Extreme PCR: Efficient and Specific DNA Amplification in 15-60 Seconds

被引:158
作者
Farrar, Jared S. [1 ]
Wittwer, Carl T. [1 ]
机构
[1] Univ Utah, Dept Pathol, Sch Med, Salt Lake City, UT 84132 USA
关键词
POLYMERASE-CHAIN-REACTION; REAL-TIME PCR; TEMPERATURE;
D O I
10.1373/clinchem.2014.228304
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
BACKGROUND: PCR is a key technology in molecular biology and diagnostics that typically amplifies and quantifies specific DNA fragments in about an hour. However, the kinetic limits of PCR are unknown. METHODS: We developed prototype instruments to temperature cycle 1- to 5-mu L samples in 0.4-2.0 s at annealing/extension temperatures of 62 degrees C-76 degrees C and denaturation temperatures of 85 degrees C-92 degrees C. Primer and polymerase concentrations were increased 10- to 20-fold above typical concentrations to match the kinetics of primer annealing and polymerase extension to the faster temperature cycling. We assessed analytical specificity and yield on agarose gels and by high-resolution melting analysis. Amplification efficiency and analytical sensitivity were demonstrated by real-time optical monitoring. RESULTS: Using single-copy genes from human genomic DNA, we amplified 45- to 102-bp targets in 15-60 s. Agarose gels showed bright single bands at the expected size, and high-resolution melting curves revealed single products without using any "hot start" technique. Amplification efficiencies were 91.7%-95.8% by use of 0.8- to 1.9-s cycles with single-molecule sensitivity. A 60-bp genomic target was amplified in 14.7 s by use of 35 cycles. CONCLUSIONS: The time required for PCR is inversely related to the concentration of critical reactants. By increasing primer and polymerase concentrations 10- to 20-fold with temperature cycles of 0.4-2.0 s, efficient (>90%), specific, high-yield PCR from human DNA is possible in <15 s. Extreme PCR demonstrates the feasibility of while-you-wait testing for infectious disease, forensics, and any application where immediate results may be critical. (C) 2014 American Association for Clinical Chemistry
引用
收藏
页码:145 / 153
页数:9
相关论文
共 26 条
[1]  
Bustin S A., 2010, The PCR Revolution: Basic Technologies and Applications, V2nd
[2]   Ultra-rapid flow-through polymerase chain reaction microfluidics using vapor pressure [J].
Fuchiwaki, Yusuke ;
Nagai, Hidenori ;
Saito, Masato ;
Tamiya, Eiichi .
BIOSENSORS & BIOELECTRONICS, 2011, 27 (01) :88-94
[3]   Rapid PCR in a continuous flow device [J].
Hashimoto, M ;
Chen, PC ;
Mitchell, MW ;
Nikitopoulos, DE ;
Soper, SA ;
Murphy, MC .
LAB ON A CHIP, 2004, 4 (06) :638-645
[4]   KINETIC PCR ANALYSIS - REAL-TIME MONITORING OF DNA AMPLIFICATION REACTIONS [J].
HIGUCHI, R ;
FOCKLER, C ;
DOLLINGER, G ;
WATSON, R .
BIO-TECHNOLOGY, 1993, 11 (09) :1026-1030
[5]   Noncontact infrared-mediated thermocycling for effective polymerase chain reaction amplification of DNA in nanoliter volumes [J].
Hühmer, AFR ;
Landers, JP .
ANALYTICAL CHEMISTRY, 2000, 72 (21) :5507-5512
[6]   Chemical amplification: Continuous-flow PCR on a chip [J].
Kopp, MU ;
de Mello, AJ ;
Manz, A .
SCIENCE, 1998, 280 (5366) :1046-1048
[7]   Exploring the limits of ultrafast polymerase chain reaction using liquid for thermal heat exchange: A proof of principle [J].
Maltezos, George ;
Johnston, Matthew ;
Taganov, Konstantin ;
Srichantaratsamee, Chutatip ;
Gorman, John ;
Baltimore, David ;
Chantratita, Wasun ;
Scherer, Axel .
APPLIED PHYSICS LETTERS, 2010, 97 (26)
[8]   Influence of PCR Reagents on DNA Polymerase Extension Rates Measured on Real-Time PCR Instruments [J].
Montgomery, Jesse L. ;
Wittwer, Carl T. .
CLINICAL CHEMISTRY, 2014, 60 (02) :334-340
[9]   Stopped-flow DNA polymerase assay by continuous monitoring of dNTP incorporation by fluorescence [J].
Montgomery, Jesse L. ;
Rejali, Nick ;
Wittwer, Carl T. .
ANALYTICAL BIOCHEMISTRY, 2013, 441 (02) :133-139
[10]  
MULLIS K.B., 1994, POLYMERASE CHAIN REA