An oil-bath-based 293 K to 473 K blackbody source

被引:10
作者
Fowler, JB
机构
[1] Natl. Inst. of Std. and Technology, Gaithersburg
关键词
aperture; blackbody; cavity; conical; emissivity; radiation; radiometry; reflectance; source; temperature; thermometer; oil bath;
D O I
10.6028/jres.101.062
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A high temperature oil-bath-based-blackbody source has been designed and constructed in the Radiometric Physics Division at the National Institute of Standards and Technology, Gaithersburg, MD. The goal of this work was to design a large aperture blackbody source with highly uniform radiance across the aperture, good temporal stability, and good reproducibility. This blackbody source operates in the 293 K to 473 K range with blackbody temperature combined standard uncertainties of 7.2 mK to 30.9 mK. The calculated emissivity of this source is 0.9997 with a standard uncertainty of 0.0003. With a 50 mm limiting aperture at the cavity entrance, the emissivity increases to 0.99996.