Multimodal intraoperative monitoring (MIOM) during cervical spine surgical procedures in 246 patients

被引:0
作者
Eggspuehler, Andreas [1 ]
Sutter, Martin A. [1 ]
Grob, Dieter [1 ]
Jeszenszky, Dezso [1 ]
Porchet, Francois [1 ]
Dvorak, Jiri [1 ]
机构
[1] Schulthess Clin, Dept Neurol, Spine Unit, CH-8008 Zurich, Switzerland
关键词
cervical spine surgery; intraoperative monitoring; sensitivity; specificity;
D O I
暂无
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
A prospective study of 246 patients who received multimodal intraoperative monitoring during cervical spine surgery between March 2000 and December 2005. To determine the sensitivity and specificity of MIOM techniques used to monitor spinal cord and nerve root function during cervical spine surgery. It is appreciated that complication rate of cervical spine surgery is low, however, there is a significant risk of neurological injury. The combination of monitoring of ascending and descending pathways may provide more sensitive and specific results giving immediate feedback information and/or alert regarding any neurological changes during the operation to the surgeon. Intraoperative somatosensory spinal and cerebral evoked potentials combined with continuous EMG and motor-evoked potentials of the spinal cord and muscles were evaluated and compared with postoperative clinical neurological changes. A total of 246 consecutive patients with cervical pathologies, majority spinal stenosis due to degenerative changes of cervical spine were monitored by means of MIOM during the surgical procedure. About 232 patients presented true negative while 2 patients false negative responses. About ten patients presented true positive responses where neurological deficit after the operation was predicted and two patients presented false positive findings. The sensitivity of MIOM applied during cervical spine procedure (anterior and/or posterior) was 83.3% and specificity of 99.2%. MIOM is an effective method of monitoring the spinal cord functional integrity during cervical spine surgery and can help to reduce the risk of neurological deficit by alerting the surgeon when monitoring changes are observed.
引用
收藏
页码:S209 / S215
页数:7
相关论文
共 10 条
[1]  
Bose Bikash, 2004, Spine J, V4, P202, DOI 10.1016/j.spinee.2003.06.001
[2]   A prospective analysis of intraoperative electromyographic monitoring of posterior cervical screw fixation [J].
Djurasovic, M ;
Dimar, JR ;
Glassman, SD ;
Edmonds, HL ;
Carreon, LY .
JOURNAL OF SPINAL DISORDERS & TECHNIQUES, 2005, 18 (06) :515-518
[3]   ADULT INTRAMEDULLARY SPINAL-CORD EPENDYMOMAS - THE RESULT OF SURGERY IN 38 PATIENTS [J].
EPSTEIN, FJ ;
FARMER, JP ;
FREED, D .
JOURNAL OF NEUROSURGERY, 1993, 79 (02) :204-209
[4]   Palsies of the fifth cervical nerve root after cervical decompression: prevention using continuous intraoperative electromyography monitoring [J].
Jimenez, JC ;
Sani, S ;
Braverman, B ;
Deutsch, H ;
Ratliff, JK .
JOURNAL OF NEUROSURGERY-SPINE, 2005, 3 (02) :92-97
[5]   Intraoperative somatosensory evoked potential monitoring during cervical spine corpectomy surgery [J].
Khan, MH ;
Smith, PN ;
Balzer, JR ;
Crammond, D ;
Welch, WC ;
Gerszten, P ;
Sclabassi, RJ ;
Kang, JD ;
Donaldson, WF .
SPINE, 2006, 31 (04) :E105-E113
[6]   The disappearing evoked potentials: A special problem of positioning patients with skeletal dysplasia - Case report [J].
Ofiram, Elisha ;
Lonstein, John E. ;
Skinner, Stan ;
Perra, Joseph H. .
SPINE, 2006, 31 (14) :E464-E470
[7]  
Sala F, 2004, ADV TECH STAND NEUR, V29, P133
[8]   Motor evoked potential monitoring improves outcome after surgery for intramedullary spinal cord tumors: A historical control study [J].
Sala, Francesco ;
Palandri, Giorgio ;
Basso, Elisabetta ;
Lanteri, Paola ;
Deletis, Vedran ;
Faccioli, Franco ;
Bricolo, Albino .
NEUROSURGERY, 2006, 58 (06) :1129-1141
[9]  
SUTTER M, 2007, UNPUB EUR SPINE J
[10]   Somatosensory evoked potential monitoring during anterior cervical discectomy and fusion [J].
Taunt, CJ ;
Sidhu, KS ;
Andrew, SA .
SPINE, 2005, 30 (17) :1970-1972