Micro-SPECT combined with 3-D optical imaging

被引:0
作者
Tsyganov, EN [1 ]
Antich, PP [1 ]
Kulkarni, PV [1 ]
Mason, RP [1 ]
Parkey, RW [1 ]
Seliuonine, SY [1 ]
Shay, JW [1 ]
Soesbe, TW [1 ]
Zinchenko, AI [1 ]
机构
[1] Univ Texas, SW Med Ctr, Dallas, TX 75390 USA
来源
2004 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-7 | 2004年
关键词
image reconstruction; single photon; emission computed tomography; small animal imaging; optical imaging;
D O I
暂无
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
A resolution enhanced NaI(TI)-scintillator micro-SPECT device using a pinhole collimator geometry has been built and tested with small animals. This device was constructed based on a depth-of-interaction (DOI) measurement using a thick scintilator crystal and a position sensitive PNIT to measure depth dependent scintillator light profiles. DOI measurement eliminates the parallax error that degrades the high spatial resolution required for small animal imaging. This novel technique for 3-D gamma-ray detection was incorporated into a micro-SPECT device and tested with a Co-57 source and Tc-98m-MDP in mice. We are now investigating the combination of micro-SPECT with 3-D optical imaging. so that the interaction of different biological aspects can e studied by each technique. A simple application permits a 3-D reconstruction of the animal ' s skin surface thus enhancing visualization and making possible depth-dependent corrections and a quantitative description of biological processes can be achieved. The combination of micro-SPECT and 3-D optical imaging, including bioluminescence 3-D reconstruction, opens new avenues in multi-modality imaging for biology and medicine.
引用
收藏
页码:3275 / 3279
页数:5
相关论文
共 10 条
  • [1] 3D position readout from thick scintillators
    Antich, P
    Malakhov, N
    Parkey, R
    Slavin, N
    Tsyganov, E
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2002, 480 (2-3) : 782 - 787
  • [2] ANTICH P, 2001, IEE NUCL SCI S MIC 2
  • [3] ANTICH P, IN PRESS IEEE T NUCL
  • [4] EDWARD N, 2004, ARXIVORGPHYSICS04061
  • [5] EDWARD N, IN PRESS IEEE T NUCL
  • [6] One-pass list-mode EM algorithm for high-resolution 3-D PET image reconstruction into large arrays
    Reader, AJ
    Ally, S
    Bakatselos, F
    Manavaki, R
    Walledge, RJ
    Jeavons, AP
    Julyan, PJ
    Zhao, S
    Hastings, DL
    Zweit, J
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2002, 49 (03) : 693 - 699
  • [7] RICHER E, 2004, REP SMI SAINT LOUIS
  • [8] Shepp L A, 1982, IEEE Trans Med Imaging, V1, P113, DOI 10.1109/TMI.1982.4307558
  • [9] TSYGANOV E, 2003, IEEE NUCL SCI S MIC
  • [10] A MONTE-CARLO MODEL FOR THE ABSORPTION AND FLUX DISTRIBUTIONS OF LIGHT IN TISSUE
    WILSON, BC
    ADAM, G
    [J]. MEDICAL PHYSICS, 1983, 10 (06) : 824 - 830