The use of static pressures of heavy gases within a quadrupole ion trap

被引:39
作者
Danell, RM
Danell, AS
Glish, GL
Vachet, RW
机构
[1] Univ N Carolina, Dept Chem, Chapel Hill, NC 27514 USA
[2] Univ Massachusetts, Dept Chem, Amherst, MA 01003 USA
关键词
D O I
10.1016/S1044-0305(03)00404-5
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The performance of quadrupole ion traps using argon or air as the buffer gas was evaluated and compared to the standard helium only operation. In all cases a pure buffer gas, not mixtures of gases, was investigated. Experiments were performed on a Bruker Esquire ion trap, a Finnigan LCQ, and a Finnigan ITMS for comparison. The heavier gases were found to have some advantages, particularly in the areas of sensitivity and collision-induced dissociation efficiency; however, there is a significant resolution loss due to dissociation and/or scattering of ions. Additionally, the heavier gases were found to affect ion activation and deactivation during MS/MS, influencing the product ion intensities observed. Finally, the specific quadrupole ion trap design and the ion ejection parameters were found to be crucial in the quality of the spectra obtained in the presence of heavy gases. Operation with static pressures of heavy gases can be beneficial under certain design and operating conditions of the quadrupole ion trap. (C) 2003 American Society for Mass Spectrometry.
引用
收藏
页码:1099 / 1109
页数:11
相关论文
共 33 条
[1]  
Asano KG, 1999, J MASS SPECTROM, V34, P691, DOI 10.1002/(SICI)1096-9888(199906)34:6<691::AID-JMS823>3.3.CO
[2]  
2-K
[3]   Bath gas temperature and the appearance of ion trap tandem mass spectra of high-mass ions [J].
Asano, KG ;
Goeringer, DE ;
Butcher, DJ ;
McLuckey, SA .
INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 1999, 191 :281-293
[4]  
BARMESS JE, 1983, VACUUM, V33, P149
[5]  
BLACK DM, 2002, P 50 ASMS C MASS SPE
[6]   COMPETITION BETWEEN RESONANCE EJECTION AND ION DISSOCIATION DURING RESONANT EXCITATION IN A QUADRUPOLE ION-TRAP [J].
CHARLES, MJ ;
MCLUCKEY, SA ;
GLISH, GL .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 1994, 5 (12) :1031-1041
[7]  
DANELL AS, UNPUB
[8]   Pulsed gas introduction for increasing peptide CID efficiency in a MALDI quadrupole ion trap mass spectrometer [J].
Doroshenko, VM ;
Cotter, RJ .
ANALYTICAL CHEMISTRY, 1996, 68 (03) :463-472
[9]   THE NONLINEAR ION-TRAP .4. MASS-SELECTIVE INSTABILITY SCAN WITH MULTIPOLE SUPERPOSITION [J].
FRANZEN, J .
INTERNATIONAL JOURNAL OF MASS SPECTROMETRY AND ION PROCESSES, 1993, 125 (2-3) :165-170
[10]   SIMULATION STUDY OF AN ION CAGE WITH SUPERIMPOSED MULTIPOLE FIELDS [J].
FRANZEN, J .
INTERNATIONAL JOURNAL OF MASS SPECTROMETRY AND ION PROCESSES, 1991, 106 :63-78