A Biomechanical Comparison of Periprosthetic Femoral Fracture Fixation in Normal and Osteoporotic Cadaveric Bone

被引:45
作者
Demos, Harry A. [1 ]
Briones, Marcus S. [2 ]
White, Peter H. [1 ]
Hogan, Kathleen A. [3 ]
Barfield, William R. [1 ,4 ]
机构
[1] Med Univ S Carolina, Dept Orthopaed Surg, Charleston, SC 29425 USA
[2] OrthoCarolina, Monroe, NC USA
[3] New Hampshire Orthopaed Ctr, Amherst, NH USA
[4] Coll Charleston, Charleston, SC 29401 USA
关键词
periprosthetic; Vancouver B1 fracture; cadaver; biomechanics; fixation; SHAFT FRACTURES; PLATE FIXATION; HIP; COMPRESSION; CABLE; LOCKING; STEM; TIP;
D O I
10.1016/j.arth.2011.08.019
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Several techniques are described for fixation of Vancouver B1 femoral shaft fractures after total hip arthroplasty. Twenty-four femurs were scanned by dual x-ray absorptiometry scanned and matched for bone mineral density. Femurs were implanted with a cemented simulated total hip prosthesis with a simulated periprosthetic femur fracture distal to the stem. Fractures were fixed with Synthes (Paoli, Pa) 12-hole curved plates and 4 different constructs proximally. Each construct was loaded to failure in axial compression. Constructs with locking and nonlocking screws demonstrated equivalent loads at failure and were superior in load at failure compared with cables. Cable constructs failed proximally. No proximal failures occurred in specimens fixed with screws and cables. A combination of locked or nonlocked screws and supplemental cable fixation is recommended for the treatment of Vancouver B1 periprosthetic femur fractures.
引用
收藏
页码:783 / 788
页数:6
相关论文
共 29 条
[1]  
[Anonymous], ORTHOPAEDIC KNOWLEDG
[2]   Locking compression plate fixation of Vancouver type-b1 periprosthetic femoral fractures [J].
Buttaro, M. A. ;
Farfalli, G. ;
Nunez, M. Paredes ;
Comba, F. ;
Piccaluga, F. .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2007, 89A (09) :1964-1969
[3]  
Chandler H P, 1993, Semin Arthroplasty, V4, P99
[4]   THE COLLAGEN FIBRIL - A MODEL SYSTEM FOR STUDYING THE STAINING AND FIXATION OF A PROTEIN [J].
CHAPMAN, JA ;
TZAPHLIDOU, M ;
MEEK, KM ;
KADLER, KE .
ELECTRON MICROSCOPY REVIEWS, 1990, 3 (01) :143-182
[5]   Fresh-frozen vs. embalmed bone: is it possible to use formalin-fixed human bone for biomechanical experiments on implants? [J].
Comert, Ayhan ;
Kokat, Ali Murat ;
Akkocaoglu, Murat ;
Tekdemir, Ibrahim ;
Akca, Kivanc ;
Cehreli, Murat Cavit .
CLINICAL ORAL IMPLANTS RESEARCH, 2009, 20 (05) :521-525
[6]   Fixation of periprosthetic femoral shaft fractures: A biomechanical comparison of two techniques [J].
Dennis, MG ;
Simon, JA ;
Kummer, FJ ;
Koval, KJ ;
Di Cesare, PE .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2001, 15 (03) :177-180
[7]   Fixation of periprosthetic femoral shaft fractures occurring at the tip of the stem - A biomechanical study of 5 techniques [J].
Dennis, MG ;
Simon, JA ;
Kummer, FJ ;
Koval, KJ ;
DiCesare, PE .
JOURNAL OF ARTHROPLASTY, 2000, 15 (04) :523-528
[8]   Biomechanics of locked plates and screws [J].
Egol, KA ;
Kubiak, EN ;
Fulkerson, E ;
Kummer, FJ ;
Koval, KJ .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2004, 18 (08) :488-493
[9]  
EMERSON RH, 1992, CLIN ORTHOP RELAT R, P35
[10]   Fixation of periprosthetic femoral shaft fractures associated with cemented femoral stems - A biomechanical comparison of locked plating and conventional cable plates [J].
Fulkerson, E ;
Koval, K ;
Preston, CF ;
Iesaka, K ;
Kummer, FJ ;
Egol, KA .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2006, 20 (02) :89-93