charlie
    Romashchenko, Dmitry
    Lundstrom, Linda
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2021, 62 (08)
  • [6] OPEN-FIELD BEHAVIOR IN 4 INBRED MOUSE STRAINS
    STRENG, J
    CANADIAN JOURNAL OF PSYCHOLOGY, 1971, 25 (01): : 62 - &
  • [7] Development of a TOF Detector for Brain PET
    Choe, Hyeok-jun
    Choi, Yong
    Jung, Jin Ho
    2014 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2014,
  • [8] EMC ANTENNA CALIBRATION AND THE DESIGN OF AN OPEN-FIELD SITE
    SALTER, MJ
    ALEXANDER, MJ
    MEASUREMENT SCIENCE AND TECHNOLOGY, 1991, 2 (06) : 510 - 519
  • [9] Mouse Behavior in the Open-field Test after Meloxicam Administration
    Antiorio, Ana T. F. B.
    Aleman-Laporte, Jilma
    Zanatto, Dennis A.
    Pereira, Marco A. A.
    Gomes, Mariana S. A. G.
    Wadt, Danilo
    Yamamoto, Pedro K.
    Bernardi, Maria M.
    Mori, M. C. Claudia
    JOURNAL OF THE AMERICAN ASSOCIATION FOR LABORATORY ANIMAL SCIENCE, 2022, 61 (03): : 270 - 274
  • [10] OPEN-FIELD BEHAVIOR IN GUINEA-PIGS - DEVELOPMENTAL AND ADAPTIVE CONSIDERATIONS
    SUAREZ, SD
    GALLUP, GG
    BEHAVIOURAL PROCESSES, 1982, 7 (03) : 267 - 274